Urine organic acid metabolomic profiling by gas chromatography mass spectrometry: Assessment of solvent extract evaporation parameters on the recovery of key diagnostic metabolites
Rachel S. Carling , Karolina Witek , Erin C Emmett , Claire Gallagher , Stuart J. Moat
{"title":"Urine organic acid metabolomic profiling by gas chromatography mass spectrometry: Assessment of solvent extract evaporation parameters on the recovery of key diagnostic metabolites","authors":"Rachel S. Carling , Karolina Witek , Erin C Emmett , Claire Gallagher , Stuart J. Moat","doi":"10.1016/j.cca.2024.120015","DOIUrl":null,"url":null,"abstract":"<div><h3>Background</h3><div>Analysis of urinary organic acids (UOAs) by gas chromatography mass-spectrometry (GC–MS) is widely used in metabolomic studies. It is a complex test with many limitations and pitfalls yet there is limited evidence in the literature to support best practice. This study investigated the impact of drying down time and temperature on the recovery of 16 key analytes from solvent extracts.</div></div><div><h3>Methods</h3><div>Pooled urine specimens were enriched with organic acids. Urine aliquots (n = 3) were acidified and extracted into diethyl ether and ethyl acetate. Extracts were dried under nitrogen at ambient temperature (25 °C); 40 °C; 60 °C then left for 0; +5; +15 min. Dried extracts were derivatised with N,O,-bis-(trimethylsilyl)trifluoroacetamide prior to analysis by GC–MS. Urine specimens from individuals with biotinidase deficiency, maple syrup urine disease (MSUD) and ketotic hypoglycemia were analysed to demonstrate the potential clinical impact.</div></div><div><h3>Results</h3><div>Recovery of shorter chain hydroxycarboxylic acids decreased significantly when extracts were dried above 25 °C (mean recovery 89 % at 60 °C, p < 0.01) or left under nitrogen post-drying (mean recovery at ambient + 15 min, 40 °C + 15mins and 60 °C + 15mins was 56 %, 12 % and 2 %, respectively, p < 0.01). Whilst dicarboxylic acids/medium chain fatty acids were unaffected by temperature (mean recovery 100 %), prolonged drying reduced recovery (mean recovery 85 % at 60 °C + 15mins, p < 0.01).</div></div><div><h3>Conclusions</h3><div>Evaporation of solvent extracts with heat and/or prolonged drying under nitrogen results in significant losses of the shorter chain hydroxycarboxylic acids. The evaporation protocol must be carefully controlled to ensure accurate and reproducible results, preventing misdiagnoses and/or misinterpretation of results.</div></div>","PeriodicalId":10205,"journal":{"name":"Clinica Chimica Acta","volume":"565 ","pages":"Article 120015"},"PeriodicalIF":3.2000,"publicationDate":"2024-10-22","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Clinica Chimica Acta","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S000989812402268X","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Background
Analysis of urinary organic acids (UOAs) by gas chromatography mass-spectrometry (GC–MS) is widely used in metabolomic studies. It is a complex test with many limitations and pitfalls yet there is limited evidence in the literature to support best practice. This study investigated the impact of drying down time and temperature on the recovery of 16 key analytes from solvent extracts.
Methods
Pooled urine specimens were enriched with organic acids. Urine aliquots (n = 3) were acidified and extracted into diethyl ether and ethyl acetate. Extracts were dried under nitrogen at ambient temperature (25 °C); 40 °C; 60 °C then left for 0; +5; +15 min. Dried extracts were derivatised with N,O,-bis-(trimethylsilyl)trifluoroacetamide prior to analysis by GC–MS. Urine specimens from individuals with biotinidase deficiency, maple syrup urine disease (MSUD) and ketotic hypoglycemia were analysed to demonstrate the potential clinical impact.
Results
Recovery of shorter chain hydroxycarboxylic acids decreased significantly when extracts were dried above 25 °C (mean recovery 89 % at 60 °C, p < 0.01) or left under nitrogen post-drying (mean recovery at ambient + 15 min, 40 °C + 15mins and 60 °C + 15mins was 56 %, 12 % and 2 %, respectively, p < 0.01). Whilst dicarboxylic acids/medium chain fatty acids were unaffected by temperature (mean recovery 100 %), prolonged drying reduced recovery (mean recovery 85 % at 60 °C + 15mins, p < 0.01).
Conclusions
Evaporation of solvent extracts with heat and/or prolonged drying under nitrogen results in significant losses of the shorter chain hydroxycarboxylic acids. The evaporation protocol must be carefully controlled to ensure accurate and reproducible results, preventing misdiagnoses and/or misinterpretation of results.
期刊介绍:
The Official Journal of the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC)
Clinica Chimica Acta is a high-quality journal which publishes original Research Communications in the field of clinical chemistry and laboratory medicine, defined as the diagnostic application of chemistry, biochemistry, immunochemistry, biochemical aspects of hematology, toxicology, and molecular biology to the study of human disease in body fluids and cells.
The objective of the journal is to publish novel information leading to a better understanding of biological mechanisms of human diseases, their prevention, diagnosis, and patient management. Reports of an applied clinical character are also welcome. Papers concerned with normal metabolic processes or with constituents of normal cells or body fluids, such as reports of experimental or clinical studies in animals, are only considered when they are clearly and directly relevant to human disease. Evaluation of commercial products have a low priority for publication, unless they are novel or represent a technological breakthrough. Studies dealing with effects of drugs and natural products and studies dealing with the redox status in various diseases are not within the journal''s scope. Development and evaluation of novel analytical methodologies where applicable to diagnostic clinical chemistry and laboratory medicine, including point-of-care testing, and topics on laboratory management and informatics will also be considered. Studies focused on emerging diagnostic technologies and (big) data analysis procedures including digitalization, mobile Health, and artificial Intelligence applied to Laboratory Medicine are also of interest.