A straightforward method for assessing the technical image quality of reconstructed and synthetic 2D images for Digital breast tomosynthesis systems

IF 2.7 3区 医学 Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Physica Medica-European Journal of Medical Physics Pub Date : 2025-02-01 Epub Date: 2025-01-21 DOI:10.1016/j.ejmp.2025.104907
V. Ravaglia , V. Ardu , G. Giovannini , S. Dalmonte , M. Porzio , R. Azzeroni , M. Campoleoni , V. Rossetti , P. Golinelli , M. Bertolini , L. Altabella , P. Berardi , L. Berta , P. Bregant , A. Bruschi , B. Cannillo , R. Castriconi , M. Cavallari , M.C. Cimmino , P.E. Colombo , N. Paruccini
{"title":"A straightforward method for assessing the technical image quality of reconstructed and synthetic 2D images for Digital breast tomosynthesis systems","authors":"V. Ravaglia ,&nbsp;V. Ardu ,&nbsp;G. Giovannini ,&nbsp;S. Dalmonte ,&nbsp;M. Porzio ,&nbsp;R. Azzeroni ,&nbsp;M. Campoleoni ,&nbsp;V. Rossetti ,&nbsp;P. Golinelli ,&nbsp;M. Bertolini ,&nbsp;L. Altabella ,&nbsp;P. Berardi ,&nbsp;L. Berta ,&nbsp;P. Bregant ,&nbsp;A. Bruschi ,&nbsp;B. Cannillo ,&nbsp;R. Castriconi ,&nbsp;M. Cavallari ,&nbsp;M.C. Cimmino ,&nbsp;P.E. Colombo ,&nbsp;N. Paruccini","doi":"10.1016/j.ejmp.2025.104907","DOIUrl":null,"url":null,"abstract":"<div><h3>Purpose</h3><div>Digital Breast Tomosynthesis (DBT) is an advanced mammography technique for which there are currently no internationally agreed methods and reference values for image quality assessment. The aim of this multicentre study was to evaluate a simple method to assess the technical image quality of reconstructed and synthetic 2D (SM) images of different models of DBT systems using commercially available phantoms.</div></div><div><h3>Methods</h3><div>The signal difference to noise ratio (SDNR) was chosen as an index of technical image quality and was evaluated for three commercial phantoms, Tomophan, Tormam and CIRS model 015, on 55 DBT systems (six vendors, nine models).</div></div><div><h3>Results</h3><div>SDNR was found to depend on several factors: detail size, average glandular dose (AGD), reconstruction algorithm, software version and applied post-processing. In particular, an increase in SDNR was observed with increasing detail size, AGD, as well as with the use of contrast-enhanced post-processing and iterative reconstruction algorithms. Most systems showed higher SDNR values in SM images respect to DBT for the largest details and a decrease for smaller details.</div></div><div><h3>Conclusions</h3><div>This study proposes a straightforward method to assess the technical image quality of reconstructed DBT and SM images using three different commercial phantoms. The algorithms used to generate SM and DBT images might perform differently when applied to phantoms with homogeneous backgrounds compared to clinical breast structures. However this method could be used to establish reference values for technical image quality during quality control (QC) which could be used to monitor the consistency of the system’s performance over time.</div></div>","PeriodicalId":56092,"journal":{"name":"Physica Medica-European Journal of Medical Physics","volume":"130 ","pages":"Article 104907"},"PeriodicalIF":2.7000,"publicationDate":"2025-02-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Physica Medica-European Journal of Medical Physics","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1120179725000171","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/1/21 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose

Digital Breast Tomosynthesis (DBT) is an advanced mammography technique for which there are currently no internationally agreed methods and reference values for image quality assessment. The aim of this multicentre study was to evaluate a simple method to assess the technical image quality of reconstructed and synthetic 2D (SM) images of different models of DBT systems using commercially available phantoms.

Methods

The signal difference to noise ratio (SDNR) was chosen as an index of technical image quality and was evaluated for three commercial phantoms, Tomophan, Tormam and CIRS model 015, on 55 DBT systems (six vendors, nine models).

Results

SDNR was found to depend on several factors: detail size, average glandular dose (AGD), reconstruction algorithm, software version and applied post-processing. In particular, an increase in SDNR was observed with increasing detail size, AGD, as well as with the use of contrast-enhanced post-processing and iterative reconstruction algorithms. Most systems showed higher SDNR values in SM images respect to DBT for the largest details and a decrease for smaller details.

Conclusions

This study proposes a straightforward method to assess the technical image quality of reconstructed DBT and SM images using three different commercial phantoms. The algorithms used to generate SM and DBT images might perform differently when applied to phantoms with homogeneous backgrounds compared to clinical breast structures. However this method could be used to establish reference values for technical image quality during quality control (QC) which could be used to monitor the consistency of the system’s performance over time.
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
一种评估数字乳房断层合成系统重建和合成二维图像技术图像质量的直接方法。
目的:数字乳腺断层合成(DBT)是一种先进的乳房x线摄影技术,目前尚无国际上一致认可的图像质量评估方法和参考价值。这项多中心研究的目的是评估一种简单的方法来评估不同模型DBT系统的重建和合成二维(SM)图像的技术图像质量。方法:选择信号差噪比(SDNR)作为技术图像质量指标,对55台DBT系统(6家厂商,9个型号)上的Tomophan、tormanm和CIRS型号015三种商用幻影进行评价。结果:发现SDNR取决于几个因素:细节大小、平均腺剂量(AGD)、重建算法、软件版本和应用后处理。特别是,随着细节尺寸、AGD的增加以及对比度增强后处理和迭代重建算法的使用,SDNR会增加。大多数系统在SM图像中,最大细节的SDNR值比DBT高,而较小细节的SDNR值比DBT低。结论:本研究提出了一种简单的方法来评估重建DBT和SM图像使用三种不同的商业幻影的技术图像质量。与临床乳房结构相比,用于生成SM和DBT图像的算法在应用于具有均匀背景的幻影时可能表现不同。然而,该方法可用于在质量控制(QC)期间建立技术图像质量参考值,可用于监测系统性能随时间的一致性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
CiteScore
6.80
自引率
14.70%
发文量
493
审稿时长
78 days
期刊介绍: Physica Medica, European Journal of Medical Physics, publishing with Elsevier from 2007, provides an international forum for research and reviews on the following main topics: Medical Imaging Radiation Therapy Radiation Protection Measuring Systems and Signal Processing Education and training in Medical Physics Professional issues in Medical Physics.
期刊最新文献
Beyond hype: Adoption and attitudes toward generative AI among Indonesian medical physicists Modelling of dose-escalated proton beam therapy for Locally Advanced Pancreatic Cancer using simulated phantoms Application of alanine and ionization chamber dosimetry with a 3D-printed prostate phantom for quality assurance in high dose rate brachytherapy Ensuring reliable digital pathology: a comparative analysis of HistoQC and PathProfiler for artefacts detection in prostate whole-slide images Personalized organ dose assessment from CT imaging during proton therapy for classical Hodgkin lymphoma patients
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1