Diagnostic accuracy of a non-invasive spot-check hemoglobin meter, Masimo Rad-67® pulse CO-Oximeter®, in detection of anemia in antenatal care settings in Kenya.

IF 2.3 Q2 OBSTETRICS & GYNECOLOGY Frontiers in global women's health Pub Date : 2024-10-14 eCollection Date: 2024-01-01 DOI:10.3389/fgwh.2024.1427261
Angela Koech, Isaac Mwaniki, Joseph Mutunga, Moses Mukhanya, Emily Mwadime, Marvine Ochieng, Grace Mwashigadi, Hiten D Mistry, Rachel Craik, Peter von Dadelszen, Marleen Temmerman, Stanley Luchters, Geoffrey Omuse
{"title":"Diagnostic accuracy of a non-invasive spot-check hemoglobin meter, Masimo Rad-67® pulse CO-Oximeter®, in detection of anemia in antenatal care settings in Kenya.","authors":"Angela Koech, Isaac Mwaniki, Joseph Mutunga, Moses Mukhanya, Emily Mwadime, Marvine Ochieng, Grace Mwashigadi, Hiten D Mistry, Rachel Craik, Peter von Dadelszen, Marleen Temmerman, Stanley Luchters, Geoffrey Omuse","doi":"10.3389/fgwh.2024.1427261","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Point of care hemoglobin meters play key roles in increasing access to anemia screening in antenatal care especially in settings with limited access to laboratories. We aimed to determine the diagnostic accuracy of a non-invasive spot-check hemoglobin (SpHb) meter, Masimo Rad-67® Pulse CO-Oximeter®, in the diagnosis of anemia in pregnant women attending antenatal care clinics in Kilifi, Kenya.</p><p><strong>Methods: </strong>This was a diagnostic accuracy study that retrospectively evaluated SpHb against a validated reference standard of laboratory assessed hemoglobin (Lab Hb) by a SYSMEX XN-330 automated hematology analyzer. The study was nested within a prospective pregnancy cohort study that recruited unselected pregnant women from antenatal care clinics in two public hospitals in Kilifi County, coastal Kenya. Records with both SpHb and Lab Hb were selected from pregnancy visits between May 2021 and December 2022. Linear regression and Bland-Altman analysis were performed to compare the two tests and diagnostic accuracy parameters obtained for the diagnosis of anemia.</p><p><strong>Results: </strong>A total of 2,975 records (from 2,203 unique participants), with paired SpHb and Lab Hb were analyzed. Linear regression showed a significant but weak positive correlation, a proportional bias of 0.44 (95% CI 0.41-0.47) and a constant of 7.59 (95% CI 7.30-7.87, <i>p</i> < 0.001). The median bias was 1.70 g/dl, with limits of agreement of -0.80 to 4.20. SpHb tended to be higher than Lab Hb on the low hemoglobin range but lower than Lab Hb on the high hemoglobin range. The sensitivity of SpHb in detecting anemia was 18.66%. Prevalence, specificity, positive predictive value, and negative predictive values were 46.37%, 96.77%, 83.33%, and 57.92% respectively.</p><p><strong>Conclusion: </strong>Overall, SpHb by Masimo Rad-67® Pulse CO-Oximeter® did not accurately identify pregnant women with anemia and many cases would be missed. We would not recommend its use in antenatal care settings.</p>","PeriodicalId":73087,"journal":{"name":"Frontiers in global women's health","volume":null,"pages":null},"PeriodicalIF":2.3000,"publicationDate":"2024-10-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11513392/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Frontiers in global women's health","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.3389/fgwh.2024.1427261","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/1/1 0:00:00","PubModel":"eCollection","JCR":"Q2","JCRName":"OBSTETRICS & GYNECOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Background: Point of care hemoglobin meters play key roles in increasing access to anemia screening in antenatal care especially in settings with limited access to laboratories. We aimed to determine the diagnostic accuracy of a non-invasive spot-check hemoglobin (SpHb) meter, Masimo Rad-67® Pulse CO-Oximeter®, in the diagnosis of anemia in pregnant women attending antenatal care clinics in Kilifi, Kenya.

Methods: This was a diagnostic accuracy study that retrospectively evaluated SpHb against a validated reference standard of laboratory assessed hemoglobin (Lab Hb) by a SYSMEX XN-330 automated hematology analyzer. The study was nested within a prospective pregnancy cohort study that recruited unselected pregnant women from antenatal care clinics in two public hospitals in Kilifi County, coastal Kenya. Records with both SpHb and Lab Hb were selected from pregnancy visits between May 2021 and December 2022. Linear regression and Bland-Altman analysis were performed to compare the two tests and diagnostic accuracy parameters obtained for the diagnosis of anemia.

Results: A total of 2,975 records (from 2,203 unique participants), with paired SpHb and Lab Hb were analyzed. Linear regression showed a significant but weak positive correlation, a proportional bias of 0.44 (95% CI 0.41-0.47) and a constant of 7.59 (95% CI 7.30-7.87, p < 0.001). The median bias was 1.70 g/dl, with limits of agreement of -0.80 to 4.20. SpHb tended to be higher than Lab Hb on the low hemoglobin range but lower than Lab Hb on the high hemoglobin range. The sensitivity of SpHb in detecting anemia was 18.66%. Prevalence, specificity, positive predictive value, and negative predictive values were 46.37%, 96.77%, 83.33%, and 57.92% respectively.

Conclusion: Overall, SpHb by Masimo Rad-67® Pulse CO-Oximeter® did not accurately identify pregnant women with anemia and many cases would be missed. We would not recommend its use in antenatal care settings.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
肯尼亚产前护理机构在检测贫血时使用无创抽查血红蛋白仪 Masimo Rad-67® pulse CO-Oximeter® 的诊断准确性。
背景:护理点血红蛋白仪在提高产前护理中贫血筛查的普及率方面发挥着关键作用,尤其是在实验室条件有限的情况下。我们的目的是确定 Masimo Rad-67® Pulse CO-Oximeter® 无创点检血红蛋白 (SpHb) 计在诊断肯尼亚基利菲产前护理诊所孕妇贫血方面的诊断准确性:这是一项诊断准确性研究,通过 SYSMEX XN-330 自动血液分析仪,对照实验室评估血红蛋白(Lab Hb)的有效参考标准,对 SpHb 进行了回顾性评估。该研究嵌套在一项前瞻性妊娠队列研究中,该研究从肯尼亚沿海基利菲县两家公立医院的产前检查诊所招募未经选择的孕妇。研究人员从 2021 年 5 月至 2022 年 12 月期间的孕妇就诊记录中选取了 SpHb 和 Lab Hb。通过线性回归和Bland-Altman分析,比较两种检测方法和诊断贫血的准确性参数:共分析了2,975条记录(来自2,203名参与者),其中SpHb和Lab Hb成对。线性回归显示出显著但微弱的正相关性,比例偏差为 0.44(95% CI 0.41-0.47),常数为 7.59(95% CI 7.30-7.87,p 结论:总体而言,MasRobot SpHb 和 Lab Hb 可用于贫血诊断:总体而言,Masimo Rad-67® Pulse CO-Oximeter® SpHb 不能准确识别贫血孕妇,许多病例会被漏诊。我们不建议在产前检查中使用该仪器。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
CiteScore
3.70
自引率
0.00%
发文量
0
审稿时长
13 weeks
期刊最新文献
Aya Contigo: evaluation of a digital intervention to support self-managed medication abortion in Venezuela. Less than one in five teenage women in Ethiopia know about emergency contraception. Determinants of hepatitis B virus infection among pregnant women in Bench Sheko zone, Southwest Ethiopia: a case-control study. Diagnostic accuracy of a non-invasive spot-check hemoglobin meter, Masimo Rad-67® pulse CO-Oximeter®, in detection of anemia in antenatal care settings in Kenya. Estradiol and progesterone from pregnancy to postpartum: a longitudinal latent class analysis.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1