How many knots are necessary to achieve knot security of two high strength suture tapes? A biomechanical comparative analysis

IF 2 3区 医学 Q2 ORTHOPEDICS Archives of Orthopaedic and Trauma Surgery Pub Date : 2024-12-16 DOI:10.1007/s00402-024-05638-2
Tatjana Pastor, Ivan Zderic, Kenneth P. van Knegsel, Till Berk, Rayna Mechkarska, Frank J. P. Beeres, Boyko Gueorguiev, Torsten Pastor
{"title":"How many knots are necessary to achieve knot security of two high strength suture tapes? A biomechanical comparative analysis","authors":"Tatjana Pastor,&nbsp;Ivan Zderic,&nbsp;Kenneth P. van Knegsel,&nbsp;Till Berk,&nbsp;Rayna Mechkarska,&nbsp;Frank J. P. Beeres,&nbsp;Boyko Gueorguiev,&nbsp;Torsten Pastor","doi":"10.1007/s00402-024-05638-2","DOIUrl":null,"url":null,"abstract":"<div><h3>Introduction</h3><p>According to current clinical practice, a minimum of 7 knots are required to provide secure hold in high-strength sutures. A new technology featuring a suture tape with a salt-infused silicon core has been recently developed, potentially reducing the number of needed knots. Aims: to (1) assess the influence of number of knots on tape security, (2) evaluate the effect of different ambient conditions on knot security, and (3) compare the biomechanical competence of the novel versus a conventional suture tape.</p><h3>Materials and methods</h3><p>A conventional suture tape (ST, SutureTape<sup>tm</sup>) was considered for knot tying together with the novel suture type (DT, Dynatape<sup>tm</sup>). Specimens were assigned to receive different number of knots, ranging from 3 to 7, and to be exposed to different media during tying―air (dry), saline solution (wet), and fat (fatty-wet). Seven specimens were considered for each suture type, knot number and ambient condition. With knotted sutures mounted between two roller bearings, quasi-static tensile ramp tests were performed to evaluate knot slippage, ultimate force at rupture, and minimum number of knots preventing suture unraveling for each suture tape and condition.</p><h3>Results</h3><p>Whereas the ST ruptured without unraveling with a minimum of 6 knots in all specimens and ambient conditions, the minimum number of knots for a DT rupture without unraveling was 6 in dry, 4 in wet, and 5 in fatty-wet condition. Ultimate force at rupture with a minimum number of needed knots did not differ significantly between ST and DT (p ≥ 0.067), in contrast to knot slippage that was significantly bigger for ST versus DT in wet and fatty-wet conditions(p ≤ 0.001).</p><h3>Conclusions</h3><p>In fatty-wet conditions―related to open surgery―the novel Dynatape<sup>tm</sup> suture tape requires 5 instead of 7 knots to achieve their security. In wet conditions―related to arthroscopic surgery―this number can be reduced to 4 knots. In contrast, the conventional SutureTape<sup>tm</sup> needs 6 knots to provide security in all conditions.</p></div>","PeriodicalId":8326,"journal":{"name":"Archives of Orthopaedic and Trauma Surgery","volume":"145 1","pages":""},"PeriodicalIF":2.0000,"publicationDate":"2024-12-16","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Archives of Orthopaedic and Trauma Surgery","FirstCategoryId":"3","ListUrlMain":"https://link.springer.com/article/10.1007/s00402-024-05638-2","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"ORTHOPEDICS","Score":null,"Total":0}
引用次数: 0

Abstract

Introduction

According to current clinical practice, a minimum of 7 knots are required to provide secure hold in high-strength sutures. A new technology featuring a suture tape with a salt-infused silicon core has been recently developed, potentially reducing the number of needed knots. Aims: to (1) assess the influence of number of knots on tape security, (2) evaluate the effect of different ambient conditions on knot security, and (3) compare the biomechanical competence of the novel versus a conventional suture tape.

Materials and methods

A conventional suture tape (ST, SutureTapetm) was considered for knot tying together with the novel suture type (DT, Dynatapetm). Specimens were assigned to receive different number of knots, ranging from 3 to 7, and to be exposed to different media during tying―air (dry), saline solution (wet), and fat (fatty-wet). Seven specimens were considered for each suture type, knot number and ambient condition. With knotted sutures mounted between two roller bearings, quasi-static tensile ramp tests were performed to evaluate knot slippage, ultimate force at rupture, and minimum number of knots preventing suture unraveling for each suture tape and condition.

Results

Whereas the ST ruptured without unraveling with a minimum of 6 knots in all specimens and ambient conditions, the minimum number of knots for a DT rupture without unraveling was 6 in dry, 4 in wet, and 5 in fatty-wet condition. Ultimate force at rupture with a minimum number of needed knots did not differ significantly between ST and DT (p ≥ 0.067), in contrast to knot slippage that was significantly bigger for ST versus DT in wet and fatty-wet conditions(p ≤ 0.001).

Conclusions

In fatty-wet conditions―related to open surgery―the novel Dynatapetm suture tape requires 5 instead of 7 knots to achieve their security. In wet conditions―related to arthroscopic surgery―this number can be reduced to 4 knots. In contrast, the conventional SutureTapetm needs 6 knots to provide security in all conditions.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
两种高强度缝合带需要打多少个结才能保证打结安全?生物力学比较分析
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
CiteScore
4.30
自引率
13.00%
发文量
424
审稿时长
2 months
期刊介绍: "Archives of Orthopaedic and Trauma Surgery" is a rich source of instruction and information for physicians in clinical practice and research in the extensive field of orthopaedics and traumatology. The journal publishes papers that deal with diseases and injuries of the musculoskeletal system from all fields and aspects of medicine. The journal is particularly interested in papers that satisfy the information needs of orthopaedic clinicians and practitioners. The journal places special emphasis on clinical relevance. "Archives of Orthopaedic and Trauma Surgery" is the official journal of the German Speaking Arthroscopy Association (AGA).
期刊最新文献
Bipolar vs. monopolar sealer in decreasing blood loss and transfusion rate in patients undergoing two-stage exchange in infected total knee arthroplasty: propensity score-matched study. Clinical and magnetic resonance imaging outcome after proximal hamstring tendon repair at mean 3 years follow-up. Scapho-metacarpal dual mobility prosthesis for TMC-1 joint salvage: technical insights. The single antegrade sling graft: a novel hamstring autograft technique for combined anterior cruciate ligament and anterolateral ligament reconstruction. Total joint arthroplasty of the thumb CMC joint.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1