Audiovisual Breathing Guidance for Improved Image Quality and Scan Efficiency of T2- and Diffusion-Weighted Liver MRI.

IF 7 1区 医学 Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Investigative Radiology Pub Date : 2025-01-14 DOI:10.1097/RLI.0000000000001150
Narine Mesropyan, Christoph Katemann, Annerieke Heuvelink-Marck, Can Yüksel, Alexander Isaak, Asadeh Lakghomi, Leon Bischoff, Tatjana Dell, Dmitrij Kravchenko, Daniel Kuetting, Claus C Pieper, Julian A Luetkens
{"title":"Audiovisual Breathing Guidance for Improved Image Quality and Scan Efficiency of T2- and Diffusion-Weighted Liver MRI.","authors":"Narine Mesropyan, Christoph Katemann, Annerieke Heuvelink-Marck, Can Yüksel, Alexander Isaak, Asadeh Lakghomi, Leon Bischoff, Tatjana Dell, Dmitrij Kravchenko, Daniel Kuetting, Claus C Pieper, Julian A Luetkens","doi":"10.1097/RLI.0000000000001150","DOIUrl":null,"url":null,"abstract":"<p><strong>Objectives: </strong>Impaired image quality and long scan times frequently occur in respiratory-triggered sequences in liver magnetic resonance imaging (MRI). We evaluated the impact of an in-bore active breathing guidance (BG) application on image quality and scan time of respiratory-triggered T2-weighted (T2) and diffusion-weighted imaging (DWI) by comparing sequences with standard triggering (T2S and DWIS) and with BG (T2BG and DWIBG).</p><p><strong>Materials and methods: </strong>In this prospective study, random patients with clinical indications for liver MRI underwent 3 T MRI with standard and BG acquisitions. The audiovisual BG application received the respiratory signal from the scanner, and animated breathing instructions were displayed using a mirror and screen behind the MRI bore. Prior to the DWIBG and T2BG acquisition, patients received a short video instruction about MRI with BG. Suitable parameters for desired breathing pattern for T2BG and DWIBG were set individually for each patient based on the patient's physical respiratory ability (ie, 4 seconds breathing followed by 4.5 seconds breath holding). Artifacts, sharpness, lesion conspicuity, and overall image quality were assessed using a Likert scale from 1 (nondiagnostic) to 5 (excellent). Scan time, apparent contrast-to-noise ratio, and apparent signal-to-noise ratio (aSNR) for all sequences were analyzed. Paired t test and Wilcoxon test were used for statistical analysis.</p><p><strong>Results: </strong>Thirty-two patients (mean age: 55 ± 13 years, 13 female) were included. T2BG showed less artifacts (4.5 ± 0.7 vs 4.1 ± 0.8, P < 0.001) and better sharpness, lesion conspicuity, and overall image quality (eg, overall image quality 4.6 ± 0.7 vs 4.4 ± 0.7, P = 0.004) compared with T2S. DWIBG demonstrated improved image quality in all categories compared with DWIS (eg, overall image quality 4.5 ± 0.5 vs 4.3 ± 0.5, P = 0.005) and less artifacts (4.1 ± 0.5 vs 3.8 ± 0.7, P = 0.007). Scan times of T2BG (286 ± 23 vs 345 ± 68 seconds, P < 0.001) and DWIBG (160 ± 4 vs 252 ± 70 seconds, P < 0.001) were reduced by 17% and 37%, respectively. aSNR and apparent contrast-to-noise ratio (eg, aSNR: 23.45 ± 11.31 [T2BG] vs 25.84 ± 10.76 [T2S]; P = 0.079) were similar for both sequences for both approaches.</p><p><strong>Conclusions: </strong>Active BG for respiratory-triggered liver T2w and DWI sequences led to significant reduction of breathing artifacts, improved image quality, and shorter scan time compared with standard acquisitions.</p>","PeriodicalId":14486,"journal":{"name":"Investigative Radiology","volume":" ","pages":""},"PeriodicalIF":7.0000,"publicationDate":"2025-01-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Investigative Radiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/RLI.0000000000001150","RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Objectives: Impaired image quality and long scan times frequently occur in respiratory-triggered sequences in liver magnetic resonance imaging (MRI). We evaluated the impact of an in-bore active breathing guidance (BG) application on image quality and scan time of respiratory-triggered T2-weighted (T2) and diffusion-weighted imaging (DWI) by comparing sequences with standard triggering (T2S and DWIS) and with BG (T2BG and DWIBG).

Materials and methods: In this prospective study, random patients with clinical indications for liver MRI underwent 3 T MRI with standard and BG acquisitions. The audiovisual BG application received the respiratory signal from the scanner, and animated breathing instructions were displayed using a mirror and screen behind the MRI bore. Prior to the DWIBG and T2BG acquisition, patients received a short video instruction about MRI with BG. Suitable parameters for desired breathing pattern for T2BG and DWIBG were set individually for each patient based on the patient's physical respiratory ability (ie, 4 seconds breathing followed by 4.5 seconds breath holding). Artifacts, sharpness, lesion conspicuity, and overall image quality were assessed using a Likert scale from 1 (nondiagnostic) to 5 (excellent). Scan time, apparent contrast-to-noise ratio, and apparent signal-to-noise ratio (aSNR) for all sequences were analyzed. Paired t test and Wilcoxon test were used for statistical analysis.

Results: Thirty-two patients (mean age: 55 ± 13 years, 13 female) were included. T2BG showed less artifacts (4.5 ± 0.7 vs 4.1 ± 0.8, P < 0.001) and better sharpness, lesion conspicuity, and overall image quality (eg, overall image quality 4.6 ± 0.7 vs 4.4 ± 0.7, P = 0.004) compared with T2S. DWIBG demonstrated improved image quality in all categories compared with DWIS (eg, overall image quality 4.5 ± 0.5 vs 4.3 ± 0.5, P = 0.005) and less artifacts (4.1 ± 0.5 vs 3.8 ± 0.7, P = 0.007). Scan times of T2BG (286 ± 23 vs 345 ± 68 seconds, P < 0.001) and DWIBG (160 ± 4 vs 252 ± 70 seconds, P < 0.001) were reduced by 17% and 37%, respectively. aSNR and apparent contrast-to-noise ratio (eg, aSNR: 23.45 ± 11.31 [T2BG] vs 25.84 ± 10.76 [T2S]; P = 0.079) were similar for both sequences for both approaches.

Conclusions: Active BG for respiratory-triggered liver T2w and DWI sequences led to significant reduction of breathing artifacts, improved image quality, and shorter scan time compared with standard acquisitions.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
视听呼吸引导提高肝脏 T2 和弥散加权磁共振成像的图像质量和扫描效率
目的:在肝脏磁共振成像(MRI)中,呼吸触发序列经常出现图像质量受损和扫描时间长。我们通过比较标准触发(T2S和DWIS)和BG (T2BG和DWIBG)序列,评估了内腔主动呼吸引导(BG)应用对呼吸触发T2加权(T2)和弥散加权成像(DWI)图像质量和扫描时间的影响。材料和方法:在这项前瞻性研究中,随机选取有肝脏MRI临床指征的患者进行3t MRI,并获得标准和BG。视听BG应用程序接收来自扫描仪的呼吸信号,并通过MRI孔后面的镜子和屏幕显示动画呼吸指示。在获得DWIBG和T2BG之前,患者接受了关于BG MRI的简短视频指导。根据患者的身体呼吸能力(即呼吸4秒,屏气4.5秒),分别为T2BG和DWIBG患者设定理想呼吸模式的合适参数。伪影、清晰度、病变显著性和整体图像质量采用李克特评分从1(非诊断性)到5(优秀)进行评估。对所有序列的扫描时间、视噪比和视信噪比进行了分析。采用配对t检验和Wilcoxon检验进行统计分析。结果:纳入32例患者,平均年龄55±13岁,其中女性13例。T2BG与T2S相比,伪影更少(4.5±0.7 vs 4.1±0.8,P < 0.001),图像清晰度、病灶显著性和整体图像质量(4.6±0.7 vs 4.4±0.7,P = 0.004)更好。与DWIS相比,DWIBG在所有类别中都表现出更高的图像质量(例如,总体图像质量为4.5±0.5 vs 4.3±0.5,P = 0.005),伪影更少(4.1±0.5 vs 3.8±0.7,P = 0.007)。T2BG(286±23 vs 345±68秒,P < 0.001)和DWIBG(160±4 vs 252±70秒,P < 0.001)扫描时间分别减少17%和37%。aSNR和视噪比(如aSNR: 23.45±11.31 [T2BG] vs 25.84±10.76 [T2S]);P = 0.079),两种方法的序列相似。结论:与标准采集相比,呼吸触发肝脏T2w和DWI序列的活性BG显著减少了呼吸伪影,提高了图像质量,缩短了扫描时间。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Investigative Radiology
Investigative Radiology 医学-核医学
CiteScore
15.10
自引率
16.40%
发文量
188
审稿时长
4-8 weeks
期刊介绍: Investigative Radiology publishes original, peer-reviewed reports on clinical and laboratory investigations in diagnostic imaging, the diagnostic use of radioactive isotopes, computed tomography, positron emission tomography, magnetic resonance imaging, ultrasound, digital subtraction angiography, and related modalities. Emphasis is on early and timely publication. Primarily research-oriented, the journal also includes a wide variety of features of interest to clinical radiologists.
期刊最新文献
Photon-Counting Detector CT Applications in Musculoskeletal Radiology. Focused Ultrasound: Noninvasive Image-Guided Therapy. Deep Learning-Enhanced Accelerated 2D TSE and 3D Superresolution Dixon TSE for Rapid Comprehensive Knee Joint Assessment. Hot Topics in Diagnostic Imaging-Encompassing Advances in MR, Photon-Counting CT, and Ultrasound. Three-Dimensional Magnetic Resonance Imaging in the Musculoskeletal System: Clinical Applications and Opportunities to Improve Imaging Speed and Resolution.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1