Evaluation of axial displacement and torque loss of Morse-type prosthetic abutments of different angular tapers to their respective implants

IF 4.3 2区 医学 Q1 DENTISTRY, ORAL SURGERY & MEDICINE Journal of Prosthetic Dentistry Pub Date : 2024-12-01 DOI:10.1016/j.prosdent.2024.08.006
José Roberto de Freitas PhD , Mariana Lima da Costa Valente PhD , Elisabeth Helena Brazão DDS , Izabela Cristina Maurício Moris PhD , João Felipe Bonatto Bruniera PhD , Yara Teresinha Corrêa Silva-Sousa PhD , Edson Alfredo PhD
{"title":"Evaluation of axial displacement and torque loss of Morse-type prosthetic abutments of different angular tapers to their respective implants","authors":"José Roberto de Freitas PhD ,&nbsp;Mariana Lima da Costa Valente PhD ,&nbsp;Elisabeth Helena Brazão DDS ,&nbsp;Izabela Cristina Maurício Moris PhD ,&nbsp;João Felipe Bonatto Bruniera PhD ,&nbsp;Yara Teresinha Corrêa Silva-Sousa PhD ,&nbsp;Edson Alfredo PhD","doi":"10.1016/j.prosdent.2024.08.006","DOIUrl":null,"url":null,"abstract":"<div><h3>Statement of problem</h3><div>The biomechanical stability of the implant-prosthesis assembly and its maintenance under function is a determining factor in the success of implant prosthesis rehabilitation, but studies of different angular tapers are lacking.</div></div><div><h3>Purpose</h3><div>The purpose of this in vitro study was to evaluate the axial displacement and torque loss of prosthetic abutments with Morse-type connections of different angular tapers after thermomechanical cycling by using microcomputed tomography (µCT) and a digital torque wrench.</div></div><div><h3>Material and methods</h3><div>Eighteen Ø3.5×11.5-mm implants were embedded in polyvinyl chloride cylinders, and the 3 different types of abutments (n=6) with angular tapers of 11.5 degrees - Alvim Cone Morse (ACM-11.5), 16.0 degrees - Helix Gran Morse (HGM-16), and 24.0 degrees - Nobel Replace Conical Connection (NRC-24) were installed with the torque recommended by the manufacturers. To calculate the axial displacements of the abutments about the implants, zirconia maxillary canine crowns were fabricated using a computer-aided design and computer-aided manufacturing (CAD-CAM) system and cemented onto the abutments. Before and after thermomechanical cycling, the assemblies were scanned using microtomography (micro-CT) to assess axial displacement, and the torque losses were calculated after the abutments were finally unscrewed. A load of 100 N, frequency of 2 Hz, and 10<sup>6</sup> cycles with temperature variation of 5 °C to 55 °C were used for the thermomechanical cycling. Analysis of variance and the Tukey post hoc test (α=.05) were used for analysis.</div></div><div><h3>Results</h3><div>A significant difference was observed between the abutments for axial displacement, measured after thermomechanical cycling (<em>P</em>=.002). The ACM-11.5 abutment showed the highest mean value (134.1 ±58.7 µm), different from HGM-16 (63.3 ±26.1 µm) (<em>P</em>=.013) and NRC-24 (42.7 ±8.7 µm) (<em>P</em>=.002); the 2 latter groups were similar to each other (<em>P</em>=.618). For the torque losses, no significant difference was found among the abutments (<em>P</em>=.928), but there were significant differences for the thermomechanical cycling (before and after) (<em>P</em>&lt;.001) in that the loss of torque was greater after thermocycling.</div></div><div><h3>Conclusions</h3><div>The smaller the taper angle of the Morse-type prosthetic abutment, the greater its axial displacement, and the thermomechanical cycling significantly reduced pretorque, regardless of taper.</div></div>","PeriodicalId":16866,"journal":{"name":"Journal of Prosthetic Dentistry","volume":"132 6","pages":"Pages 1311.e1-1311.e7"},"PeriodicalIF":4.3000,"publicationDate":"2024-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Prosthetic Dentistry","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0022391324005717","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 0

Abstract

Statement of problem

The biomechanical stability of the implant-prosthesis assembly and its maintenance under function is a determining factor in the success of implant prosthesis rehabilitation, but studies of different angular tapers are lacking.

Purpose

The purpose of this in vitro study was to evaluate the axial displacement and torque loss of prosthetic abutments with Morse-type connections of different angular tapers after thermomechanical cycling by using microcomputed tomography (µCT) and a digital torque wrench.

Material and methods

Eighteen Ø3.5×11.5-mm implants were embedded in polyvinyl chloride cylinders, and the 3 different types of abutments (n=6) with angular tapers of 11.5 degrees - Alvim Cone Morse (ACM-11.5), 16.0 degrees - Helix Gran Morse (HGM-16), and 24.0 degrees - Nobel Replace Conical Connection (NRC-24) were installed with the torque recommended by the manufacturers. To calculate the axial displacements of the abutments about the implants, zirconia maxillary canine crowns were fabricated using a computer-aided design and computer-aided manufacturing (CAD-CAM) system and cemented onto the abutments. Before and after thermomechanical cycling, the assemblies were scanned using microtomography (micro-CT) to assess axial displacement, and the torque losses were calculated after the abutments were finally unscrewed. A load of 100 N, frequency of 2 Hz, and 106 cycles with temperature variation of 5 °C to 55 °C were used for the thermomechanical cycling. Analysis of variance and the Tukey post hoc test (α=.05) were used for analysis.

Results

A significant difference was observed between the abutments for axial displacement, measured after thermomechanical cycling (P=.002). The ACM-11.5 abutment showed the highest mean value (134.1 ±58.7 µm), different from HGM-16 (63.3 ±26.1 µm) (P=.013) and NRC-24 (42.7 ±8.7 µm) (P=.002); the 2 latter groups were similar to each other (P=.618). For the torque losses, no significant difference was found among the abutments (P=.928), but there were significant differences for the thermomechanical cycling (before and after) (P<.001) in that the loss of torque was greater after thermocycling.

Conclusions

The smaller the taper angle of the Morse-type prosthetic abutment, the greater its axial displacement, and the thermomechanical cycling significantly reduced pretorque, regardless of taper.
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
评估不同角度锥度的莫尔斯型修复基台与各自种植体之间的轴向位移和扭矩损失。
目的本体外研究的目的是通过使用微计算机断层扫描(μCT)和数字扭矩扳手,评估不同角度锥度的莫氏型连接修复基台在热机械循环后的轴向位移和扭矩损失。材料和方法将 18 个直径为 3.5×11.5 mm 的种植体嵌入聚氯乙烯圆柱体中,按照制造商推荐的扭矩安装 3 种不同类型的基台(n=6),其角度锥度分别为 11.5 度 - Alvim Cone Morse (ACM-11.5)、16.0 度 - Helix Gran Morse (HGM-16) 和 24.0 度 - Nobel Replace Conical Connection (NRC-24)。为了计算基台对种植体的轴向位移,使用计算机辅助设计和计算机辅助制造(CAD-CAM)系统制作了氧化锆上颌犬冠,并将其粘结在基台上。在热力学循环之前和之后,使用显微层析成像(micro-CT)扫描组件以评估轴向位移,并计算基台最终拧下后的扭矩损失。热机械循环采用的载荷为 100 N,频率为 2 Hz,循环次数为 106 次,温度变化范围为 5 °C 至 55 °C。结果在热机械循环后测量的轴向位移方面,观察到不同基台之间存在显著差异(P=.002)。ACM-11.5 基台的平均值最高(134.1 ±58.7 µm),与 HGM-16 基台(63.3 ±26.1 µm)(P=.013)和 NRC-24 基台(42.7 ±8.7 µm)(P=.002)不同;后两组基台的平均值相近(P=.618)。结论Morse型修复基台的锥度角越小,其轴向位移越大,无论锥度如何,热机械循环都会显著降低预扭力。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Journal of Prosthetic Dentistry
Journal of Prosthetic Dentistry 医学-牙科与口腔外科
CiteScore
7.00
自引率
13.00%
发文量
599
审稿时长
69 days
期刊介绍: The Journal of Prosthetic Dentistry is the leading professional journal devoted exclusively to prosthetic and restorative dentistry. The Journal is the official publication for 24 leading U.S. international prosthodontic organizations. The monthly publication features timely, original peer-reviewed articles on the newest techniques, dental materials, and research findings. The Journal serves prosthodontists and dentists in advanced practice, and features color photos that illustrate many step-by-step procedures. The Journal of Prosthetic Dentistry is included in Index Medicus and CINAHL.
期刊最新文献
Reliability, failure mode, and stress distribution in conical connections with different types of retention. A hybrid workflow for fabricating diagnostic trial restorations using a two-piece additively manufactured index. Adjacent free gingival grafting in the anterior region to avoid color inconsistency: A clinical report. Can researchers attract serendipity? Effect of different preparations and restorative materials on partial posterior restorations: A 3D FEA study using μCT data.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1