Md Mohiuddin Soliman , Mohammad Tariqul Islam , Phumin Kirawanich , Muhammad E.H. Chowdhury , Touhidul Alam , Ayed M. Alrashdi , Norbahiah Misran , Mohamed S. Soliman
{"title":"Contact Pressure, sliding distance and wear rate analysis at trunnion of hip implant for daily Activities: A finite element approach","authors":"Md Mohiuddin Soliman , Mohammad Tariqul Islam , Phumin Kirawanich , Muhammad E.H. Chowdhury , Touhidul Alam , Ayed M. Alrashdi , Norbahiah Misran , Mohamed S. Soliman","doi":"10.1016/j.bbe.2025.02.001","DOIUrl":null,"url":null,"abstract":"<div><div>This research analyses contact pressure, sliding distance, and wear rate at the trunnion interface of hip implants during various activities to understand post-hip replacement outcomes. The study uses a numerical model and ISO-7206–6:2013 constraints with an AML hip implant. Greater F<sub>x</sub>, F<sub>y</sub>, and F<sub>z</sub> forces broaden contact pressure distribution. The highest pressure occurs on the proximal superolateral surface, with the lowest on the anterior and posterior surfaces. The HIGH100 (individuals weighing above 100 kg) weight category demonstrates 2 times higher maximum and average contact pressure compared to AVG75 (individuals weighing 75 kg) for sit-down and knee bend activities. Force components and the duration of a full gait cycle influence sliding distance. Stance activities show the highest sliding distance due to rapid changes in force load during the gait cycle. For sit-down and knee bend activities, the total sliding distance is 2.5 times higher in the HIGH100 weight category compared to AVG75. Sliding distance primarily occurs at the proximal superolateral-inferomedial-anterior-posterior contact surface, decreasing distally. Based on contact pressure, sliding distance, and wear volume rate, jogging and stance activities pose the highest risk for hip replacement patients, while cycling is the safest. The HIGH100 weight group exhibits 5- and 4-times greater wear volume rates than AVG75 in sit-down and knee bend activities, respectively. The research findings align with wear degradation observed in retrieved hip implants, validating the study. These insights can assist patients in making informed decisions about performing activities after surgery while enabling physicians to provide accurate guidelines.</div></div>","PeriodicalId":55381,"journal":{"name":"Biocybernetics and Biomedical Engineering","volume":"45 2","pages":"Pages 137-153"},"PeriodicalIF":5.3000,"publicationDate":"2025-02-24","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Biocybernetics and Biomedical Engineering","FirstCategoryId":"5","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0208521625000063","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 0
Abstract
This research analyses contact pressure, sliding distance, and wear rate at the trunnion interface of hip implants during various activities to understand post-hip replacement outcomes. The study uses a numerical model and ISO-7206–6:2013 constraints with an AML hip implant. Greater Fx, Fy, and Fz forces broaden contact pressure distribution. The highest pressure occurs on the proximal superolateral surface, with the lowest on the anterior and posterior surfaces. The HIGH100 (individuals weighing above 100 kg) weight category demonstrates 2 times higher maximum and average contact pressure compared to AVG75 (individuals weighing 75 kg) for sit-down and knee bend activities. Force components and the duration of a full gait cycle influence sliding distance. Stance activities show the highest sliding distance due to rapid changes in force load during the gait cycle. For sit-down and knee bend activities, the total sliding distance is 2.5 times higher in the HIGH100 weight category compared to AVG75. Sliding distance primarily occurs at the proximal superolateral-inferomedial-anterior-posterior contact surface, decreasing distally. Based on contact pressure, sliding distance, and wear volume rate, jogging and stance activities pose the highest risk for hip replacement patients, while cycling is the safest. The HIGH100 weight group exhibits 5- and 4-times greater wear volume rates than AVG75 in sit-down and knee bend activities, respectively. The research findings align with wear degradation observed in retrieved hip implants, validating the study. These insights can assist patients in making informed decisions about performing activities after surgery while enabling physicians to provide accurate guidelines.
期刊介绍:
Biocybernetics and Biomedical Engineering is a quarterly journal, founded in 1981, devoted to publishing the results of original, innovative and creative research investigations in the field of Biocybernetics and biomedical engineering, which bridges mathematical, physical, chemical and engineering methods and technology to analyse physiological processes in living organisms as well as to develop methods, devices and systems used in biology and medicine, mainly in medical diagnosis, monitoring systems and therapy. The Journal''s mission is to advance scientific discovery into new or improved standards of care, and promotion a wide-ranging exchange between science and its application to humans.