Investigating the Impact of Voxel Size and Postfiltering on Quantitative Analysis of Positron Emission Tomography/Computed Tomography: A Phantom Study.
Ahmed Abdel Mohymen, Hamed Ibrahim Farag, Sameh M Reda, Ahmed Soltan Monem, Said A Ali
{"title":"Investigating the Impact of Voxel Size and Postfiltering on Quantitative Analysis of Positron Emission Tomography/Computed Tomography: A Phantom Study.","authors":"Ahmed Abdel Mohymen, Hamed Ibrahim Farag, Sameh M Reda, Ahmed Soltan Monem, Said A Ali","doi":"10.4103/jmp.jmp_123_24","DOIUrl":null,"url":null,"abstract":"<p><strong>Aim: </strong>This study aims to investigate the influence of voxel size and postfiltering on the quantification of standardized uptake value (SUV) in positron emission tomography/computed tomography (PET/CT) images.</p><p><strong>Materials and methods: </strong>National Electrical Manufacturers Association phantom with the spheres of different sizes were utilized to simulate the lesions. The phantom was scanned using a PET/CT scanner, and the acquired images were reconstructed using two different matrix sizes, (192 × 192) and (256 × 256), and a wide range of postfiltering values.</p><p><strong>Results: </strong>The findings demonstrated that postfiltering significantly affected SUV measurements. The changes in postfiltering values can result in overestimation or underestimation of SUV values, highlighting the importance of carefully selecting appropriate filters. Increasing the matrix size improved SUVmax and SUVmean values, particularly for small-sized spheres. Smaller voxel reconstructions slightly reduced partial volume effects and partially enhanced SUV quantification.</p><p><strong>Conclusions: </strong>Careful consideration of postfiltering values and matrix size selection can lead to better SUV quantification. These findings emphasize the need to optimize the reconstruction parameters to enhance the clinical utility of PET/CT in detecting and evaluating malignant lesions.</p>","PeriodicalId":51719,"journal":{"name":"Journal of Medical Physics","volume":"49 4","pages":"597-607"},"PeriodicalIF":0.7000,"publicationDate":"2024-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11801078/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Medical Physics","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.4103/jmp.jmp_123_24","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/12/18 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
Aim: This study aims to investigate the influence of voxel size and postfiltering on the quantification of standardized uptake value (SUV) in positron emission tomography/computed tomography (PET/CT) images.
Materials and methods: National Electrical Manufacturers Association phantom with the spheres of different sizes were utilized to simulate the lesions. The phantom was scanned using a PET/CT scanner, and the acquired images were reconstructed using two different matrix sizes, (192 × 192) and (256 × 256), and a wide range of postfiltering values.
Results: The findings demonstrated that postfiltering significantly affected SUV measurements. The changes in postfiltering values can result in overestimation or underestimation of SUV values, highlighting the importance of carefully selecting appropriate filters. Increasing the matrix size improved SUVmax and SUVmean values, particularly for small-sized spheres. Smaller voxel reconstructions slightly reduced partial volume effects and partially enhanced SUV quantification.
Conclusions: Careful consideration of postfiltering values and matrix size selection can lead to better SUV quantification. These findings emphasize the need to optimize the reconstruction parameters to enhance the clinical utility of PET/CT in detecting and evaluating malignant lesions.
期刊介绍:
JOURNAL OF MEDICAL PHYSICS is the official journal of Association of Medical Physicists of India (AMPI). The association has been bringing out a quarterly publication since 1976. Till the end of 1993, it was known as Medical Physics Bulletin, which then became Journal of Medical Physics. The main objective of the Journal is to serve as a vehicle of communication to highlight all aspects of the practice of medical radiation physics. The areas covered include all aspects of the application of radiation physics to biological sciences, radiotherapy, radiodiagnosis, nuclear medicine, dosimetry and radiation protection. Papers / manuscripts dealing with the aspects of physics related to cancer therapy / radiobiology also fall within the scope of the journal.