Elena Kuzmanova, Angela McIntyre, Maaz Syed, Zaid Iskandar, David E Newby, Matt J Bown, Anna-Maria Choy, Jeffrey Tj Huang
{"title":"Accelerating isotope dilution LC-MS-based desmosine quantification for estimating elastin turnover.","authors":"Elena Kuzmanova, Angela McIntyre, Maaz Syed, Zaid Iskandar, David E Newby, Matt J Bown, Anna-Maria Choy, Jeffrey Tj Huang","doi":"10.1080/17576180.2025.2452723","DOIUrl":null,"url":null,"abstract":"<p><strong>Aims: </strong>Circulating total desmosine, representing endogenous systemic elastin degradation activity, is an emerging biomarker for mortality risk in several diseases and aging. However, the existing analytical method takes more than 23 hours to complete, limiting its potential applications. The objective of this study was to shorten the turnover time of a stable isotope dilution liquid chromatogram mass spectrometry-based desmosine assay.</p><p><strong>Materials & methods: </strong>Plasma samples were analyzed using acid hydrolysis followed by solid-phase extraction and LC-MS. Two approaches to reduce assay time were tested: microwave-assisted acid hydrolysis and direct injection following solid-phase extraction.</p><p><strong>Results: </strong>The combination of acid hydrolysis at 180°C for 8 minutes and a low-volume elution design for solid-phase extraction reduced the overall assay time to ~ 30 minutes. The assay was validated with intra-day precision and accuracy ranging from 4% to 14%, and -7% to 9%, respectively, while inter-day precision and accuracy were 0% to 9% and 1% to 3%, respectively. The assay was tested in a cohort of patients with acute aortic dissection and control subjects, where desmosine concentrations were approximately three-fold higher in patients.</p><p><strong>Conclusions: </strong>These results demonstrated that rapid desmosine analysis can be achieved with the use of both microwave-assisted hydrolysis and streamlined solid-phase extraction.</p>","PeriodicalId":8797,"journal":{"name":"Bioanalysis","volume":" ","pages":"99-104"},"PeriodicalIF":1.9000,"publicationDate":"2025-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11801354/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Bioanalysis","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1080/17576180.2025.2452723","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/2/1 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"BIOCHEMICAL RESEARCH METHODS","Score":null,"Total":0}
引用次数: 0
Abstract
Aims: Circulating total desmosine, representing endogenous systemic elastin degradation activity, is an emerging biomarker for mortality risk in several diseases and aging. However, the existing analytical method takes more than 23 hours to complete, limiting its potential applications. The objective of this study was to shorten the turnover time of a stable isotope dilution liquid chromatogram mass spectrometry-based desmosine assay.
Materials & methods: Plasma samples were analyzed using acid hydrolysis followed by solid-phase extraction and LC-MS. Two approaches to reduce assay time were tested: microwave-assisted acid hydrolysis and direct injection following solid-phase extraction.
Results: The combination of acid hydrolysis at 180°C for 8 minutes and a low-volume elution design for solid-phase extraction reduced the overall assay time to ~ 30 minutes. The assay was validated with intra-day precision and accuracy ranging from 4% to 14%, and -7% to 9%, respectively, while inter-day precision and accuracy were 0% to 9% and 1% to 3%, respectively. The assay was tested in a cohort of patients with acute aortic dissection and control subjects, where desmosine concentrations were approximately three-fold higher in patients.
Conclusions: These results demonstrated that rapid desmosine analysis can be achieved with the use of both microwave-assisted hydrolysis and streamlined solid-phase extraction.
BioanalysisBIOCHEMICAL RESEARCH METHODS-CHEMISTRY, ANALYTICAL
CiteScore
3.30
自引率
16.70%
发文量
88
审稿时长
2 months
期刊介绍:
Reliable data obtained from selective, sensitive and reproducible analysis of xenobiotics and biotics in biological samples is a fundamental and crucial part of every successful drug development program. The same principles can also apply to many other areas of research such as forensic science, toxicology and sports doping testing.
The bioanalytical field incorporates sophisticated techniques linking sample preparation and advanced separations with MS and NMR detection systems, automation and robotics. Standards set by regulatory bodies regarding method development and validation increasingly define the boundaries between speed and quality.
Bioanalysis is a progressive discipline for which the future holds many exciting opportunities to further reduce sample volumes, analysis cost and environmental impact, as well as to improve sensitivity, specificity, accuracy, efficiency, assay throughput, data quality, data handling and processing.
The journal Bioanalysis focuses on the techniques and methods used for the detection or quantitative study of analytes in human or animal biological samples. Bioanalysis encourages the submission of articles describing forward-looking applications, including biosensors, microfluidics, miniaturized analytical devices, and new hyphenated and multi-dimensional techniques.
Bioanalysis delivers essential information in concise, at-a-glance article formats. Key advances in the field are reported and analyzed by international experts, providing an authoritative but accessible forum for the modern bioanalyst.