Comparison of intra-assay and inter-assay reproducibility and positive detection times of two different (BacT/Alert 3D and Autobio BC) commercial blood culture systems.
{"title":"Comparison of intra-assay and inter-assay reproducibility and positive detection times of two different (BacT/Alert 3D and Autobio BC) commercial blood culture systems.","authors":"Nilgün Kansak, Nilay Zeynep Kalender, Neslihan Arıcı, Rıza Adaleti, Sebahat Aksaray, Handan Ankaralı, Nevriye Gönüllü","doi":"10.1016/j.ijmmb.2024.100754","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>In our study, we aimed to compare the performance of the BacT/Alert 3D (bioMerieux, France) system, which is currently used in our laboratory, and the Autobio BC (Autobio, China) system, which was newly introduced in our country, using standard and clinical isolates.</p><p><strong>Methods: </strong>Bacterial suspension was prepared by two technicians on the same day and three consecutive days from five different standard strains with 0.5 McFarland turbidity, then serial dilution to a final concentration was adjusted and was simultaneously inoculated in aerobic blood culture bottles. The bacterial concentration was measured by making a quantitative counting plate. The same procedure was also performed for 55 clinical isolates belonging to eleven species. After simulated bacteremia with standard and clinical isolates, the growth results were confirmed by inoculation from positive blood culture bottles onto solid medium and identification was made in the next day with MALDI-TOF MS (bioMérieux). In each study, sterile saline and blood was inoculated into the bottles as a negative control to check contamination. Intra-assay and inter-assay reproducibility of recovery rates and detection times of standard strains; recovery rates and detection times of clinical isolates were compared for both systems.</p><p><strong>Results: </strong>Recovery rates were 100 % in both systems, and when positive detection times were compared, it was found that there was no difference between the two devices in clinical isolates (p:0.262) but that Autobio BC gave significantly (p < 0.001) earlier results in standard strains.</p><p><strong>Conclusions: </strong>In our simulated bloodstream infection study, Autobio BC was found to be comparable with BacT/Alert 3D, both recovery rates and growth detection time performance were found to be very good, and it can be used in routine microbiology laboratories.</p>","PeriodicalId":13284,"journal":{"name":"Indian Journal of Medical Microbiology","volume":" ","pages":"100754"},"PeriodicalIF":1.4000,"publicationDate":"2024-11-21","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Indian Journal of Medical Microbiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1016/j.ijmmb.2024.100754","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"IMMUNOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose: In our study, we aimed to compare the performance of the BacT/Alert 3D (bioMerieux, France) system, which is currently used in our laboratory, and the Autobio BC (Autobio, China) system, which was newly introduced in our country, using standard and clinical isolates.
Methods: Bacterial suspension was prepared by two technicians on the same day and three consecutive days from five different standard strains with 0.5 McFarland turbidity, then serial dilution to a final concentration was adjusted and was simultaneously inoculated in aerobic blood culture bottles. The bacterial concentration was measured by making a quantitative counting plate. The same procedure was also performed for 55 clinical isolates belonging to eleven species. After simulated bacteremia with standard and clinical isolates, the growth results were confirmed by inoculation from positive blood culture bottles onto solid medium and identification was made in the next day with MALDI-TOF MS (bioMérieux). In each study, sterile saline and blood was inoculated into the bottles as a negative control to check contamination. Intra-assay and inter-assay reproducibility of recovery rates and detection times of standard strains; recovery rates and detection times of clinical isolates were compared for both systems.
Results: Recovery rates were 100 % in both systems, and when positive detection times were compared, it was found that there was no difference between the two devices in clinical isolates (p:0.262) but that Autobio BC gave significantly (p < 0.001) earlier results in standard strains.
Conclusions: In our simulated bloodstream infection study, Autobio BC was found to be comparable with BacT/Alert 3D, both recovery rates and growth detection time performance were found to be very good, and it can be used in routine microbiology laboratories.
期刊介绍:
Manuscripts of high standard in the form of original research, multicentric studies, meta analysis, are accepted. Current reports can be submitted as brief communications. Case reports must include review of current literature, clinical details, outcome and follow up. Letters to the editor must be a comment on or pertain to a manuscript already published in the IJMM or in relation to preliminary communication of a larger study.