Estimating Human Fat and Muscle Conductivity From 100 Hz to 1 MHz Using Measurements and Modelling

IF 1.2 3区 生物学 Q3 BIOLOGY Bioelectromagnetics Pub Date : 2025-01-18 DOI:10.1002/bem.22541
Otto Kangasmaa, Ilkka Laakso, Gernot Schmid
{"title":"Estimating Human Fat and Muscle Conductivity From 100 Hz to 1 MHz Using Measurements and Modelling","authors":"Otto Kangasmaa,&nbsp;Ilkka Laakso,&nbsp;Gernot Schmid","doi":"10.1002/bem.22541","DOIUrl":null,"url":null,"abstract":"<p>The electrical conductivity of human tissues is a major source of uncertainty when modelling the interactions between electromagnetic fields and the human body. The aim of this study is to estimate human tissue conductivities in vivo over the low-frequency range, from 30 Hz to 1 MHz. Noninvasive impedance measurements, medical imaging, and 3D surface scanning were performed on the forearms of ten volunteer test subjects. This data set was used to create subject-specific forearm models, numerically solve an electrostatic forward problem, after which the tissue conductivities could be estimated by solving a probabilistic inverse problem. The electrical conductivity of skeletal muscle was found to be highly anisotropic at frequencies below 10 kHz, with conductivities of 0.13 (95% credible interval (CrI): 0.10–0.16) S/m perpendicular and 0.56 (CrI: 0.52–0.60) S/m parallel to the muscle fibre direction. This anisotropy decreased with increasing frequency with these values being 0.65 (CrI: 0.48–1.00) S/m and 0.78 (CrI: 0.72–0.85) S/m at 1 MHz. The conductivity of subcutaneous fat was found to be almost constant across the considered frequency range, with values of 0.21 (CrI: 0.12–0.31) S/m and 0.22 (CrI: 0.07–0.37) S/m at 10 kHz and 1 MHz, respectively. Our study provides robust uncertainty bounds for human tissue conductivity values, which are crucial in the computational assessment of human electromagnetic field exposure. Additionally, our findings are applicable to other fields of modelling such as medical stimulation or measurement technologies.</p>","PeriodicalId":8956,"journal":{"name":"Bioelectromagnetics","volume":"46 1","pages":""},"PeriodicalIF":1.2000,"publicationDate":"2025-01-18","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11742663/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Bioelectromagnetics","FirstCategoryId":"99","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/bem.22541","RegionNum":3,"RegionCategory":"生物学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"BIOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

The electrical conductivity of human tissues is a major source of uncertainty when modelling the interactions between electromagnetic fields and the human body. The aim of this study is to estimate human tissue conductivities in vivo over the low-frequency range, from 30 Hz to 1 MHz. Noninvasive impedance measurements, medical imaging, and 3D surface scanning were performed on the forearms of ten volunteer test subjects. This data set was used to create subject-specific forearm models, numerically solve an electrostatic forward problem, after which the tissue conductivities could be estimated by solving a probabilistic inverse problem. The electrical conductivity of skeletal muscle was found to be highly anisotropic at frequencies below 10 kHz, with conductivities of 0.13 (95% credible interval (CrI): 0.10–0.16) S/m perpendicular and 0.56 (CrI: 0.52–0.60) S/m parallel to the muscle fibre direction. This anisotropy decreased with increasing frequency with these values being 0.65 (CrI: 0.48–1.00) S/m and 0.78 (CrI: 0.72–0.85) S/m at 1 MHz. The conductivity of subcutaneous fat was found to be almost constant across the considered frequency range, with values of 0.21 (CrI: 0.12–0.31) S/m and 0.22 (CrI: 0.07–0.37) S/m at 10 kHz and 1 MHz, respectively. Our study provides robust uncertainty bounds for human tissue conductivity values, which are crucial in the computational assessment of human electromagnetic field exposure. Additionally, our findings are applicable to other fields of modelling such as medical stimulation or measurement technologies.

Abstract Image

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
估计人体脂肪和肌肉电导率从100赫兹至1兆赫使用测量和建模。
在模拟电磁场与人体之间的相互作用时,人体组织的电导率是一个主要的不确定性来源。本研究的目的是在30赫兹至1兆赫的低频范围内估计体内人体组织的电导率。对10名志愿者的前臂进行了无创阻抗测量、医学成像和3D表面扫描。该数据集用于创建受试者特定的前臂模型,数值解决静电正向问题,之后可以通过求解概率逆问题来估计组织电导率。在低于10 kHz的频率下,骨骼肌的电导率具有高度的各向异性,垂直方向的电导率为0.13(95%可信区间(CrI): 0.10-0.16) S/m,平行方向的电导率为0.56 (CrI: 0.52-0.60) S/m。这种各向异性随频率的增加而降低,在1 MHz时,这些值分别为0.65 (CrI: 0.48-1.00) S/m和0.78 (CrI: 0.72-0.85) S/m。皮下脂肪的电导率在所考虑的频率范围内几乎是恒定的,在10 kHz和1 MHz时分别为0.21 (CrI: 0.12-0.31) S/m和0.22 (CrI: 0.07-0.37) S/m。我们的研究为人体组织电导率值提供了强大的不确定性界限,这在人体电磁场暴露的计算评估中至关重要。此外,我们的发现也适用于其他领域的建模,如医疗刺激或测量技术。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Bioelectromagnetics
Bioelectromagnetics 生物-生物物理
CiteScore
4.60
自引率
0.00%
发文量
44
审稿时长
6-12 weeks
期刊介绍: Bioelectromagnetics is published by Wiley-Liss, Inc., for the Bioelectromagnetics Society and is the official journal of the Bioelectromagnetics Society and the European Bioelectromagnetics Association. It is a peer-reviewed, internationally circulated scientific journal that specializes in reporting original data on biological effects and applications of electromagnetic fields that range in frequency from zero hertz (static fields) to the terahertz undulations and visible light. Both experimental and clinical data are of interest to the journal''s readers as are theoretical papers or reviews that offer novel insights into or criticism of contemporary concepts and theories of field-body interactions. The Bioelectromagnetics Society, which sponsors the journal, also welcomes experimental or clinical papers on the domains of sonic and ultrasonic radiation.
期刊最新文献
Issue Information Classification of Cancer Tissue With Machine Learning Algorithms Using Microwave Datasets RF-EMF Risk Perception and Trust in Radiation Protection Authorities: A Comparative Study on Precautionary Information in Germany and Greece. Different Orientations Moderate Static Magnetic Fields Prevent Bone Loss and Improve the Mechanical Properties in Ovariectomized Mice Towards a Planetary Health Impact Assessment Framework: Exploring Expert Knowledge and Artificial Intelligence for a RF-EMF Exposure Case-Study
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1