{"title":"Importance of Considering Temporal Variations in Pulse Wave Velocity for Accurate Blood Pressure Prediction.","authors":"Aditya Satishkumar Bantwal, Amit Kumar Bhayadia, Hui Meng","doi":"10.1007/s10439-025-03681-7","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>Continuous, cuffless blood pressure (BP) monitoring devices based on measuring pulse wave velocity (PWV) or pulse transit time (PTT) are emerging but are often plagued by large prediction errors. A key issue is that these techniques typically rely on a single PWV value, assuming a linear response and small arterial wall deformations. However, arterial response to BP is inherently nonlinear, with PWV varying over time [PWV(t)] by up to 50% during a cardiac cycle. This study evaluates the impact of assuming a single PWV on BP prediction accuracy.</p><p><strong>Method: </strong>Using a Fluid-structure Interaction (FSI) testbed, we simulate the radial and common carotid arteries with the Holzapfel-Gasser-Ogden (HGO) constitutive model to capture nonlinear arterial behavior under a pulsatile physiological blood flow. Pressure data from FSI simulation are used as the ground truth, while inner area A(t) and two PWV values, at diastole and systole, serve as inputs to BP prediction models. Two models are tested: one using a single PWV value, emulating existing PWV-based BP prediction methods; another using the two PWV values to account for PWV(t).</p><p><strong>Results: </strong>The single-PWV BP model produced prediction errors of 17.44 mmHg and 6.57 mmHg for the radial and carotid arteries, respectively. The model incorporating two PWV values reduced these errors by 90.6% and 96.8%, respectively.</p><p><strong>Conclusion: </strong>Relying on a single PWV in BP prediction models can lead to significant errors. To improve BP accuracy, future efforts should focus on incorporating PWV(t), or at least both diastolic and systolic PWV values, into these models.</p>","PeriodicalId":7986,"journal":{"name":"Annals of Biomedical Engineering","volume":" ","pages":""},"PeriodicalIF":3.0000,"publicationDate":"2025-02-06","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Annals of Biomedical Engineering","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1007/s10439-025-03681-7","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose: Continuous, cuffless blood pressure (BP) monitoring devices based on measuring pulse wave velocity (PWV) or pulse transit time (PTT) are emerging but are often plagued by large prediction errors. A key issue is that these techniques typically rely on a single PWV value, assuming a linear response and small arterial wall deformations. However, arterial response to BP is inherently nonlinear, with PWV varying over time [PWV(t)] by up to 50% during a cardiac cycle. This study evaluates the impact of assuming a single PWV on BP prediction accuracy.
Method: Using a Fluid-structure Interaction (FSI) testbed, we simulate the radial and common carotid arteries with the Holzapfel-Gasser-Ogden (HGO) constitutive model to capture nonlinear arterial behavior under a pulsatile physiological blood flow. Pressure data from FSI simulation are used as the ground truth, while inner area A(t) and two PWV values, at diastole and systole, serve as inputs to BP prediction models. Two models are tested: one using a single PWV value, emulating existing PWV-based BP prediction methods; another using the two PWV values to account for PWV(t).
Results: The single-PWV BP model produced prediction errors of 17.44 mmHg and 6.57 mmHg for the radial and carotid arteries, respectively. The model incorporating two PWV values reduced these errors by 90.6% and 96.8%, respectively.
Conclusion: Relying on a single PWV in BP prediction models can lead to significant errors. To improve BP accuracy, future efforts should focus on incorporating PWV(t), or at least both diastolic and systolic PWV values, into these models.
期刊介绍:
Annals of Biomedical Engineering is an official journal of the Biomedical Engineering Society, publishing original articles in the major fields of bioengineering and biomedical engineering. The Annals is an interdisciplinary and international journal with the aim to highlight integrated approaches to the solutions of biological and biomedical problems.