Development and validation of polymer-based porphyrin-incorporated reference materials for calibration of quantitative light-induced fluorescence device

IF 3.1 3区 医学 Q2 ONCOLOGY Photodiagnosis and Photodynamic Therapy Pub Date : 2024-11-01 DOI:10.1016/j.pdpdt.2024.104391
Sang-Eun Park , Hong-Cheol Yoon , Elbert de Josselin de Jong , Kwang-Mahn Kim , Jae-Sung Kwon
{"title":"Development and validation of polymer-based porphyrin-incorporated reference materials for calibration of quantitative light-induced fluorescence device","authors":"Sang-Eun Park ,&nbsp;Hong-Cheol Yoon ,&nbsp;Elbert de Josselin de Jong ,&nbsp;Kwang-Mahn Kim ,&nbsp;Jae-Sung Kwon","doi":"10.1016/j.pdpdt.2024.104391","DOIUrl":null,"url":null,"abstract":"<div><h3>Background</h3><div>This study aimed at developing and validating polymer-based reference materials with varying amounts of porphyrin to accurately assess and calibrate quantitative light-induced fluorescence (QLF) device.</div></div><div><h3>Methods</h3><div>Reference materials with porphyrin concentrations ranging from 0 to 0.08 wt.% were prepared. The surface properties of the materials were analyzed via gloss and roughness measurements. Color analysis of the specimens was performed on black and white backgrounds with or without filters using a spectrophotometer. This approach revealed the correlations between fluorescence and color. The fluorescence emitted by the specimens was analyzed by measuring ΔR and ΔRmax values using a QLF-D Biluminator™.</div></div><div><h3>Results</h3><div>The surface gloss and roughness of the reference materials were not affected by the porphyrin content (<em>p</em> &gt; 0.005). Spectrophotometric measurements revealed significant color differences among most specimen groups depending on the background color and the presence of a filter. QLF-D imaging revealed significant differences in fluorescence (ΔR and ΔRmax) among all specimen groups regardless of the background. The fluorescence values observed on the black backgrounds were higher than those observed on the white backgrounds (<em>p</em> &lt; 0.05).</div></div><div><h3>Conclusions</h3><div>The developed polymer-based porphyrin-incorporated materials serve as reliable reference standards for accurate assessment and calibration of QLF devices. This study demonstrates the importance of background conditions in fluorescence detection and highlights the potential of these materials as standards for QLF device calibration.</div></div>","PeriodicalId":20141,"journal":{"name":"Photodiagnosis and Photodynamic Therapy","volume":"50 ","pages":"Article 104391"},"PeriodicalIF":3.1000,"publicationDate":"2024-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Photodiagnosis and Photodynamic Therapy","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1572100024004289","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"ONCOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Background

This study aimed at developing and validating polymer-based reference materials with varying amounts of porphyrin to accurately assess and calibrate quantitative light-induced fluorescence (QLF) device.

Methods

Reference materials with porphyrin concentrations ranging from 0 to 0.08 wt.% were prepared. The surface properties of the materials were analyzed via gloss and roughness measurements. Color analysis of the specimens was performed on black and white backgrounds with or without filters using a spectrophotometer. This approach revealed the correlations between fluorescence and color. The fluorescence emitted by the specimens was analyzed by measuring ΔR and ΔRmax values using a QLF-D Biluminator™.

Results

The surface gloss and roughness of the reference materials were not affected by the porphyrin content (p > 0.005). Spectrophotometric measurements revealed significant color differences among most specimen groups depending on the background color and the presence of a filter. QLF-D imaging revealed significant differences in fluorescence (ΔR and ΔRmax) among all specimen groups regardless of the background. The fluorescence values observed on the black backgrounds were higher than those observed on the white backgrounds (p < 0.05).

Conclusions

The developed polymer-based porphyrin-incorporated materials serve as reliable reference standards for accurate assessment and calibration of QLF devices. This study demonstrates the importance of background conditions in fluorescence detection and highlights the potential of these materials as standards for QLF device calibration.
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
用于校准光诱导荧光定量装置的聚合物基卟啉掺杂参考材料的开发与验证。
背景:本研究旨在开发和验证不同卟啉含量的聚合物基参考材料,以准确评估和校准定量光诱导荧光(QLF)装置:方法:制备了卟啉浓度为 0 至 0.08 wt.%的参考材料。通过光泽度和粗糙度测量分析了材料的表面特性。使用分光光度计在有滤光片或无滤光片的黑白背景上对试样进行颜色分析。这种方法揭示了荧光和颜色之间的相关性。使用 QLF-D BiluminatorTM 测量 ΔR 和 ΔRmax 值,分析试样发出的荧光:结果:参考材料的表面光泽度和粗糙度不受卟啉含量的影响(p > 0.005)。分光光度测量结果显示,大多数试样组的颜色差异显著,这取决于背景颜色和是否有滤光片。QLF-D 成像显示,无论背景如何,所有标本组的荧光(ΔR 和 ΔRmax)都存在显著差异。黑色背景上观察到的荧光值高于白色背景上观察到的荧光值(p < 0.05):结论:所开发的聚合物基卟啉掺杂材料是准确评估和校准 QLF 设备的可靠参考标准。这项研究证明了背景条件在荧光检测中的重要性,并强调了这些材料作为 QLF 设备校准标准的潜力。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
CiteScore
5.80
自引率
24.20%
发文量
509
审稿时长
50 days
期刊介绍: Photodiagnosis and Photodynamic Therapy is an international journal for the dissemination of scientific knowledge and clinical developments of Photodiagnosis and Photodynamic Therapy in all medical specialties. The journal publishes original articles, review articles, case presentations, "how-to-do-it" articles, Letters to the Editor, short communications and relevant images with short descriptions. All submitted material is subject to a strict peer-review process.
期刊最新文献
The ability of changes in oct angiography values to predict visual field and retinal nerve layer tomography progression in patients with glaucoma progression: A prospective study The effect of antimicrobial photodynamic therapy using different concentrations of phycocyanin against Streptococcus mutans Posterior corneal elevation after small incision lenticule extraction (SMILE) in eyes of different myopia severity Comparative analysis of central corneal thickness in ocular hypertension and healthy individuals Development and validation of polymer-based porphyrin-incorporated reference materials for calibration of quantitative light-induced fluorescence device
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1