{"title":"Use the external radiotherapy StarTrack detector to study the electron beam profile.","authors":"Safa Sami Hassan","doi":"10.47391/JPMA-BAGH-16-18","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>To assess the electron beam dose verification, and to investigate the dependability and stability of the Startrack two-dimensional array.</p><p><strong>Methods: </strong>The is cross-secional study was conducted from January to April 2021 at the Baghdad Centre for Radiotherapy and Nuclear Medicine, Baghdad Medical City, Baghdad, Iraq. Quality assurance measurements were made using a StarTrack two-dimensional detector on an electron beam with an energy of 12MeV at 1.5cm depth for field sizes 6cm×6cm and 14cm×14cm. The detector was positioned 100cm from the source to the surface of the linear accelerator infinity. Testing was done using the International Electrotechnical Commissioning protocol.</p><p><strong>Results: </strong>There was a significant difference between field sizes 6cmx6cm and 14cmx14cm (-4.97±0.83, -2.01±0.68, respectively) (p<0.05). The beam profile measurements of penumbra, flatness and symmetry were within limits ±2 cm, and ±2 cm, 2%, respectively. The percentage of output dose curve against distance showed instability in crossline for 6cm x 6cm and 14cm x 14cm field size.</p><p><strong>Conclusions: </strong>The output dose was found to be above the tolerance level, and should be corrected before patient treatment for 12MeV energy electron beam therapy.</p>","PeriodicalId":54369,"journal":{"name":"Journal of the Pakistan Medical Association","volume":"74 10 (Supple-8)","pages":"S76-S80"},"PeriodicalIF":0.8000,"publicationDate":"2024-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of the Pakistan Medical Association","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.47391/JPMA-BAGH-16-18","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"MEDICINE, GENERAL & INTERNAL","Score":null,"Total":0}
引用次数: 0
Abstract
Objective: To assess the electron beam dose verification, and to investigate the dependability and stability of the Startrack two-dimensional array.
Methods: The is cross-secional study was conducted from January to April 2021 at the Baghdad Centre for Radiotherapy and Nuclear Medicine, Baghdad Medical City, Baghdad, Iraq. Quality assurance measurements were made using a StarTrack two-dimensional detector on an electron beam with an energy of 12MeV at 1.5cm depth for field sizes 6cm×6cm and 14cm×14cm. The detector was positioned 100cm from the source to the surface of the linear accelerator infinity. Testing was done using the International Electrotechnical Commissioning protocol.
Results: There was a significant difference between field sizes 6cmx6cm and 14cmx14cm (-4.97±0.83, -2.01±0.68, respectively) (p<0.05). The beam profile measurements of penumbra, flatness and symmetry were within limits ±2 cm, and ±2 cm, 2%, respectively. The percentage of output dose curve against distance showed instability in crossline for 6cm x 6cm and 14cm x 14cm field size.
Conclusions: The output dose was found to be above the tolerance level, and should be corrected before patient treatment for 12MeV energy electron beam therapy.
期刊介绍:
Primarily being a medical journal, JPMA publishes scholarly research focusing on the various fields in the areas of health and medical education. It publishes original research describing recent advances in health particularly clinical studies, clinical trials, assessments of pathogens of diagnostic importance, medical genetics and epidemiological studies. Review articles highlighting importance of various issues in the domain of public health, drug research and medical education are also accepted. As a leading journal of South Asia, JPMA remains cognizant of the recent advances in the rapidly growing fields of biomedical sciences, it invites and encourages scholars to write short reviews and invited editorials on the emerging issues. We particularly aim to promote health standards of developing countries by encouraging manuscript submissions on issues affecting the public health and health delivery services.