Validation of the CONTEC08A upper-arm blood pressure monitor in adult Chinese.

IF 1.2 4区 医学 Q4 PERIPHERAL VASCULAR DISEASE Blood Pressure Monitoring Pub Date : 2025-06-01 DOI:10.1097/MBP.0000000000000743
Yi Zhou, Yi Zhou, Ruo-Yi Zou, Yan Li, Ji-Guang Wang
{"title":"Validation of the CONTEC08A upper-arm blood pressure monitor in adult Chinese.","authors":"Yi Zhou, Yi Zhou, Ruo-Yi Zou, Yan Li, Ji-Guang Wang","doi":"10.1097/MBP.0000000000000743","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>The aim of this study was to evaluate the accuracy of the CONTEC08A oscillometric upper-arm blood pressure (BP) monitor for BP measurement in adults according to the International Organization for Standardization (ISO) 81060-2:2018 and amendment (Amd)1:2020 standard.</p><p><strong>Methods: </strong>Eighty-five subjects (male 40 and female 45) were recruited to fulfill the age, sex, BP, and cuff distribution criteria of the ISO standard in the general population, and had a mean age of 40.2 years. The same arm sequential BP measurement method was used with three differently sized cuffs for the arm circumference ranging from 18 to 26 cm (small), 22-32 cm (medium), and 22-43 cm (large), respectively, for the test device, and two differently sized cuffs for the arm circumference ≤32 cm (standard) and >32 cm (large), respectively, for the mercury sphygmomanometer.</p><p><strong>Results: </strong>Two hundred and fifty-five comparison pairs were obtained for analysis. For validation criterion 1, the mean ± SD of the differences between the test device and reference SBP/DBP readings was 1.1 ± 6.7/3.0 ± 5.3 mmHg. For validation criterion 2, the SD of the averaged SBP/DBP differences between the test device and reference BP per subject was 5.63/4.68 mmHg.</p><p><strong>Conclusion: </strong>The automated upper-arm BP monitor CONTEC08A has passed the requirements of the ISO Universal Standard in the general population, and can be recommended for BP measurement in adults.</p>","PeriodicalId":8950,"journal":{"name":"Blood Pressure Monitoring","volume":" ","pages":"109-112"},"PeriodicalIF":1.2000,"publicationDate":"2025-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Blood Pressure Monitoring","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/MBP.0000000000000743","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"PERIPHERAL VASCULAR DISEASE","Score":null,"Total":0}
引用次数: 0

Abstract

Objective: The aim of this study was to evaluate the accuracy of the CONTEC08A oscillometric upper-arm blood pressure (BP) monitor for BP measurement in adults according to the International Organization for Standardization (ISO) 81060-2:2018 and amendment (Amd)1:2020 standard.

Methods: Eighty-five subjects (male 40 and female 45) were recruited to fulfill the age, sex, BP, and cuff distribution criteria of the ISO standard in the general population, and had a mean age of 40.2 years. The same arm sequential BP measurement method was used with three differently sized cuffs for the arm circumference ranging from 18 to 26 cm (small), 22-32 cm (medium), and 22-43 cm (large), respectively, for the test device, and two differently sized cuffs for the arm circumference ≤32 cm (standard) and >32 cm (large), respectively, for the mercury sphygmomanometer.

Results: Two hundred and fifty-five comparison pairs were obtained for analysis. For validation criterion 1, the mean ± SD of the differences between the test device and reference SBP/DBP readings was 1.1 ± 6.7/3.0 ± 5.3 mmHg. For validation criterion 2, the SD of the averaged SBP/DBP differences between the test device and reference BP per subject was 5.63/4.68 mmHg.

Conclusion: The automated upper-arm BP monitor CONTEC08A has passed the requirements of the ISO Universal Standard in the general population, and can be recommended for BP measurement in adults.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
CONTEC08A上臂血压计在中国成人中的验证。
目的:本研究的目的是根据国际标准化组织(ISO) 81060-2:2018和修订(Amd)1:2020标准,评估CONTEC08A振荡式上臂血压(BP)监测仪用于成人血压测量的准确性。方法:招募符合一般人群年龄、性别、血压、袖带分布标准的85例受试者,男40岁,女45岁,平均年龄40.2岁。采用相同的手臂顺序血压测量方法,测试装置的臂围分别为18 - 26 cm(小)、22-32 cm(中)和22-43 cm(大),分别为三个不同尺寸的袖口;水银血压计的臂围分别为≤32 cm(标准)和>32 cm(大)。结果:共获得255对对照对进行分析。对于验证标准1,测试装置与参考收缩压/舒张压读数差异的平均值±SD为1.1±6.7/3.0±5.3 mmHg。对于验证标准2,测试装置与参考血压之间的平均收缩压/舒张压差的标准差为5.63/4.68 mmHg。结论:全自动上臂血压监测仪CONTEC08A在普通人群中已通过ISO通用标准的要求,可推荐用于成人血压测量。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Blood Pressure Monitoring
Blood Pressure Monitoring 医学-外周血管病
CiteScore
2.00
自引率
7.70%
发文量
110
审稿时长
>12 weeks
期刊介绍: Blood Pressure Monitoring is devoted to original research in blood pressure measurement and blood pressure variability. It includes device technology, analytical methodology of blood pressure over time and its variability, clinical trials - including, but not limited to, pharmacology - involving blood pressure monitoring, blood pressure reactivity, patient evaluation, and outcomes and effectiveness research. This innovative journal contains papers dealing with all aspects of manual, automated, and ambulatory monitoring. Basic and clinical science papers are considered although the emphasis is on clinical medicine. Submitted articles undergo a preliminary review by the editor. Some articles may be returned to authors without further consideration. Those being considered for publication will undergo further assessment and peer-review by the editors and those invited to do so from a reviewer pool.
期刊最新文献
Validation of the automated oscillometric upper-arm cuff home blood pressure monitor Microlife BP3KV1-5X (BP B6 Connect) in a general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization Universal Standard (ISO 81060-2:2018). Association between the direction of the blood pressure discrepancy and vascular aging in a large-scale screening population. Validation of the automated oscillometric upper-arm cuff home blood pressure monitor Microlife BP3KT1-4X (BP B3 Comfort PC) in a general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization Universal Standard (ISO 81060-2:2018). Combination of treatment intensity score and interankle diastolic blood pressure is associated with cardiovascular prognosis among adults with controlled hypertension who have pre-existing cardiovascular disease. Validation of the BTL ambulatory blood pressure monitoring NG ambulatory blood pressure monitoring device according to the Universal Standard International Organization for Standardization 81060-2:2018.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1