Echocardiographic Evaluation of Chronic Aortic Regurgitation: Comparison With Cardiac Magnetic Resonance and Implications for Guideline Recommendations.

IF 12.8 1区 医学 Q1 CARDIAC & CARDIOVASCULAR SYSTEMS JACC. Cardiovascular imaging Pub Date : 2024-10-30 DOI:10.1016/j.jcmg.2024.08.013
Rowa Attar, Maan Malahfji, Clara Angulo, Duc T Nguyen, Edward A Graviss, Dipan J Shah, William A Zoghbi
{"title":"Echocardiographic Evaluation of Chronic Aortic Regurgitation: Comparison With Cardiac Magnetic Resonance and Implications for Guideline Recommendations.","authors":"Rowa Attar, Maan Malahfji, Clara Angulo, Duc T Nguyen, Edward A Graviss, Dipan J Shah, William A Zoghbi","doi":"10.1016/j.jcmg.2024.08.013","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Guidelines for echocardiographic evaluation of aortic regurgitation (AR) have not been validated against an independent quantitative standard.</p><p><strong>Objectives: </strong>The aim of this study was to evaluate the accuracy of the ASE (American Society of Echocardiography) AR guidelines against cardiac magnetic resonance (CMR) and to develop simplified approaches for detection of significant AR.</p><p><strong>Methods: </strong>Patients with AR underwent echocardiography and CMR <4 hours apart. AR severity was graded according to ASE guidelines. Quantitation of regurgitant volume (RegV) was performed with pulsed Doppler at the mitral annulus and right ventricular outflow compared with left ventricular (LV) outflow, and with proximal isovelocity surface area.</p><p><strong>Results: </strong>The authors studied 81 patients; median age was 52 years, and 58% had a bicuspid aortic valve. According to echo, 35 (43%) patients had mild AR, 18 (22%) moderate, 12 (15%) moderate to severe, and 16 (20%) had severe AR. The area under the curve (AUC) for detection of severe AR by CMR using ASE grading was 0.9 (82.4% sensitivity and 96.9% specificity). Feasibility of RegV quantitation was >88% using either echo volumetric method, and it was low for proximal isovelocity surface area (37%). The highest accuracy for echo parameters against CMR was seen with vena contracta width, jet width, and LV end-diastolic volume index (AUC: 0.86-0.89); pressure half-time had the lowest accuracy. Without RegV quantitation, a vena contracta width ≥0.5 cm and indexed LV end-diastolic volume ≥82 mL/m<sup>2</sup> had 95.5% positive predictive value and 87.5% negative predictive value for identifying ≥moderate to severe AR by CMR (AUC: 0.89).</p><p><strong>Conclusions: </strong>The ASE guidelines display very good performance in identifying significant AR. A simplified approach using vena contracta width and LV volumes can be used to reliably identify significant AR. Further validation of the findings in larger cohorts and against clinical outcomes is needed.</p>","PeriodicalId":14767,"journal":{"name":"JACC. Cardiovascular imaging","volume":" ","pages":""},"PeriodicalIF":12.8000,"publicationDate":"2024-10-30","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"JACC. Cardiovascular imaging","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1016/j.jcmg.2024.08.013","RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0

Abstract

Background: Guidelines for echocardiographic evaluation of aortic regurgitation (AR) have not been validated against an independent quantitative standard.

Objectives: The aim of this study was to evaluate the accuracy of the ASE (American Society of Echocardiography) AR guidelines against cardiac magnetic resonance (CMR) and to develop simplified approaches for detection of significant AR.

Methods: Patients with AR underwent echocardiography and CMR <4 hours apart. AR severity was graded according to ASE guidelines. Quantitation of regurgitant volume (RegV) was performed with pulsed Doppler at the mitral annulus and right ventricular outflow compared with left ventricular (LV) outflow, and with proximal isovelocity surface area.

Results: The authors studied 81 patients; median age was 52 years, and 58% had a bicuspid aortic valve. According to echo, 35 (43%) patients had mild AR, 18 (22%) moderate, 12 (15%) moderate to severe, and 16 (20%) had severe AR. The area under the curve (AUC) for detection of severe AR by CMR using ASE grading was 0.9 (82.4% sensitivity and 96.9% specificity). Feasibility of RegV quantitation was >88% using either echo volumetric method, and it was low for proximal isovelocity surface area (37%). The highest accuracy for echo parameters against CMR was seen with vena contracta width, jet width, and LV end-diastolic volume index (AUC: 0.86-0.89); pressure half-time had the lowest accuracy. Without RegV quantitation, a vena contracta width ≥0.5 cm and indexed LV end-diastolic volume ≥82 mL/m2 had 95.5% positive predictive value and 87.5% negative predictive value for identifying ≥moderate to severe AR by CMR (AUC: 0.89).

Conclusions: The ASE guidelines display very good performance in identifying significant AR. A simplified approach using vena contracta width and LV volumes can be used to reliably identify significant AR. Further validation of the findings in larger cohorts and against clinical outcomes is needed.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
慢性主动脉瓣反流的超声心动图评估:与心脏磁共振的比较及对指南建议的影响。
背景:主动脉瓣反流(AR)的超声心动图评估指南尚未经过独立定量标准的验证:主动脉瓣反流(AR)的超声心动图评估指南尚未根据独立的定量标准进行验证:本研究旨在评估美国超声心动图学会(ASE)主动脉瓣反流指南与心脏磁共振(CMR)的准确性,并制定检测明显主动脉瓣反流的简化方法:AR患者接受超声心动图和CMR检查:作者研究了 81 名患者;中位年龄为 52 岁,58% 患有双尖瓣主动脉瓣。根据回波检查结果,35 例(43%)患者为轻度 AR,18 例(22%)为中度,12 例(15%)为中至重度,16 例(20%)为重度 AR。使用 ASE 分级通过 CMR 检测重度 AR 的曲线下面积 (AUC) 为 0.9(灵敏度为 82.4%,特异度为 96.9%)。使用任何一种回波容积法进行 RegV 定量的可行性都大于 88%,而近端等速表面积的可行性较低(37%)。回波参数与 CMR 比较,准确度最高的是静脉收缩宽度、射流宽度和 LV 舒张末期容积指数(AUC:0.86-0.89);压力半衰期的准确度最低。在没有RegV定量的情况下,收缩静脉宽度≥0.5 cm和左心室舒张末期容积指数≥82 mL/m2对通过CMR识别≥中度至重度AR具有95.5%的阳性预测值和87.5%的阴性预测值(AUC:0.89):结论:ASE指南在识别严重AR方面表现非常出色。使用收缩静脉宽度和左心室容积的简化方法可以可靠地鉴别出明显的 AR。需要在更大的队列中并根据临床结果进一步验证这些发现。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
JACC. Cardiovascular imaging
JACC. Cardiovascular imaging CARDIAC & CARDIOVASCULAR SYSTEMS-RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
CiteScore
24.90
自引率
5.70%
发文量
330
审稿时长
4-8 weeks
期刊介绍: JACC: Cardiovascular Imaging, part of the prestigious Journal of the American College of Cardiology (JACC) family, offers readers a comprehensive perspective on all aspects of cardiovascular imaging. This specialist journal covers original clinical research on both non-invasive and invasive imaging techniques, including echocardiography, CT, CMR, nuclear, optical imaging, and cine-angiography. JACC. Cardiovascular imaging highlights advances in basic science and molecular imaging that are expected to significantly impact clinical practice in the next decade. This influence encompasses improvements in diagnostic performance, enhanced understanding of the pathogenetic basis of diseases, and advancements in therapy. In addition to cutting-edge research,the content of JACC: Cardiovascular Imaging emphasizes practical aspects for the practicing cardiologist, including advocacy and practice management.The journal also features state-of-the-art reviews, ensuring a well-rounded and insightful resource for professionals in the field of cardiovascular imaging.
期刊最新文献
Mechanism and Impact of Left Atrial Dyssynchrony on Long-Term Clinical Outcome During Cardiac Resynchronization Therapy. Editorial Board Cardiac Magnetic Resonance Evaluation of LV Remodeling Post-Myocardial Infarction Longitudinal Evaluation of Coronary Arteries and Myocardium in Breast Cancer Using Coronary Computed Tomographic Angiography Myocardial Characteristics, Cardiac Structure, and Cardiac Function in Systemic Light-Chain Amyloidosis
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1