Femke Wouters, Henri Gruwez, Christophe Smeets, Anessa Pijalovic, Wouter Wilms, Julie Vranken, Zoë Pieters, Hugo Van Herendael, Dieter Nuyens, Maximo Rivero-Ayerza, Pieter Vandervoort, Peter Haemers, Laurent Pison
{"title":"Comparative Evaluation of Consumer Wearable Devices for Atrial Fibrillation Detection: Validation Study.","authors":"Femke Wouters, Henri Gruwez, Christophe Smeets, Anessa Pijalovic, Wouter Wilms, Julie Vranken, Zoë Pieters, Hugo Van Herendael, Dieter Nuyens, Maximo Rivero-Ayerza, Pieter Vandervoort, Peter Haemers, Laurent Pison","doi":"10.2196/65139","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Consumer-oriented wearable devices (CWDs) such as smartphones and smartwatches have gained prominence for their ability to detect atrial fibrillation (AF) through proprietary algorithms using electrocardiography or photoplethysmography (PPG)-based digital recordings. Despite numerous individual validation studies, a direct comparison of interdevice performance is lacking.</p><p><strong>Objective: </strong>This study aimed to evaluate and compare the ability of CWDs to distinguish between sinus rhythm and AF.</p><p><strong>Methods: </strong>Patients exhibiting sinus rhythm or AF were enrolled through a cardiology outpatient clinic. The participants were instructed to perform heart rhythm measurements using a handheld 6-lead electrocardiogram (ECG) device (KardiaMobile 6L), a smartwatch-derived single-lead ECG (Apple Watch), and two PPG-based smartphone apps (FibriCheck and Preventicus) in a random sequence, with simultaneous 12-lead reference ECG as the gold standard.</p><p><strong>Results: </strong>A total of 122 participants were included in the study: median age 69 (IQR 61-77) years, 63.9% (n=78) men, 25% (n=30) with AF, 9.8% (n=12) without prior smartphone experience, and 73% (n=89) without experience in using a smartwatch. The sensitivity to detect AF was 100% for all devices. The specificity to detect sinus rhythm was 96.4% (95% CI 89.5%-98.8%) for KardiaMobile 6L, 97.8% (95% CI 91.6%-99.5%) for Apple Watch, 98.9% (95% CI 92.5%-99.8%) for FibriCheck, and 97.8% (95% CI 91.5%-99.4%) for Preventicus (P=.50). Insufficient quality measurements were observed in 10.7% (95% CI 6.3%-17.5%) of cases for both KardiaMobile 6L and Apple Watch, 7.4% (95% CI 3.9%-13.6%) for FibriCheck, and 14.8% (95% CI 9.5%-22.2%) for Preventicus (P=.21). Participants preferred Apple Watch over the other devices to monitor their heart rhythm.</p><p><strong>Conclusions: </strong>In this study population, the discrimination between sinus rhythm and AF using CWDs based on ECG or PPG was highly accurate, with no significant variations in performance across the examined devices.</p>","PeriodicalId":14841,"journal":{"name":"JMIR Formative Research","volume":"9 ","pages":"e65139"},"PeriodicalIF":2.0000,"publicationDate":"2025-01-09","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11737281/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"JMIR Formative Research","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.2196/65139","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"HEALTH CARE SCIENCES & SERVICES","Score":null,"Total":0}
引用次数: 0
Abstract
Background: Consumer-oriented wearable devices (CWDs) such as smartphones and smartwatches have gained prominence for their ability to detect atrial fibrillation (AF) through proprietary algorithms using electrocardiography or photoplethysmography (PPG)-based digital recordings. Despite numerous individual validation studies, a direct comparison of interdevice performance is lacking.
Objective: This study aimed to evaluate and compare the ability of CWDs to distinguish between sinus rhythm and AF.
Methods: Patients exhibiting sinus rhythm or AF were enrolled through a cardiology outpatient clinic. The participants were instructed to perform heart rhythm measurements using a handheld 6-lead electrocardiogram (ECG) device (KardiaMobile 6L), a smartwatch-derived single-lead ECG (Apple Watch), and two PPG-based smartphone apps (FibriCheck and Preventicus) in a random sequence, with simultaneous 12-lead reference ECG as the gold standard.
Results: A total of 122 participants were included in the study: median age 69 (IQR 61-77) years, 63.9% (n=78) men, 25% (n=30) with AF, 9.8% (n=12) without prior smartphone experience, and 73% (n=89) without experience in using a smartwatch. The sensitivity to detect AF was 100% for all devices. The specificity to detect sinus rhythm was 96.4% (95% CI 89.5%-98.8%) for KardiaMobile 6L, 97.8% (95% CI 91.6%-99.5%) for Apple Watch, 98.9% (95% CI 92.5%-99.8%) for FibriCheck, and 97.8% (95% CI 91.5%-99.4%) for Preventicus (P=.50). Insufficient quality measurements were observed in 10.7% (95% CI 6.3%-17.5%) of cases for both KardiaMobile 6L and Apple Watch, 7.4% (95% CI 3.9%-13.6%) for FibriCheck, and 14.8% (95% CI 9.5%-22.2%) for Preventicus (P=.21). Participants preferred Apple Watch over the other devices to monitor their heart rhythm.
Conclusions: In this study population, the discrimination between sinus rhythm and AF using CWDs based on ECG or PPG was highly accurate, with no significant variations in performance across the examined devices.