Acute Electrical Synchronization Achieved With Dynamic Atrioventricular Delays During Biventricular and Left Ventricular MultiPoint Pacing

IF 2.5 Q2 CARDIAC & CARDIOVASCULAR SYSTEMS CJC Open Pub Date : 2025-02-01 DOI:10.1016/j.cjco.2024.11.003
Bernard Thibault MD , Peter Waddingham MD , Nima Badie PhD , Jan O. Mangual PhD , Luke C. McSpadden PhD , Tim R. Betts MD , Leonardo Calò MD , Domenico Grieco MD , Francisco Leyva MD , Anthony Chow MD
{"title":"Acute Electrical Synchronization Achieved With Dynamic Atrioventricular Delays During Biventricular and Left Ventricular MultiPoint Pacing","authors":"Bernard Thibault MD ,&nbsp;Peter Waddingham MD ,&nbsp;Nima Badie PhD ,&nbsp;Jan O. Mangual PhD ,&nbsp;Luke C. McSpadden PhD ,&nbsp;Tim R. Betts MD ,&nbsp;Leonardo Calò MD ,&nbsp;Domenico Grieco MD ,&nbsp;Francisco Leyva MD ,&nbsp;Anthony Chow MD","doi":"10.1016/j.cjco.2024.11.003","DOIUrl":null,"url":null,"abstract":"<div><h3>Background</h3><div>Cardiac resynchronization therapy (CRT) response relies on 2 factors: when and where to pace. These factors may be enhanced by dynamic atrioventricular delays (AVDs) (eg, SyncAV CRT, Abbott Cardiovascular, Abbott Park, IL) and multisite left ventricular (LV) pacing (eg, MultiPoint Pacing [MPP], Abbott). Their individual and combined synchronization contributions have not been evaluated across a comprehensive spectrum of pacing configurations. The objective is to distinguish the acute electrical synchrony achieved by static vs dynamic AVDs, single-site vs multisite LV pacing, and with vs without right ventricular (RV) pacing.</div></div><div><h3>Methods</h3><div>CRT-indicated patients with left bundle branch block (LBBB) and intact atrioventricular (AV) conduction (PR &lt; 250 ms) were enrolled and evaluated during implant. Acute changes in 12-lead electrocardiographic (ECG) QRS duration (QRSd) were evaluated during intrinsic conduction, biventricular pacing (BiV), biventricular MPP, LV-only single-site pacing (LVSS), and LV-only MPP (LVMPP). CRT modes were evaluated with static AVDs and optimized SyncAV AVDs.</div></div><div><h3>Results</h3><div>CRT implant and QRSd evaluation were completed in 85 patients (71% male, 34% ischemic, 179 ms PR). The median intrinsic QRSd of 165 ms was reduced by BiV, MPP, LVSS, and LVMPP without SyncAV to 144 ms (by 14%), 142 ms (16%), 155 ms (8%), and 149 ms (12%), respectively (<em>P</em> &lt; 0.01 vs intrinsic). BiV + SyncAV, MPP + SyncAV, LVSS + SyncAV, and LVMPP + SyncAV reduced the intrinsic QRSd significantly further to 128 ms (by 23%), 124 ms (26%), 131 ms (21%), and 129 ms (24%) (<em>P</em> &lt; 0.0001, each corresponding pair).</div></div><div><h3>Conclusions</h3><div>MPP combined with SyncAV achieved the narrowest QRSd, in the overall population and in the most patients, by delivering ventricular pacing from all available sites (RV + LV1 + LV2) while timed with dynamic AVDs.</div></div><div><h3>Clinical Registration Number</h3><div>NCT03567096.</div></div>","PeriodicalId":36924,"journal":{"name":"CJC Open","volume":"7 2","pages":"Pages 166-175"},"PeriodicalIF":2.5000,"publicationDate":"2025-02-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"CJC Open","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2589790X24005201","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0

Abstract

Background

Cardiac resynchronization therapy (CRT) response relies on 2 factors: when and where to pace. These factors may be enhanced by dynamic atrioventricular delays (AVDs) (eg, SyncAV CRT, Abbott Cardiovascular, Abbott Park, IL) and multisite left ventricular (LV) pacing (eg, MultiPoint Pacing [MPP], Abbott). Their individual and combined synchronization contributions have not been evaluated across a comprehensive spectrum of pacing configurations. The objective is to distinguish the acute electrical synchrony achieved by static vs dynamic AVDs, single-site vs multisite LV pacing, and with vs without right ventricular (RV) pacing.

Methods

CRT-indicated patients with left bundle branch block (LBBB) and intact atrioventricular (AV) conduction (PR < 250 ms) were enrolled and evaluated during implant. Acute changes in 12-lead electrocardiographic (ECG) QRS duration (QRSd) were evaluated during intrinsic conduction, biventricular pacing (BiV), biventricular MPP, LV-only single-site pacing (LVSS), and LV-only MPP (LVMPP). CRT modes were evaluated with static AVDs and optimized SyncAV AVDs.

Results

CRT implant and QRSd evaluation were completed in 85 patients (71% male, 34% ischemic, 179 ms PR). The median intrinsic QRSd of 165 ms was reduced by BiV, MPP, LVSS, and LVMPP without SyncAV to 144 ms (by 14%), 142 ms (16%), 155 ms (8%), and 149 ms (12%), respectively (P < 0.01 vs intrinsic). BiV + SyncAV, MPP + SyncAV, LVSS + SyncAV, and LVMPP + SyncAV reduced the intrinsic QRSd significantly further to 128 ms (by 23%), 124 ms (26%), 131 ms (21%), and 129 ms (24%) (P < 0.0001, each corresponding pair).

Conclusions

MPP combined with SyncAV achieved the narrowest QRSd, in the overall population and in the most patients, by delivering ventricular pacing from all available sites (RV + LV1 + LV2) while timed with dynamic AVDs.

Clinical Registration Number

NCT03567096.
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
求助全文
约1分钟内获得全文 去求助
来源期刊
CJC Open
CJC Open Medicine-Cardiology and Cardiovascular Medicine
CiteScore
3.30
自引率
0.00%
发文量
143
审稿时长
60 days
期刊最新文献
First-in-Human Abdominal Aortic Aneurysms Trial with Tricaprin (F-HAAAT): Study Design and Protocol Inflammatory Mediators in Pericardial Fluid in Patients Undergoing Cardiac Surgery Implications of Inaccurate Blood Pressure Measurement on Hypertension Prevalence Femoral Vein Pulsatility and Neurocognitive Disorder in Cardiac Surgery Comparing ECG Lead Subsets for Heart Arrhythmia/ECG Pattern Classification: Convolutional Neural Networks and Random Forest
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1