Abdellah Khallouqi, Hamza Sekkat, Omar El Rhazouani, Abdellah Halimi
{"title":"Investigation of organs dosimetry precision using ATOM phantom and optically stimulated luminescence detectors in computed tomography.","authors":"Abdellah Khallouqi, Hamza Sekkat, Omar El Rhazouani, Abdellah Halimi","doi":"10.1088/2057-1976/ad992e","DOIUrl":null,"url":null,"abstract":"<p><p>The primary objective of this study was to compare organ doses measured using optically stimulated luminescent dosimeters (OSLDs) with those estimated by the CT-EXPO software for common CT protocols. An anthropomorphic ATOM phantom was employed to measure organ doses across head, chest, and abdominal CT scans performed on a Hitachi Supria 16-slice CT scanner. These OSLD measurements were then compared to the estimates provided by the widely used CT-EXPO software. Organ doses were assessed using OSLDs placed in an adult anthropomorphic phantom, with calibration performed through a comprehensive process involving multiple tube potentials and sensitivity corrections. Results from three CT acquisitions per protocol were compared to estimates provided by CT-EXPO software. Findings reveal significant discrepancies between measured and estimated organ doses, with p-values consistently below 0.05 across all organs. For head CT, measured eye lens doses averaged 33.51 mGy, 6.0% lower than the estimated 35.65 mGy. In chest CT, the thyroid dose was 9.82 mGy, 13.5% higher than the estimated 8.65 mGy. For abdominal CT, the liver dose measured 12.11 mGy, 9.6% higher than the estimated 11.05 mGy. Measured doses for the rest of organs were generally lower than those predicted by CT-EXPO, showing some limitations in current estimation models and the importance of precise dosimetry. This study highlights the potential of OSLD measurements as a complementary method for organ dose assessment in CT imaging, emphasizing the need for more accurate organ dose measurement to optimize patient care.</p>","PeriodicalId":8896,"journal":{"name":"Biomedical Physics & Engineering Express","volume":" ","pages":""},"PeriodicalIF":1.3000,"publicationDate":"2024-12-20","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Biomedical Physics & Engineering Express","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1088/2057-1976/ad992e","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
The primary objective of this study was to compare organ doses measured using optically stimulated luminescent dosimeters (OSLDs) with those estimated by the CT-EXPO software for common CT protocols. An anthropomorphic ATOM phantom was employed to measure organ doses across head, chest, and abdominal CT scans performed on a Hitachi Supria 16-slice CT scanner. These OSLD measurements were then compared to the estimates provided by the widely used CT-EXPO software. Organ doses were assessed using OSLDs placed in an adult anthropomorphic phantom, with calibration performed through a comprehensive process involving multiple tube potentials and sensitivity corrections. Results from three CT acquisitions per protocol were compared to estimates provided by CT-EXPO software. Findings reveal significant discrepancies between measured and estimated organ doses, with p-values consistently below 0.05 across all organs. For head CT, measured eye lens doses averaged 33.51 mGy, 6.0% lower than the estimated 35.65 mGy. In chest CT, the thyroid dose was 9.82 mGy, 13.5% higher than the estimated 8.65 mGy. For abdominal CT, the liver dose measured 12.11 mGy, 9.6% higher than the estimated 11.05 mGy. Measured doses for the rest of organs were generally lower than those predicted by CT-EXPO, showing some limitations in current estimation models and the importance of precise dosimetry. This study highlights the potential of OSLD measurements as a complementary method for organ dose assessment in CT imaging, emphasizing the need for more accurate organ dose measurement to optimize patient care.
期刊介绍:
BPEX is an inclusive, international, multidisciplinary journal devoted to publishing new research on any application of physics and/or engineering in medicine and/or biology. Characterized by a broad geographical coverage and a fast-track peer-review process, relevant topics include all aspects of biophysics, medical physics and biomedical engineering. Papers that are almost entirely clinical or biological in their focus are not suitable. The journal has an emphasis on publishing interdisciplinary work and bringing research fields together, encompassing experimental, theoretical and computational work.