Low-dose lung CT: Optimizing diagnostic radiation dose – A phantom study

IF 1.8 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING European Journal of Radiology Open Pub Date : 2024-11-24 DOI:10.1016/j.ejro.2024.100614
Michael Zellner , Sebastian Tschauner , Mathias S. Weyland , Peter Eggenberger Hotz , Stephan Scheidegger , Christian J. Kellenberger
{"title":"Low-dose lung CT: Optimizing diagnostic radiation dose – A phantom study","authors":"Michael Zellner ,&nbsp;Sebastian Tschauner ,&nbsp;Mathias S. Weyland ,&nbsp;Peter Eggenberger Hotz ,&nbsp;Stephan Scheidegger ,&nbsp;Christian J. Kellenberger","doi":"10.1016/j.ejro.2024.100614","DOIUrl":null,"url":null,"abstract":"<div><h3>Background/purpose</h3><div>To investigate a quantitative method for assessing image quality of low dose lung computed tomography (CT) and find the lowest exposure dose providing diagnostic images.</div></div><div><h3>Methods</h3><div>Axial volumetric lung CT acquisitions (256 slice scanner) were performed on three different sized anthropomorphic phantoms at different dose levels. The maximum steepness of sigmoid curves fitted to line density profiles was measured at lung-to-pleura interfaces. For each phantom, image sharpness was calculated as the median of 468 measurements from 39 different locations. Diagnostic image quality for the adult and paediatric phantom was rated by three radiologists using 4-point Likert scales. The image sharpness cut-off for obtaining adequate image quality was determined from qualitative ratings.</div></div><div><h3>Results</h3><div>Adequate diagnostic image quality was reached at a median steepness of 713 HU/mm in the adult phantom with a corresponding CTDIvol of 0.14 mGy and an effective dose of 0.13 mSv at a dose level of 100 kVp and 10 mA. In the paediatric phantom diagnostic image quality was reached at a median steepness of 1139 HU/mm with a corresponding CTDIvol of 0.13 mGy and an effective dose of 0.08 mSv at a dose level of 100 kVp and 10 mA.</div></div><div><h3>Conclusions</h3><div>Determination of image sharpness on line density profiles can be used as quantitative measure for image quality of lung CT. Sufficient-quality lung CT can be achieved at effective radiation doses of 0.13 mSv (adult phantom) and 0.08 mSv (paediatric phantom). These findings suggest that substantial dose reduction is feasible without compromising diagnostic accuracy.</div></div>","PeriodicalId":38076,"journal":{"name":"European Journal of Radiology Open","volume":"13 ","pages":"Article 100614"},"PeriodicalIF":1.8000,"publicationDate":"2024-11-24","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"European Journal of Radiology Open","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2352047724000698","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Background/purpose

To investigate a quantitative method for assessing image quality of low dose lung computed tomography (CT) and find the lowest exposure dose providing diagnostic images.

Methods

Axial volumetric lung CT acquisitions (256 slice scanner) were performed on three different sized anthropomorphic phantoms at different dose levels. The maximum steepness of sigmoid curves fitted to line density profiles was measured at lung-to-pleura interfaces. For each phantom, image sharpness was calculated as the median of 468 measurements from 39 different locations. Diagnostic image quality for the adult and paediatric phantom was rated by three radiologists using 4-point Likert scales. The image sharpness cut-off for obtaining adequate image quality was determined from qualitative ratings.

Results

Adequate diagnostic image quality was reached at a median steepness of 713 HU/mm in the adult phantom with a corresponding CTDIvol of 0.14 mGy and an effective dose of 0.13 mSv at a dose level of 100 kVp and 10 mA. In the paediatric phantom diagnostic image quality was reached at a median steepness of 1139 HU/mm with a corresponding CTDIvol of 0.13 mGy and an effective dose of 0.08 mSv at a dose level of 100 kVp and 10 mA.

Conclusions

Determination of image sharpness on line density profiles can be used as quantitative measure for image quality of lung CT. Sufficient-quality lung CT can be achieved at effective radiation doses of 0.13 mSv (adult phantom) and 0.08 mSv (paediatric phantom). These findings suggest that substantial dose reduction is feasible without compromising diagnostic accuracy.
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
低剂量肺部 CT:优化诊断辐射剂量 - 一项模型研究
背景/目的 研究一种评估低剂量肺部计算机断层扫描(CT)图像质量的定量方法,并找出能提供诊断图像的最低曝光剂量。方法 在三个不同大小的拟人化模型上以不同剂量水平进行轴向容积肺部 CT 采集(256 片扫描仪)。在肺-胸膜界面测量了线密度剖面拟合的sigmoid曲线的最大陡度。每个模型的图像清晰度是根据 39 个不同位置 468 次测量结果的中位数计算得出的。成人和儿童模型的诊断图像质量由三位放射科医生使用 4 点李克特量表进行评分。结果成人模型的中位陡度为 713 HU/mm,相应的 CTDIvol 为 0.14 mGy,有效剂量为 0.13 mSv(剂量水平为 100 kVp 和 10 mA)时,诊断图像质量达到合格。在儿童模型中,诊断图像质量的中位陡度为 1139 HU/mm,相应的 CTDIvol 为 0.13 mGy,有效剂量为 0.08 mSv,剂量水平为 100 kVp 和 10 mA。在有效辐射剂量为 0.13 毫西弗特(成人模型)和 0.08 毫西弗特(儿童模型)的情况下,可以获得足够质量的肺部 CT。这些研究结果表明,在不影响诊断准确性的前提下大幅降低剂量是可行的。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
European Journal of Radiology Open
European Journal of Radiology Open Medicine-Radiology, Nuclear Medicine and Imaging
CiteScore
4.10
自引率
5.00%
发文量
55
审稿时长
51 days
期刊最新文献
An online national quality assessment survey of prostate MRI reading: interreader variability in prostate volume measurement and PI-RADS classification. Deep learning enabled near-isotropic CAIPIRINHA VIBE in the nephrogenic phase improves image quality and renal lesion conspicuity. Multidisciplinary quantitative and qualitative assessment of IDH-mutant gliomas with full diagnostic deep learning image reconstruction. Role of pre-procedure CCTA in predicting failed percutaneous coronary intervention for chronic total occlusions Enhancing mortality prediction in patients with spontaneous intracerebral hemorrhage: Radiomics and supervised machine learning on non-contrast computed tomography
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1