A-306 High Sensitivity Capability of an Anti-müllerian Hormone Assay on the Beckman Coulter DxI 9000 Immunoassay Analyzer*, a Next Step Toward Enabling Ovarian Failure Staging?

IF 7.1 2区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY Clinical chemistry Pub Date : 2024-10-02 DOI:10.1093/clinchem/hvae106.301
S Kuklok, J Thomforde, L Pearson, K Pritchard, T Pfingston, C Bunkelman, L Pearson, T Her, B Bolstad, B Bilyeu, S Santiago, S Thornburgh, M Posnansky, N Malikowski-Hoffarth, K Walt, D Lovett, M Holland, M Szabo, C Carlson
{"title":"A-306 High Sensitivity Capability of an Anti-müllerian Hormone Assay on the Beckman Coulter DxI 9000 Immunoassay Analyzer*, a Next Step Toward Enabling Ovarian Failure Staging?","authors":"S Kuklok, J Thomforde, L Pearson, K Pritchard, T Pfingston, C Bunkelman, L Pearson, T Her, B Bolstad, B Bilyeu, S Santiago, S Thornburgh, M Posnansky, N Malikowski-Hoffarth, K Walt, D Lovett, M Holland, M Szabo, C Carlson","doi":"10.1093/clinchem/hvae106.301","DOIUrl":null,"url":null,"abstract":"Background Measurement of anti-müllerian hormone (AMH) has proven value in the evaluation of ovarian reserve using assays such as the Beckman Coulter Access AMH assay. If improved sensitivity of AMH measurement can be achieved, enhanced clinical utility during the assessment of reproductive aging and ovarian response to enable staging of ovarian failure is possible (Practice Committee of the American Society of Reproductive Medicine 2020, Cappola et al. 2023). The Beckman Coulter DxI 9000 Immunoassay Analyzer* has been designed with new technology that can be leveraged to improve assay sensitivity. Such technological advancements include the new Lumi-Phos PRO chemiluminescent substrate which yields markedly increased signal-to-noise, a new luminometer with increased dynamic range, and improved low-volume pipetting capabilities. Data herein summarize results from analytical verification studies of the Access AMH assay on the DxI 9000 Immunoassay Analyzer. Results demonstrate an example of DxI 9000 technologies enabling sensitivity claims that are up to 20-fold improved. * The official name is DxI 9000 Access Immunoassay Analyzer. Methods A method comparison study was performed based on CLSI guideline EP09c, 3rd ed. to compare results for the Access AMH assay on DxI 9000 to the predicate device, the Access AMH assay on the Access 2 Immunoassay System. Within-laboratory precision was also evaluated following CLSI EP05-A3. Finally, performance at low analyte levels was evaluated following CLSI EP17-A2, and linearity was evaluated following CLSI EP06-Ed2. Results Method comparison of the Access AMH assay on DxI 9000 compared to the Access AMH assay on the Access 2 yielded excellent agreement with a Passing-Bablok slope of 1.02 for N=126 samples. Further, bias estimates for a number of key medical decision points suggest minimal bias (≤ 2%) across the assay range, and minimal non-linearity was observed. 20-day imprecision studies yielded CV’s between 2 and 5% across a broad range of concentrations evaluated. Sensitivity studies yielded estimates of limit of quantitation (LoQ) between 0.001 and 0.003 ng/mL. This represents an approximately 10- to 20-fold improvement in sensitivity claims compared to the AMH assay on the Access 2 and other commercially available automated immunoassay analyzers. Conclusions The data herein present performance data for the Access AMH assay on the DxI 9000 Immunoassay Analyzer. The assay demonstrates excellent accuracy relative to the predicate device, while showing good linearity and imprecision. Of note, the Access AMH assay exhibits exceptional low-end performance on the DxI 9000 analyzer as demonstrated by the ability to enable sensitivity claims that are up to 20-fold improved in comparison to existing immunoassays for the measurement of AMH. The sensitivity capabilities of the Access AMH assay on DxI 9000 suggest an opportunity for the addition of further intended uses for scenarios where extremely low AMH measurements can provide additional value of clinical significance. Further studies are needed to fully understand additional clinical utilities of the Access AMH assay with high sensitivity capability on DxI 9000, including the assessment of reproductive aging for menopause diagnosis or ovarian response for fertility treatment.","PeriodicalId":10690,"journal":{"name":"Clinical chemistry","volume":"26 1","pages":""},"PeriodicalIF":7.1000,"publicationDate":"2024-10-02","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Clinical chemistry","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1093/clinchem/hvae106.301","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Background Measurement of anti-müllerian hormone (AMH) has proven value in the evaluation of ovarian reserve using assays such as the Beckman Coulter Access AMH assay. If improved sensitivity of AMH measurement can be achieved, enhanced clinical utility during the assessment of reproductive aging and ovarian response to enable staging of ovarian failure is possible (Practice Committee of the American Society of Reproductive Medicine 2020, Cappola et al. 2023). The Beckman Coulter DxI 9000 Immunoassay Analyzer* has been designed with new technology that can be leveraged to improve assay sensitivity. Such technological advancements include the new Lumi-Phos PRO chemiluminescent substrate which yields markedly increased signal-to-noise, a new luminometer with increased dynamic range, and improved low-volume pipetting capabilities. Data herein summarize results from analytical verification studies of the Access AMH assay on the DxI 9000 Immunoassay Analyzer. Results demonstrate an example of DxI 9000 technologies enabling sensitivity claims that are up to 20-fold improved. * The official name is DxI 9000 Access Immunoassay Analyzer. Methods A method comparison study was performed based on CLSI guideline EP09c, 3rd ed. to compare results for the Access AMH assay on DxI 9000 to the predicate device, the Access AMH assay on the Access 2 Immunoassay System. Within-laboratory precision was also evaluated following CLSI EP05-A3. Finally, performance at low analyte levels was evaluated following CLSI EP17-A2, and linearity was evaluated following CLSI EP06-Ed2. Results Method comparison of the Access AMH assay on DxI 9000 compared to the Access AMH assay on the Access 2 yielded excellent agreement with a Passing-Bablok slope of 1.02 for N=126 samples. Further, bias estimates for a number of key medical decision points suggest minimal bias (≤ 2%) across the assay range, and minimal non-linearity was observed. 20-day imprecision studies yielded CV’s between 2 and 5% across a broad range of concentrations evaluated. Sensitivity studies yielded estimates of limit of quantitation (LoQ) between 0.001 and 0.003 ng/mL. This represents an approximately 10- to 20-fold improvement in sensitivity claims compared to the AMH assay on the Access 2 and other commercially available automated immunoassay analyzers. Conclusions The data herein present performance data for the Access AMH assay on the DxI 9000 Immunoassay Analyzer. The assay demonstrates excellent accuracy relative to the predicate device, while showing good linearity and imprecision. Of note, the Access AMH assay exhibits exceptional low-end performance on the DxI 9000 analyzer as demonstrated by the ability to enable sensitivity claims that are up to 20-fold improved in comparison to existing immunoassays for the measurement of AMH. The sensitivity capabilities of the Access AMH assay on DxI 9000 suggest an opportunity for the addition of further intended uses for scenarios where extremely low AMH measurements can provide additional value of clinical significance. Further studies are needed to fully understand additional clinical utilities of the Access AMH assay with high sensitivity capability on DxI 9000, including the assessment of reproductive aging for menopause diagnosis or ovarian response for fertility treatment.
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
A-306 Beckman Coulter DxI 9000 免疫测定分析仪*上的高灵敏度抗苗勒管激素测定,实现卵巢功能衰竭分期的下一步?
背景使用贝克曼库尔特公司的Access AMH测定法等测定抗缪勒氏管激素(AMH)在评估卵巢储备功能方面的价值已得到证实。如果能提高 AMH 测量的灵敏度,就有可能在评估生殖衰老和卵巢反应时提高临床实用性,从而对卵巢功能衰竭进行分期(美国生殖医学会实践委员会 2020 年,Cappola 等 2023 年)。贝克曼库尔特 DxI 9000 免疫分析仪*的设计采用了可用于提高检测灵敏度的新技术。这些技术进步包括:可显著提高信噪比的新型 Lumi-Phos PRO 化学发光底物、可提高动态范围的新型发光仪以及改进的低容量移液能力。本文总结了在 DxI 9000 免疫分析仪上对 Access AMH 检测法进行分析验证研究的结果。结果展示了 DxI 9000 技术使灵敏度提高 20 倍的实例。* 正式名称为 DxI 9000 Access 免疫分析仪。方法 根据 CLSI 指南 EP09c 第 3 版进行了方法比较研究,将 DxI 9000 上的 Access AMH 检测结果与同类设备(Access 2 免疫测定系统上的 Access AMH 检测结果)进行比较。此外,还按照 CLSI EP05-A3 对实验室内的精密度进行了评估。最后,根据 CLSI EP17-A2 评估了低分析物水平下的性能,并根据 CLSI EP06-Ed2 评估了线性度。结果 将 DxI 9000 上的 Access AMH 检测法与 Access 2 上的 Access AMH 检测法进行方法比较,结果显示两者的一致性非常好,N=126 个样本的 Passing-Bablok 斜率为 1.02。此外,对一些关键医疗决策点的偏倚估计表明,整个检测范围内的偏倚极小(≤ 2%),而且观察到的非线性极小。20 天不精密度研究显示,在广泛的评估浓度范围内,CV 值介于 2% 和 5% 之间。灵敏度研究得出的定量限 (LoQ) 估计值介于 0.001 和 0.003 纳克/毫升之间。与 Access 2 和其他市售自动免疫分析仪上的 AMH 检测相比,灵敏度提高了约 10 到 20 倍。结论 本文数据显示了 Access AMH 检测法在 DxI 9000 免疫分析仪上的性能数据。该检测法与同类设备相比具有极佳的准确性,同时还显示出良好的线性和不精密度。值得注意的是,Access AMH 检测试剂盒在 DxI 9000 分析仪上表现出卓越的低端性能,其灵敏度比现有的 AMH 免疫测定试剂盒提高了 20 倍。DxI 9000 上 Access AMH 检测法的灵敏度能力表明,在极低 AMH 测量可提供具有临床意义的额外价值的情况下,有机会增加更多预期用途。要充分了解 DxI 9000 上高灵敏度的 Access AMH 检测法的其他临床用途,包括用于更年期诊断的生殖衰老评估或用于生育治疗的卵巢反应评估,还需要进一步的研究。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Clinical chemistry
Clinical chemistry 医学-医学实验技术
CiteScore
11.30
自引率
4.30%
发文量
212
审稿时长
1.7 months
期刊介绍: Clinical Chemistry is a peer-reviewed scientific journal that is the premier publication for the science and practice of clinical laboratory medicine. It was established in 1955 and is associated with the Association for Diagnostics & Laboratory Medicine (ADLM). The journal focuses on laboratory diagnosis and management of patients, and has expanded to include other clinical laboratory disciplines such as genomics, hematology, microbiology, and toxicology. It also publishes articles relevant to clinical specialties including cardiology, endocrinology, gastroenterology, genetics, immunology, infectious diseases, maternal-fetal medicine, neurology, nutrition, oncology, and pediatrics. In addition to original research, editorials, and reviews, Clinical Chemistry features recurring sections such as clinical case studies, perspectives, podcasts, and Q&A articles. It has the highest impact factor among journals of clinical chemistry, laboratory medicine, pathology, analytical chemistry, transfusion medicine, and clinical microbiology. The journal is indexed in databases such as MEDLINE and Web of Science.
期刊最新文献
Virtual Gene Panels Have a Superior Diagnostic Yield for Inherited Rare Diseases Relative to Static Panels Xylazine Pharmacokinetics in Patients Testing Positive for Fentanyl and Xylazine. Prospective and External Validation of an Ensemble Learning Approach to Sensitively Detect Intravenous Fluid Contamination in Basic Metabolic Panels Niacin and Risk of Cardiovascular Events: Deciphering the Paradox. Alzheimer Disease Blood-Based Biomarkers: Translation from Research into Clinical Use.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1