Establishing an in-house quality control program for nine autoantibody assays using donor sera

IF 2.5 3区 医学 Q2 MEDICAL LABORATORY TECHNOLOGY Clinical biochemistry Pub Date : 2025-03-01 DOI:10.1016/j.clinbiochem.2025.110900
Mary Kathryn Bohn , Lusia Sepiashvili
{"title":"Establishing an in-house quality control program for nine autoantibody assays using donor sera","authors":"Mary Kathryn Bohn ,&nbsp;Lusia Sepiashvili","doi":"10.1016/j.clinbiochem.2025.110900","DOIUrl":null,"url":null,"abstract":"<div><h3>Background</h3><div>Clinical laboratories performing semi-quantitative autoantibody testing face challenges in developing quality systems to assure performance. Careful consideration is required to evaluate assay imprecision, including quality control (QC) matrix, autoantibody titer targets, and performance goals. The objective of this study was to evaluate the performance of a novel patient-based QC solution relative to vendor-provided QC for nine autoantibodies over a four-year period at a tertiary pediatric care centre.</div></div><div><h3>Methods</h3><div>Internal QC data were extracted over a 4-year period for nine autoantibodies measured via chemiluminescent immunoassays (anti-double stranded DNA (dsDNA), anti-ribonuclear protein, anti-Ro52, anti-Ro60, anti-La, anti-Smith, anti-proteinase 3, anti-myeloperoxidase, and anti-tissue transglutaminase IgA). QC evaluated during study period included vendor-based QC and third-party patient-based QC using defibrinated plasma specimens from autoantibody positive single donors. Lot-specific QC coefficients of variation (CV), titer means, and standard deviations were calculated and compared across QC matrices.</div></div><div><h3>Results</h3><div>Approximately 500 QC values per autoantibody were evaluated. Mean CV across lots ranged from 8.2 to 14.4% for negative vendor-based QC, 8.2–14.5% for positive vendor-based QC, and 9.8–17.8% for positive patient-based QC across evaluated autoantibodies. Imprecision estimates were similar between positive vendor and patient-based QC for all autoantibodies with the exception of anti-dsDNA (vendor: 4.8–13.2%, patient-based: 14.3–21.6%) and anti-Ro52 (vendor: 6.9–12.4%; patient-based: 9.7–18.9%).</div></div><div><h3>Conclusions</h3><div>This study characterizes imprecision across different QC matrices for nine autoantibodies measured semi-quantitatively using chemiluminescent immunoassays. Results may serve as a benchmark tool to assess imprecision goals for commonly measured autoantibodies in clinical laboratories and as a framework for utilizing third-party patient-based QC solutions.</div></div>","PeriodicalId":10172,"journal":{"name":"Clinical biochemistry","volume":"136 ","pages":"Article 110900"},"PeriodicalIF":2.5000,"publicationDate":"2025-03-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Clinical biochemistry","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0009912025000293","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Background

Clinical laboratories performing semi-quantitative autoantibody testing face challenges in developing quality systems to assure performance. Careful consideration is required to evaluate assay imprecision, including quality control (QC) matrix, autoantibody titer targets, and performance goals. The objective of this study was to evaluate the performance of a novel patient-based QC solution relative to vendor-provided QC for nine autoantibodies over a four-year period at a tertiary pediatric care centre.

Methods

Internal QC data were extracted over a 4-year period for nine autoantibodies measured via chemiluminescent immunoassays (anti-double stranded DNA (dsDNA), anti-ribonuclear protein, anti-Ro52, anti-Ro60, anti-La, anti-Smith, anti-proteinase 3, anti-myeloperoxidase, and anti-tissue transglutaminase IgA). QC evaluated during study period included vendor-based QC and third-party patient-based QC using defibrinated plasma specimens from autoantibody positive single donors. Lot-specific QC coefficients of variation (CV), titer means, and standard deviations were calculated and compared across QC matrices.

Results

Approximately 500 QC values per autoantibody were evaluated. Mean CV across lots ranged from 8.2 to 14.4% for negative vendor-based QC, 8.2–14.5% for positive vendor-based QC, and 9.8–17.8% for positive patient-based QC across evaluated autoantibodies. Imprecision estimates were similar between positive vendor and patient-based QC for all autoantibodies with the exception of anti-dsDNA (vendor: 4.8–13.2%, patient-based: 14.3–21.6%) and anti-Ro52 (vendor: 6.9–12.4%; patient-based: 9.7–18.9%).

Conclusions

This study characterizes imprecision across different QC matrices for nine autoantibodies measured semi-quantitatively using chemiluminescent immunoassays. Results may serve as a benchmark tool to assess imprecision goals for commonly measured autoantibodies in clinical laboratories and as a framework for utilizing third-party patient-based QC solutions.
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
求助全文
约1分钟内获得全文 去求助
来源期刊
Clinical biochemistry
Clinical biochemistry 医学-医学实验技术
CiteScore
5.10
自引率
0.00%
发文量
151
审稿时长
25 days
期刊介绍: Clinical Biochemistry publishes articles relating to clinical chemistry, molecular biology and genetics, therapeutic drug monitoring and toxicology, laboratory immunology and laboratory medicine in general, with the focus on analytical and clinical investigation of laboratory tests in humans used for diagnosis, prognosis, treatment and therapy, and monitoring of disease.
期刊最新文献
Cardiac troponin I is associated with ICU admission in pediatric patients with RSV Establishing an in-house quality control program for nine autoantibody assays using donor sera Knowledge about globin genetics for precision diagnosis of hemoglobinopathies: A case Study Secretoneurin is not associated with cardiovascular events or mortality in patients treated with hemodialysis: A prospective multicenter cohort study Diagnostic accuracy of breath tests based on volatile organic compounds for cancer: A systematic review and meta-analysis
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1