Global Perspectives on Mycotoxin Reference Materials (Part I): Insights from Multi-Supplier Comparison Study Including Aflatoxin B1, Deoxynivalenol and Zearalenone
David Steiner, Tibor Bartók, Michael Sulyok, András Szekeres, Mónika Varga, Levente Horváth, Helmut Rost
{"title":"Global Perspectives on Mycotoxin Reference Materials (Part I): Insights from Multi-Supplier Comparison Study Including Aflatoxin B1, Deoxynivalenol and Zearalenone","authors":"David Steiner, Tibor Bartók, Michael Sulyok, András Szekeres, Mónika Varga, Levente Horváth, Helmut Rost","doi":"10.3390/toxins16090397","DOIUrl":null,"url":null,"abstract":"We conducted a comprehensive examination of liquid mycotoxin reference standards. A total of 30 different standards were tested, each containing 10 samples of three distinct substances: Aflatoxin B1, Deoxynivalenol, and Zearalenone. The standards were sourced from 10 different global market leading manufacturers. To facilitate comparison, all the standard sets were adjusted to the same concentration level. The standards were analyzed using the techniques LC-MS/MS, HPLC-DAD, and LC-HRMS to assess their quality attributes. Regarding the validation of the reference values, it was observed that 30% of the suppliers provided reference standards that were either below the lower acceptance limit or above the higher acceptance limit, confirmed by both the LC-MS/MS and HPLC-DAD methods. Furthermore, a total of 12 impurities were found in the DON standards, 10 in the AFB1 standards, and 8 in the ZON standards, distributed across all the suppliers. Therefore, this study suggests relevant adjustments to the ISO 17034 standard, proposing that the purity of a raw material should be uniformly based on q-NMR analysis, as most manufacturers state the purity of their certificates is determined using HPLC-UV or LC-MS/MS. Liquid standards with a shelf life of ≤1 year should not exceed an uncertainty of 3%. Standards that have a longer shelf life should not have more than 5% uncertainty. This study also emphasizes the importance of stability. The standards should undergo continuous long-term monitoring; otherwise, products may exhibit a target value of only 80%, as seen in one instance. It is also recommended to include proof of HPLC and LC-MS/MS analyses on the certificate of each released batch of a final product.","PeriodicalId":23119,"journal":{"name":"Toxins","volume":"28 1","pages":""},"PeriodicalIF":3.9000,"publicationDate":"2024-09-17","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Toxins","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.3390/toxins16090397","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"FOOD SCIENCE & TECHNOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
We conducted a comprehensive examination of liquid mycotoxin reference standards. A total of 30 different standards were tested, each containing 10 samples of three distinct substances: Aflatoxin B1, Deoxynivalenol, and Zearalenone. The standards were sourced from 10 different global market leading manufacturers. To facilitate comparison, all the standard sets were adjusted to the same concentration level. The standards were analyzed using the techniques LC-MS/MS, HPLC-DAD, and LC-HRMS to assess their quality attributes. Regarding the validation of the reference values, it was observed that 30% of the suppliers provided reference standards that were either below the lower acceptance limit or above the higher acceptance limit, confirmed by both the LC-MS/MS and HPLC-DAD methods. Furthermore, a total of 12 impurities were found in the DON standards, 10 in the AFB1 standards, and 8 in the ZON standards, distributed across all the suppliers. Therefore, this study suggests relevant adjustments to the ISO 17034 standard, proposing that the purity of a raw material should be uniformly based on q-NMR analysis, as most manufacturers state the purity of their certificates is determined using HPLC-UV or LC-MS/MS. Liquid standards with a shelf life of ≤1 year should not exceed an uncertainty of 3%. Standards that have a longer shelf life should not have more than 5% uncertainty. This study also emphasizes the importance of stability. The standards should undergo continuous long-term monitoring; otherwise, products may exhibit a target value of only 80%, as seen in one instance. It is also recommended to include proof of HPLC and LC-MS/MS analyses on the certificate of each released batch of a final product.
期刊介绍:
Toxins (ISSN 2072-6651) is an international, peer-reviewed open access journal which provides an advanced forum for studies related to toxins and toxinology. It publishes reviews, regular research papers and short communications. Our aim is to encourage scientists to publish their experimental and theoretical results in as much detail as possible. There is no restriction on the length of the papers. The full experimental details must be provided so that the results can be reproduced.