High concordance of blood glucose measurement in dogs between a beta prototype glucometer device and a reference laboratory standard in a clinical setting

IF 1.9 2区 农林科学 Q2 VETERINARY SCIENCES Journal of Small Animal Practice Pub Date : 2025-02-25 DOI:10.1111/jsap.13847
K. Palladino, S. Velineni, S. Meyer, A. Rhodes, S. D. Stewart
{"title":"High concordance of blood glucose measurement in dogs between a beta prototype glucometer device and a reference laboratory standard in a clinical setting","authors":"K. Palladino,&nbsp;S. Velineni,&nbsp;S. Meyer,&nbsp;A. Rhodes,&nbsp;S. D. Stewart","doi":"10.1111/jsap.13847","DOIUrl":null,"url":null,"abstract":"<div>\n \n <section>\n \n <h3> Objectives</h3>\n \n <p>The objective of this study was to confirm the accuracy of a new portable beta prototype glucometer device for monitoring blood glucose levels in dogs.</p>\n </section>\n \n <section>\n \n <h3> Materials and Methods</h3>\n \n <p>One hundred and three client-owned dogs with dysglycemia or normoglycemia were enrolled. A 3 mL whole blood sample was collected from a peripheral vein from each dog and divided into four aliquots. Blood glucose was measured from three of the aliquots on the beta prototype glucometer in triplicate. The remaining aliquot was used to measure packed cell volume, total solids, and blood glucose via a reference laboratory chemistry analyzer. Accuracy was assessed in accordance with the International Organization for Standardization guidelines.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>The analytical performance of the beta prototype met the International Organization for Standardization accuracy requirements, with 95.1% of measurements falling within the International Organization for Standardization accuracy threshold as well as 100% of measurements being within zones A and B of the consensus error grid. Within the full range of glucose concentrations (20 to 750 mg/dL) the mean difference between the AT3 and reference lab analyzer was 2.53 mg/dL (95% limit of agreement: −45.27 to 50.34).</p>\n </section>\n \n <section>\n \n <h3> Clinical Significance</h3>\n \n <p>The beta prototype glucometer device had a strong correlation with the reference laboratory standard and can be used confidently in clinical patients for the accurate monitoring of blood glucose levels in dogs.</p>\n </section>\n </div>","PeriodicalId":17062,"journal":{"name":"Journal of Small Animal Practice","volume":"66 7","pages":"451-457"},"PeriodicalIF":1.9000,"publicationDate":"2025-02-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Small Animal Practice","FirstCategoryId":"97","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1111/jsap.13847","RegionNum":2,"RegionCategory":"农林科学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"VETERINARY SCIENCES","Score":null,"Total":0}
引用次数: 0

Abstract

Objectives

The objective of this study was to confirm the accuracy of a new portable beta prototype glucometer device for monitoring blood glucose levels in dogs.

Materials and Methods

One hundred and three client-owned dogs with dysglycemia or normoglycemia were enrolled. A 3 mL whole blood sample was collected from a peripheral vein from each dog and divided into four aliquots. Blood glucose was measured from three of the aliquots on the beta prototype glucometer in triplicate. The remaining aliquot was used to measure packed cell volume, total solids, and blood glucose via a reference laboratory chemistry analyzer. Accuracy was assessed in accordance with the International Organization for Standardization guidelines.

Results

The analytical performance of the beta prototype met the International Organization for Standardization accuracy requirements, with 95.1% of measurements falling within the International Organization for Standardization accuracy threshold as well as 100% of measurements being within zones A and B of the consensus error grid. Within the full range of glucose concentrations (20 to 750 mg/dL) the mean difference between the AT3 and reference lab analyzer was 2.53 mg/dL (95% limit of agreement: −45.27 to 50.34).

Clinical Significance

The beta prototype glucometer device had a strong correlation with the reference laboratory standard and can be used confidently in clinical patients for the accurate monitoring of blood glucose levels in dogs.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
在狗的血糖测量之间的β原型血糖仪设备和参考实验室标准在临床设置的高度一致性。
目的:本研究的目的是确认一种用于监测狗血糖水平的新型便携式beta原型血糖仪的准确性。材料与方法:入选103只血糖异常或血糖正常的客户犬。从每只狗的外周静脉采集3ml全血样本,分成四等份。在beta原型血糖仪上的三个等分液中测量血糖,一式三份。剩余的等温物通过参考实验室化学分析仪测量堆积细胞体积、总固体和血糖。准确性根据国际标准化组织的指导方针进行评估。结果:beta原型的分析性能满足国际标准化组织的精度要求,95.1%的测量值落在国际标准化组织的精度阈值内,100%的测量值落在共识误差网格的A区和B区内。在整个葡萄糖浓度范围内(20至750 mg/dL), AT3和参考实验室分析仪之间的平均差异为2.53 mg/dL(95%一致性限:-45.27至50.34)。临床意义:该beta原型血糖仪设备与实验室参考标准相关性强,可放心用于临床患者,准确监测犬的血糖水平。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Journal of Small Animal Practice
Journal of Small Animal Practice 农林科学-兽医学
CiteScore
3.30
自引率
6.20%
发文量
117
审稿时长
12-24 weeks
期刊介绍: Journal of Small Animal Practice (JSAP) is a monthly peer-reviewed publication integrating clinical research papers and case reports from international sources, covering all aspects of medicine and surgery relating to dogs, cats and other small animals. These papers facilitate the dissemination and implementation of new ideas and techniques relating to clinical veterinary practice, with the ultimate aim of promoting best practice. JSAP publishes high quality original articles, as well as other scientific and educational information. New developments are placed in perspective, encompassing new concepts and peer commentary. The target audience is veterinarians primarily engaged in the practise of small animal medicine and surgery. In addition to original articles, JSAP will publish invited editorials (relating to a manuscript in the same issue or a topic of current interest), review articles, which provide in-depth discussion of important clinical issues, and other scientific and educational information from around the world. The final decision on publication of a manuscript rests with the Editorial Board and ultimately with the Editor. All papers, regardless of type, represent the opinion of the authors and not necessarily that of the Editor, the Association or the Publisher. The Journal of Small Animal Practice is published on behalf of the British Small Animal Veterinary Association and is also the official scientific journal of the World Small Animal Veterinary Association
期刊最新文献
Bilateral ischaemic neuromyopathy due to axillary artery thrombosis in a dog. A cross-sectional radiographic study on the prevalence and distribution of dentigerous cysts in unerupted teeth in adult dogs. BSAVA Petsavers Article: Validation of an ELISA assay for measurement of the metabolite of serotonin, 5-hydroxyindole acetic acid (5-HIAA), in canine urine. Phenotypic characteristics of paroxysmal dyskinesia in 25 cats. Bilateral surgery of humeral intracondylar fissures and humeral condylar fractures is associated with increased complication rates.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1