Comparison of haemoglobin concentration measurements using HemoCue-301 and Sysmex XN-Series 1500: A survey among anaemic Gambian infants aged 6-12 months.

IF 2.9 3区 综合性期刊 Q1 MULTIDISCIPLINARY SCIENCES PLoS ONE Pub Date : 2024-11-25 eCollection Date: 2024-01-01 DOI:10.1371/journal.pone.0310577
Mamadou Bah, Hans Verhoef, Abdou Camara, Morris Nden Ngom, Demba Jallow, Kebba Bajo, Foday Bah, Maarten Pleij, Maaike Klappe, Alasana Saidykhan, Emmanuel Okoh, Abdoulie Bah, Carla Cerami
{"title":"Comparison of haemoglobin concentration measurements using HemoCue-301 and Sysmex XN-Series 1500: A survey among anaemic Gambian infants aged 6-12 months.","authors":"Mamadou Bah, Hans Verhoef, Abdou Camara, Morris Nden Ngom, Demba Jallow, Kebba Bajo, Foday Bah, Maarten Pleij, Maaike Klappe, Alasana Saidykhan, Emmanuel Okoh, Abdoulie Bah, Carla Cerami","doi":"10.1371/journal.pone.0310577","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>In low-income countries, point-of-care photometers are used in the screening and management of anaemia in individuals, but also in the assessment of population iron status when evaluating efficacy of intervention studies or public health interventions.</p><p><strong>Aims: </strong>We aimed to evaluate the accuracy of a commonly used photometer, HemoCue-301, in determining haemoglobin concentration among anaemic children aged 6-12 months in a field setting in rural Africa.</p><p><strong>Methods: </strong>This report concerns a secondary analysis of data from Gambian infants being screened for an ongoing randomized controlled trial. In those found to be anaemic by HemoCue-301, haemoglobin concentration was measured by Sysmex XN-1500, an automated haematology analyser that was used as a reference. Passing-Bablok regression analysis was used to estimate the regression constant (systematic deviation between two measurement methods that remain consistent across the range of measurements) and proportional bias (systematic deviation between two measurement methods that change in magnitude relative to the value being measured).</p><p><strong>Results: </strong>Analysis was based on 227 participants. There was strong evidence of absolute bias among moderately anaemic participants (haemoglobin concentration at 8.0g/dL) (absolute bias: 1.12g/dL; 95% CI: 0.91 to 1.37g/dL; proportional bias: 14.0%; 95% CI: 11.4% to 17.1%) in haemoglobin concentrations measured by HemoCue-301 compared to those measured by Sysmex XN-Series1500. Bias was marginal at haemoglobin concentration of 11.0g/dL (absolute bias: -0.08g/dL; 95% CI: -0.18 to 0.07g/dL; proportional bias: -7.3%; 95% CI: -6.5% to 0.6%).</p><p><strong>Conclusion: </strong>Haemoglobin measurements by HemoCue-301 seem substantially biased in participants with haemoglobin less than 8.0g/dL.</p>","PeriodicalId":20189,"journal":{"name":"PLoS ONE","volume":"19 11","pages":"e0310577"},"PeriodicalIF":2.9000,"publicationDate":"2024-11-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"PLoS ONE","FirstCategoryId":"103","ListUrlMain":"https://doi.org/10.1371/journal.pone.0310577","RegionNum":3,"RegionCategory":"综合性期刊","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/1/1 0:00:00","PubModel":"eCollection","JCR":"Q1","JCRName":"MULTIDISCIPLINARY SCIENCES","Score":null,"Total":0}
引用次数: 0

Abstract

Background: In low-income countries, point-of-care photometers are used in the screening and management of anaemia in individuals, but also in the assessment of population iron status when evaluating efficacy of intervention studies or public health interventions.

Aims: We aimed to evaluate the accuracy of a commonly used photometer, HemoCue-301, in determining haemoglobin concentration among anaemic children aged 6-12 months in a field setting in rural Africa.

Methods: This report concerns a secondary analysis of data from Gambian infants being screened for an ongoing randomized controlled trial. In those found to be anaemic by HemoCue-301, haemoglobin concentration was measured by Sysmex XN-1500, an automated haematology analyser that was used as a reference. Passing-Bablok regression analysis was used to estimate the regression constant (systematic deviation between two measurement methods that remain consistent across the range of measurements) and proportional bias (systematic deviation between two measurement methods that change in magnitude relative to the value being measured).

Results: Analysis was based on 227 participants. There was strong evidence of absolute bias among moderately anaemic participants (haemoglobin concentration at 8.0g/dL) (absolute bias: 1.12g/dL; 95% CI: 0.91 to 1.37g/dL; proportional bias: 14.0%; 95% CI: 11.4% to 17.1%) in haemoglobin concentrations measured by HemoCue-301 compared to those measured by Sysmex XN-Series1500. Bias was marginal at haemoglobin concentration of 11.0g/dL (absolute bias: -0.08g/dL; 95% CI: -0.18 to 0.07g/dL; proportional bias: -7.3%; 95% CI: -6.5% to 0.6%).

Conclusion: Haemoglobin measurements by HemoCue-301 seem substantially biased in participants with haemoglobin less than 8.0g/dL.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
背景:在低收入国家,护理点光度计不仅用于筛查和管理贫血患儿,还用于评估干预研究或公共卫生干预措施的效果时对人群铁状况的评估:本报告是对正在进行的随机对照试验所筛查的冈比亚婴儿数据的二次分析。在通过 HemoCue-301 发现贫血的婴儿中,使用 Sysmex XN-1500 自动血液分析仪测量血红蛋白浓度,并将其作为参考。采用 Passing-Bablok 回归分析法估算回归常数(两种测量方法之间的系统偏差,在整个测量范围内保持一致)和比例偏差(两种测量方法之间的系统偏差,相对于测量值的变化幅度):分析基于 227 名参与者。与 Sysmex XN-Series1500 测量的血红蛋白浓度相比,HemoCue-301 测量的血红蛋白浓度在中度贫血者(血红蛋白浓度为 8.0g/dL)中存在明显的绝对偏差(绝对偏差:1.12g/dL;95% CI:0.91 至 1.37g/dL;比例偏差:14.0%;95% CI:11.4% 至 17.1%)。当血红蛋白浓度为 11.0g/dL 时,偏差很小(绝对偏差:-0.08g/dL;95% CI:-0.18 至 0.07g/dL;比例偏差:-7.3%;95% CI:-6.5% 至 0.6%):结论:HemoCue-301 对血红蛋白低于 8.0g/dL 的参与者进行的血红蛋白测量似乎存在严重偏差。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
PLoS ONE
PLoS ONE 生物-生物学
CiteScore
6.20
自引率
5.40%
发文量
14242
审稿时长
3.7 months
期刊介绍: PLOS ONE is an international, peer-reviewed, open-access, online publication. PLOS ONE welcomes reports on primary research from any scientific discipline. It provides: * Open-access—freely accessible online, authors retain copyright * Fast publication times * Peer review by expert, practicing researchers * Post-publication tools to indicate quality and impact * Community-based dialogue on articles * Worldwide media coverage
期刊最新文献
Chemical fractions of potassium in arid region calcareous soils: The impact of microclimates and physiographic variability. Is transcutaneous auricular vagus nerve stimulation effective and safe for primary insomnia? A PRISMA-compliant protocol for a systematic review and meta-analysis. Comparison of haemoglobin concentration measurements using HemoCue-301 and Sysmex XN-Series 1500: A survey among anaemic Gambian infants aged 6-12 months. Clinical predictors of physiologic change after treatment with immunosuppression in hypersensitivity pneumonitis. Juvenile detachment, an early sign of departure from parental care, in the leech Orientobdelloides siamensis (Oka, 1917).
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1