Huaiming Wang, Wenlong Feng, Xue Ren, Quan Tao, Liangliang Rong, Yiping P Du, Hui Dong
{"title":"Acquisition Acceleration of Ultra-low Field MRI with Parallel Imaging and Compressed Sensing in Microtesla Fields.","authors":"Huaiming Wang, Wenlong Feng, Xue Ren, Quan Tao, Liangliang Rong, Yiping P Du, Hui Dong","doi":"10.1109/TBME.2024.3466929","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>In recent years, ultra-low field (ULF) magnetic resonance imaging (MRI) has gained widespread attention due to its advantages, such as low cost, light weight, and portability. However, the low signal-to-noise ratio (SNR) leads to a long scan time. Herein, we study the acceleration performance of parallel imaging (PI) and compressed sensing (CS) in different kspace sampling strategies at 0.12 mT.</p><p><strong>Methods: </strong>This study employs phantoms to assess the efficiency of acceleration methods at ULF MRI, in which signals are detected by ultra-sensitive superconducting quantum interference devices (SQUIDs). We compare the performance of fast Fourier transform (FFT), generalized auto-calibrating partially parallel acquisitions (GRAPPA), and eigenvector-based SPIRiT (ESPIRiT) in Cartesian sampling, while also evaluating non-uniform FFT (NUFFT), GRAPPA operator gridding, and ESPIRiT in nonCartesian sampling. We design a resolution phantom to investigate the effectiveness of these methods in maintaining image resolution.</p><p><strong>Results: </strong>In Cartesian sampling, GRAPPA and ESPIRiT jointly regularized by total variation and ℓ1-norm (TVJℓ1 -ESPIRiT) methods reconstructed good-quality phantom images with an acceleration factor of R = 2. In contrast, TVJℓ1-ESPIRiT exhibited improved image quality and much less signal loss even for R = 4. In radial sampling, TVJℓ1-ESPIRiT reduced the acquisition time to 1.69 minutes at R = 4, with a respective improvement of 12.26 dB in peak SNR compared to NUFFT. The resolution phantom imaging showed that the reconstructions by PI and CS maintained the original resolution of 2 mm.</p><p><strong>Conclusion and significance: </strong>This study improves the practicality of ULF MRI at microtesla fields by implementing imaging acceleration with PI and CS in different k-space sampling.</p>","PeriodicalId":13245,"journal":{"name":"IEEE Transactions on Biomedical Engineering","volume":"PP ","pages":""},"PeriodicalIF":4.4000,"publicationDate":"2024-09-24","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"IEEE Transactions on Biomedical Engineering","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1109/TBME.2024.3466929","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 0
Abstract
Objective: In recent years, ultra-low field (ULF) magnetic resonance imaging (MRI) has gained widespread attention due to its advantages, such as low cost, light weight, and portability. However, the low signal-to-noise ratio (SNR) leads to a long scan time. Herein, we study the acceleration performance of parallel imaging (PI) and compressed sensing (CS) in different kspace sampling strategies at 0.12 mT.
Methods: This study employs phantoms to assess the efficiency of acceleration methods at ULF MRI, in which signals are detected by ultra-sensitive superconducting quantum interference devices (SQUIDs). We compare the performance of fast Fourier transform (FFT), generalized auto-calibrating partially parallel acquisitions (GRAPPA), and eigenvector-based SPIRiT (ESPIRiT) in Cartesian sampling, while also evaluating non-uniform FFT (NUFFT), GRAPPA operator gridding, and ESPIRiT in nonCartesian sampling. We design a resolution phantom to investigate the effectiveness of these methods in maintaining image resolution.
Results: In Cartesian sampling, GRAPPA and ESPIRiT jointly regularized by total variation and ℓ1-norm (TVJℓ1 -ESPIRiT) methods reconstructed good-quality phantom images with an acceleration factor of R = 2. In contrast, TVJℓ1-ESPIRiT exhibited improved image quality and much less signal loss even for R = 4. In radial sampling, TVJℓ1-ESPIRiT reduced the acquisition time to 1.69 minutes at R = 4, with a respective improvement of 12.26 dB in peak SNR compared to NUFFT. The resolution phantom imaging showed that the reconstructions by PI and CS maintained the original resolution of 2 mm.
Conclusion and significance: This study improves the practicality of ULF MRI at microtesla fields by implementing imaging acceleration with PI and CS in different k-space sampling.