{"title":"Accelerometer-based portable navigation system shows no superior accuracy over pelvic alignment guide for acetabular cup placement in total hip arthroplasty in lateral decubitus position.","authors":"Masanori Okamoto, Yasuhiko Takegami, Yusuke Osawa, Shinya Tanaka, Tatsuya Nobori, Shiro Imagama","doi":"10.1007/s00402-024-05571-4","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Total hip arthroplasty (THA) using a portable navigation system (PNS) incurs costs per procedure. However, it does not require a large console. This study aimed to compare the accuracy of acetabular cup placement using a pelvic alignment guide (PAG) attached to the pelvis and an accelerometer-based PNS in THA performed in the lateral decubitus position.</p><p><strong>Methods: </strong>We retrospectively analyzed 100 hips that underwent primary THA in the lateral decubitus position between July 2018 and January 2021. The PAG was used in 50 hips, whereas the PNS was used in the other 50. Cup placement accuracy was measured using postoperative computed tomography scans, comparing errors in inclination and anteversion angles. The surgical time, blood loss, and complications were recorded. The follow-up period was at least 2 years in all cases.</p><p><strong>Results: </strong>The mean absolute error of the inclination angle was similar between the groups (the PAG group: 3.7° ± 2.3° [range, 0.0-9.0]; the PNS group: 3.7° ± 2.3° [range, 0.2-10.5], p = 0.705). The mean absolute error of the anteversion angle was significantly smaller in the PAG group than in the PNS group (3.0° ± 2.4° [range, 0.0-9.7] vs. 6.5° ± 4.8° [range, 0.3-17.3], p < 0.001). The PAG group had a higher proportion of hips within 5° and 10° of the target angle (64 vs. 42%, P = 0.028, and 100 vs. 74%, p < 0.001, respectively). The PNS group had six hips with anteversion errors of 15° or more. Surgical time and blood loss were lower in the PAG group. The PNS group had one dislocation, whereas the PAG group did not.</p><p><strong>Conclusion: </strong>The accelerometer-based PNS did not demonstrate superior cup alignment accuracy compared to the PAG in THA performed in the lateral decubitus position. This finding informs surgeons that computer-assisted surgery is not necessarily superior to conventional THA using a PAG.</p>","PeriodicalId":8326,"journal":{"name":"Archives of Orthopaedic and Trauma Surgery","volume":" ","pages":"4849-4856"},"PeriodicalIF":2.0000,"publicationDate":"2024-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Archives of Orthopaedic and Trauma Surgery","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s00402-024-05571-4","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/9/30 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"ORTHOPEDICS","Score":null,"Total":0}
引用次数: 0
Abstract
Background: Total hip arthroplasty (THA) using a portable navigation system (PNS) incurs costs per procedure. However, it does not require a large console. This study aimed to compare the accuracy of acetabular cup placement using a pelvic alignment guide (PAG) attached to the pelvis and an accelerometer-based PNS in THA performed in the lateral decubitus position.
Methods: We retrospectively analyzed 100 hips that underwent primary THA in the lateral decubitus position between July 2018 and January 2021. The PAG was used in 50 hips, whereas the PNS was used in the other 50. Cup placement accuracy was measured using postoperative computed tomography scans, comparing errors in inclination and anteversion angles. The surgical time, blood loss, and complications were recorded. The follow-up period was at least 2 years in all cases.
Results: The mean absolute error of the inclination angle was similar between the groups (the PAG group: 3.7° ± 2.3° [range, 0.0-9.0]; the PNS group: 3.7° ± 2.3° [range, 0.2-10.5], p = 0.705). The mean absolute error of the anteversion angle was significantly smaller in the PAG group than in the PNS group (3.0° ± 2.4° [range, 0.0-9.7] vs. 6.5° ± 4.8° [range, 0.3-17.3], p < 0.001). The PAG group had a higher proportion of hips within 5° and 10° of the target angle (64 vs. 42%, P = 0.028, and 100 vs. 74%, p < 0.001, respectively). The PNS group had six hips with anteversion errors of 15° or more. Surgical time and blood loss were lower in the PAG group. The PNS group had one dislocation, whereas the PAG group did not.
Conclusion: The accelerometer-based PNS did not demonstrate superior cup alignment accuracy compared to the PAG in THA performed in the lateral decubitus position. This finding informs surgeons that computer-assisted surgery is not necessarily superior to conventional THA using a PAG.
期刊介绍:
"Archives of Orthopaedic and Trauma Surgery" is a rich source of instruction and information for physicians in clinical practice and research in the extensive field of orthopaedics and traumatology. The journal publishes papers that deal with diseases and injuries of the musculoskeletal system from all fields and aspects of medicine. The journal is particularly interested in papers that satisfy the information needs of orthopaedic clinicians and practitioners. The journal places special emphasis on clinical relevance.