Diagnostic value of a point-of-care cardiac troponin-I assay (i-STAT®) for clinical application in canine and feline cardiology

IF 1.5 2区 农林科学 Q2 VETERINARY SCIENCES Journal of Veterinary Cardiology Pub Date : 2024-08-22 DOI:10.1016/j.jvc.2024.07.006
L. Ferasin , H. Ferasin , J. Farminer , E. Hudson , K. Lamb
{"title":"Diagnostic value of a point-of-care cardiac troponin-I assay (i-STAT®) for clinical application in canine and feline cardiology","authors":"L. Ferasin ,&nbsp;H. Ferasin ,&nbsp;J. Farminer ,&nbsp;E. Hudson ,&nbsp;K. Lamb","doi":"10.1016/j.jvc.2024.07.006","DOIUrl":null,"url":null,"abstract":"<div><h3><strong>Introduction/Objectives</strong></h3><p>This study was performed to evaluate the diagnostic value and net benefits, including cost-effectiveness, of a point-of-care analyser (i-STAT®) for measurement of cardiac troponin I (cTnI) in dogs and cats.</p></div><div><h3><strong>Animals, Materials and Methods</strong></h3><p>120 dogs and 120 cats presented with signs of cardiac disease and suspected myocardial insult on cardiac assessment. This was a validation study expressed as agreement between the i-STAT® analyser and two common commercial reference methods (IMMULITE® 2000 and ACCESS® hsTnI).</p></div><div><h3><strong>Results</strong></h3><p>The comparison between methods showed a negative bias between the i–STAT® and the two commercial cTnI assays. The bias was more evident when the i-STAT® values were compared to the traditional cTnI assay (IMMULITE® 2000), with a calculated difference of −1.14 ng/mL (dogs) and −0.96 ng/mL (cats). However, the bias was distinctly lower when the i-STAT® measurements were compared to the high-sensitivity cTnI assay (ACCESS®), namely −0.3 ng/mL in dogs and −0.17 in cats.</p></div><div><h3><strong>Discussion</strong></h3><p>The i-STAT® method can reliably detect normal, low and elevated cTnI values, which is fundamental to differentiate pets with and without myocardial damage and, with the rapid availability of results, this confirms the clinical utility of the i-STAT® method.</p></div><div><h3>Conclusions</h3><p>cTnI concentrations measured with the i-STAT® have good comparability with those obtained with both commercial assays for low and elevated cTnI values. However, results should be cautiously interpreted for high troponin values, especially if a strict cut-off value is adopted for diagnostic or prognostic purposes in critical clinical conditions, such as myocarditis or acute myocardial ischaemia.</p></div>","PeriodicalId":48788,"journal":{"name":"Journal of Veterinary Cardiology","volume":"56 ","pages":"Pages 35-43"},"PeriodicalIF":1.5000,"publicationDate":"2024-08-22","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Veterinary Cardiology","FirstCategoryId":"97","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1760273424000791","RegionNum":2,"RegionCategory":"农林科学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"VETERINARY SCIENCES","Score":null,"Total":0}
引用次数: 0

Abstract

Introduction/Objectives

This study was performed to evaluate the diagnostic value and net benefits, including cost-effectiveness, of a point-of-care analyser (i-STAT®) for measurement of cardiac troponin I (cTnI) in dogs and cats.

Animals, Materials and Methods

120 dogs and 120 cats presented with signs of cardiac disease and suspected myocardial insult on cardiac assessment. This was a validation study expressed as agreement between the i-STAT® analyser and two common commercial reference methods (IMMULITE® 2000 and ACCESS® hsTnI).

Results

The comparison between methods showed a negative bias between the i–STAT® and the two commercial cTnI assays. The bias was more evident when the i-STAT® values were compared to the traditional cTnI assay (IMMULITE® 2000), with a calculated difference of −1.14 ng/mL (dogs) and −0.96 ng/mL (cats). However, the bias was distinctly lower when the i-STAT® measurements were compared to the high-sensitivity cTnI assay (ACCESS®), namely −0.3 ng/mL in dogs and −0.17 in cats.

Discussion

The i-STAT® method can reliably detect normal, low and elevated cTnI values, which is fundamental to differentiate pets with and without myocardial damage and, with the rapid availability of results, this confirms the clinical utility of the i-STAT® method.

Conclusions

cTnI concentrations measured with the i-STAT® have good comparability with those obtained with both commercial assays for low and elevated cTnI values. However, results should be cautiously interpreted for high troponin values, especially if a strict cut-off value is adopted for diagnostic or prognostic purposes in critical clinical conditions, such as myocarditis or acute myocardial ischaemia.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
床旁心肌肌钙蛋白-I 检测法(i-STAT®)在犬猫心脏病学临床应用中的诊断价值
这项研究旨在评估用于测量猫狗心肌肌钙蛋白 I (cTnI) 的护理点分析仪 (i-STAT®) 的诊断价值和净效益,包括成本效益。120 只狗和 120 只猫在心脏评估时出现心脏疾病症状并怀疑心肌受损。这是一项以 i-STAT® 分析仪与两种常用商业参考方法(IMMULITE® 2000 和 ACCESS® hsTnI)之间的一致性表示的验证研究。方法间的比较显示,i-STAT® 和两种商用 cTnI 检测方法之间存在负偏差。当 i-STAT® 值与传统 cTnI 检测法(IMMULITE® 2000)进行比较时,总体负偏差更为明显,计算得出的差异为-1.14 纳克/毫升(狗)和-0.96 纳克/毫升(猫)。然而,当 i-STAT® 测量值与高灵敏度 cTnI 检测法(ACCESS®)相比时,偏差明显降低,即犬为-0.3 纳克/毫升,猫为-0.17 纳克/毫升。本研究结果表明,使用 i-STAT® 分析仪测量的 cTnI 浓度与使用 IMMULITE® 2000 或 ACCESS ® 检测仪测量的低值和高值 cTnI 浓度具有良好的可比性。不过,对于高托品蛋白值的结果应谨慎解读,尤其是在心肌炎或急性心肌缺血等危重临床情况下,如果采用严格的临界值进行诊断或预后时。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Journal of Veterinary Cardiology
Journal of Veterinary Cardiology VETERINARY SCIENCES-
CiteScore
2.50
自引率
25.00%
发文量
66
审稿时长
154 days
期刊介绍: The mission of the Journal of Veterinary Cardiology is to publish peer-reviewed reports of the highest quality that promote greater understanding of cardiovascular disease, and enhance the health and well being of animals and humans. The Journal of Veterinary Cardiology publishes original contributions involving research and clinical practice that include prospective and retrospective studies, clinical trials, epidemiology, observational studies, and advances in applied and basic research. The Journal invites submission of original manuscripts. Specific content areas of interest include heart failure, arrhythmias, congenital heart disease, cardiovascular medicine, surgery, hypertension, health outcomes research, diagnostic imaging, interventional techniques, genetics, molecular cardiology, and cardiovascular pathology, pharmacology, and toxicology.
期刊最新文献
Echocardiographic estimates of stroke volume in healthy dogs: comparability, reference intervals, and reproducibility Efficacy and safety of different antiarrhythmic protocols used for rate control in dogs with secondary atrial fibrillation Diagnosis and hybrid interventional and surgical removal of ectopic heartworms (Dirofilaria immitis) in a dog Transjugular patent ductus arteriosus occlusion in a cat using the Vet-PDA Occluder™ device Agreement of echocardiographic and catheterization-based methods of transpulmonary pressure gradient measurement in dogs
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1