Characterization of myocardial perfusion imaging systems - an extension of quality metrics

IF 3.3 3区 医学 Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Physica Medica-European Journal of Medical Physics Pub Date : 2024-09-01 DOI:10.1016/j.ejmp.2024.104510
{"title":"Characterization of myocardial perfusion imaging systems - an extension of quality metrics","authors":"","doi":"10.1016/j.ejmp.2024.104510","DOIUrl":null,"url":null,"abstract":"<div><h3>Purpose</h3><p>The aim of this study was to evaluate the performance of myocardial perfusion imaging (MPI) systems in detecting perfusion defects (PDs). The defect perfusion index (DPI) was introduced to extend and further advance the current MPI quality metrics.</p></div><div><h3>Methods</h3><p>An anthropomorphic phantom simulating normal and pathological myocardial perfusion conditions was imaged by various NaI-crystal detector systems with and without corrections for scatter (SC) and attenuation (AC) (Symbia, Symbia + SC, Symbia IQ + SCAC, Symbia IQ), and cadmium-zinc-telluride detector systems without corrections (DSPECT, D530c). The extent of PD and the summed score (SS) were obtained by comparing polar maps with ad hoc normal databases created for each MPI system by using phantom polar maps with normal perfusion. The segmental uptake (SU) and the global uniformity (GU) were evaluated. The DPI was calculated on segments included in the PD to minimize attenuation artifacts outside the PD. The 17 segmental model was used.</p></div><div><h3>Results</h3><p>The highest level of uniformity of polar map was obtained for Symbia IQ + SCAC. D530c showed the highest extent of PD and dependence of the extent on the PD position. It showed in general the lowest SU values and the highest GU due to attenuation artifacts. Nevertheless, D530c outperforms other MPI systems in terms of PD detection, showing the highest DPI value. DSPECT system showed the lowest SS value, and DPI values comparable to NaI-crystal detector systems.</p></div><div><h3>Conclusion</h3><p>The DPI can be evaluated to investigate the intrinsic ability of MPI systems to detect PDs, whatever the quantitative post-processing software used.</p></div>","PeriodicalId":56092,"journal":{"name":"Physica Medica-European Journal of Medical Physics","volume":null,"pages":null},"PeriodicalIF":3.3000,"publicationDate":"2024-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Physica Medica-European Journal of Medical Physics","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1120179724002552","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose

The aim of this study was to evaluate the performance of myocardial perfusion imaging (MPI) systems in detecting perfusion defects (PDs). The defect perfusion index (DPI) was introduced to extend and further advance the current MPI quality metrics.

Methods

An anthropomorphic phantom simulating normal and pathological myocardial perfusion conditions was imaged by various NaI-crystal detector systems with and without corrections for scatter (SC) and attenuation (AC) (Symbia, Symbia + SC, Symbia IQ + SCAC, Symbia IQ), and cadmium-zinc-telluride detector systems without corrections (DSPECT, D530c). The extent of PD and the summed score (SS) were obtained by comparing polar maps with ad hoc normal databases created for each MPI system by using phantom polar maps with normal perfusion. The segmental uptake (SU) and the global uniformity (GU) were evaluated. The DPI was calculated on segments included in the PD to minimize attenuation artifacts outside the PD. The 17 segmental model was used.

Results

The highest level of uniformity of polar map was obtained for Symbia IQ + SCAC. D530c showed the highest extent of PD and dependence of the extent on the PD position. It showed in general the lowest SU values and the highest GU due to attenuation artifacts. Nevertheless, D530c outperforms other MPI systems in terms of PD detection, showing the highest DPI value. DSPECT system showed the lowest SS value, and DPI values comparable to NaI-crystal detector systems.

Conclusion

The DPI can be evaluated to investigate the intrinsic ability of MPI systems to detect PDs, whatever the quantitative post-processing software used.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
心肌灌注成像系统的特征--质量指标的延伸
目的本研究旨在评估心肌灌注成像(MPI)系统在检测灌注缺损(PD)方面的性能。方法模拟正常和病理心肌灌注条件的拟人化模型,用不同的碘化钠晶体探测器系统(带或不带散射(SC)和衰减(AC)校正功能)(Symbia、Symbia + SC、Symbia IQ + SCAC、Symbia IQ)和不带校正功能的碲锌镉探测器系统(DSPECT、D530c)进行成像。通过使用正常灌注的幻影极坐标图,将极坐标图与为每种 MPI 系统创建的特设正常数据库进行比较,从而得出 PD 的范围和总分 (SS)。评估了节段摄取(SU)和全局均匀性(GU)。DPI 是根据 PD 中的节段计算的,以尽量减少 PD 以外的衰减伪影。结果 Symbia IQ + SCAC 获得的极坐标图均匀度最高。D530c 显示的极坐标范围最大,而且其范围与极坐标位置有关。总体而言,D530c 的 SU 值最低,衰减伪影导致的 GU 值最高。尽管如此,D530c 在 PD 检测方面仍优于其他 MPI 系统,显示出最高的 DPI 值。DSPECT系统的SS值最低,DPI值与NaI晶体探测器系统相当。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
CiteScore
6.80
自引率
14.70%
发文量
493
审稿时长
78 days
期刊介绍: Physica Medica, European Journal of Medical Physics, publishing with Elsevier from 2007, provides an international forum for research and reviews on the following main topics: Medical Imaging Radiation Therapy Radiation Protection Measuring Systems and Signal Processing Education and training in Medical Physics Professional issues in Medical Physics.
期刊最新文献
Comparison of two biologically effective dose calculation models applied to single fraction stereotactic radiosurgery Advanced mathematical modeling for preciseestimation of CT energy spectrum using a calibration phantom LatticeOpt: An automatic tool for planning optimisation of spatially fractionated stereotactic body radiotherapy Development of a Monte Carlo simulation platform for the systematic evaluation of photon-counting detector-based micro-CT Investigate potential clinical benefits and linear energy transfer sparing utilizing proton arc therapy for hepatocellular carcinoma
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1