Real-world comparison of efficacy and safety of XEN45 implant with phacoemulsification versus iStent inject® W with phacoemulsification

IF 1.2 4区 医学 Q3 OPHTHALMOLOGY Journal Francais D Ophtalmologie Pub Date : 2025-03-08 DOI:10.1016/j.jfo.2025.104469
A.L. Mendoza-Moreira , M.T. Marcos-Parra , S. de Bari , J. Ezbakhe , E. España-Vera , M.C. Calatayud-Hernández , J.J. Pérez-Santonja
{"title":"Real-world comparison of efficacy and safety of XEN45 implant with phacoemulsification versus iStent inject® W with phacoemulsification","authors":"A.L. Mendoza-Moreira ,&nbsp;M.T. Marcos-Parra ,&nbsp;S. de Bari ,&nbsp;J. Ezbakhe ,&nbsp;E. España-Vera ,&nbsp;M.C. Calatayud-Hernández ,&nbsp;J.J. Pérez-Santonja","doi":"10.1016/j.jfo.2025.104469","DOIUrl":null,"url":null,"abstract":"<div><h3>Purpose</h3><div>To compare iStent inject® W<!--> <!-->+<!--> <!-->phacoemulsification with XEN45 implant<!--> <!-->+<!--> <!-->phacoemulsification over 12 months in mild to moderate glaucoma patients needing cataract surgery at the Hospital General Universitario Dr. Balmis, Alicante, Spain.</div></div><div><h3>Methods</h3><div>Retrospective clinical cohort study. Consecutive patients who underwent phacoemulsification either combined with XEN45 implantation or iStent inject® W from 2020 to 2022 were included. The main outcome measure was surgical success at 12 months postoperatively. Intraocular pressure (IOP), number of antiglaucoma medications, intraoperative and postoperative complications, number of revision surgeries, use of the operating room and required visits were also evaluated.</div></div><div><h3>Results</h3><div>A total of 167 eyes were included in the study, with 101 eyes of 61 patients undergoing iStent<!--> <!-->+<!--> <!-->PHACO and 66 eyes of 46 patients undergoing XEN45<!--> <!-->+<!--> <!-->PHACO. The proportion of patients achieving a complete success (IOP<!--> <!-->≥<!--> <!-->5 and ≤ 18<!--> <!-->mmHg without medication) at the conclusion of the 12-month follow-up was 47.9% (35/73) in the iStent<!--> <!-->+<!--> <!-->PHACO and 48.2% (27/56) in the XEN45<!--> <!-->+<!--> <!-->PHACO surgery group, <em>P</em> <!-->=<!--> <!-->1.000. The mean (95% CI) IOP reduction at the conclusion of the study follow-up was –2.4 (–3.5 to –1.3<!--> <!-->mmHg, <em>P</em> <!-->&lt;<!--> <!-->0.001, iStent<!--> <!-->+<!--> <!-->PHACO) and –3.10 (–4.8 to –1.4<!--> <!-->mmHg, <em>P</em> <!-->&lt;<!--> <!-->0.001, XEN45<!--> <!-->+<!--> <!-->PHACO). The mean number of antiglaucoma medications was significantly reduced in both study groups. Rates of intraoperative complications (5 vs. 15.2%), postoperative complications (1 vs. 46.2%), reoperations (0 vs. 24.6%), mean number of operating room visits (1 vs. 1.41), and mean number of required postoperative visits (4.50 vs. 11.13) were all statistically significantly higher in the XEN45<!--> <!-->+<!--> <!-->PHACO group (&lt; 0.001).</div></div><div><h3>Conclusions</h3><div>Both procedures achieved similar surgical success rates, with comparable reductions in IOP and the number of antiglaucoma medications. After one year, a modest IOP reduction of 10.83% was observed in the iStent group and 13.65% in the Xen group. However, the iStent inject® W demonstrated a better intraoperative and postoperative safety profile.</div></div><div><h3>Objectif</h3><div>L’objectif était de comparer l’iStent inject® W associé à une phacoémulsification avec l’implant XEN45 associé à une phacoémulsification sur une période de 12 mois chez des patients atteints de glaucome léger à modéré nécessitant une chirurgie de la cataracte à l’Hôpital General Universitario Dr Balmis, à Alicante, en Espagne.</div></div><div><h3>Méthodes</h3><div>Étude clinique rétrospective de cohorte. Des patients consécutifs ayant subi une phacoémulsification combinée avec l’implantation du XEN45 ou de l’iStent inject® W entre 2020 et 2022 ont été inclus. La principale mesure de résultat était le succès chirurgical à 12 mois postopératoires. La pression intraoculaire (PIO), le nombre de médicaments antiglaucomateux, les complications peropératoires et postopératoires, le nombre de chirurgies de révision, l’utilisation de la salle d’opération et les visites nécessaires ont également été évalués.</div></div><div><h3>Résultats</h3><div>Au total, 167 yeux ont été inclus dans l’étude, avec 101 yeux de 61 patients ayant subi la procédure iStent<!--> <!-->+<!--> <!-->PHACO et 66 yeux de 46 patients ayant subi la procédure XEN45<!--> <!-->+<!--> <!-->PHACO. La proportion de patients ayant atteint un succès complet (PIO<!--> <!-->≥<!--> <!-->5 et ≤ 18<!--> <!-->mmHg sans traitement) à la fin des 12 mois de suivi était de 47,9 % (35/73) dans le groupe iStent<!--> <!-->+<!--> <!-->PHACO et de 48,2 % (27/56) dans le groupe XEN45<!--> <!-->+<!--> <!-->PHACO, <em>p</em> <!-->=<!--> <!-->1,000. La réduction moyenne de la PIO (IC 95 %) à la fin de l’étude était de –2,4 (–3,5 à –1,3<!--> <!-->mmHg, <em>p</em> <!-->&lt;<!--> <!-->0,001, iStent<!--> <!-->+<!--> <!-->PHACO) et de –3,10 (–4,8 à –1,4<!--> <!-->mmHg, <em>p</em> <!-->&lt;<!--> <!-->0,001, XEN45<!--> <!-->+<!--> <!-->PHACO). Le nombre moyen de médicaments antiglaucomateux a été significativement réduit dans les deux groupes de l’étude. Les taux de complications peropératoires (5 vs 15,2 %), de complications postopératoires (1 vs 46,2 %), de réinterventions (0 vs 24,6 %), le nombre moyen de visites en salle d’opération (1 vs 1,41), et le nombre moyen de visites postopératoires nécessaires (4,50 vs 11,13) étaient tous significativement plus élevés dans le groupe XEN45<!--> <!-->+<!--> <!-->PHACO (<em>p</em> <!-->&lt;<!--> <!-->0,001).</div></div><div><h3>Conclusion</h3><div>Les deux procédures ont atteint des taux de succès chirurgicaux similaires, avec des réductions comparables de la PIO et du nombre de médicaments antiglaucomateux. Après un an, une réduction modeste de la PIO de 10,83 % a été observée dans le groupe iStent et de 13,65 % dans le groupe XEN-45. Cependant, l’iStent inject® W a démontré un meilleur profil de sécurité peropératoire et postopératoire.</div></div>","PeriodicalId":14777,"journal":{"name":"Journal Francais D Ophtalmologie","volume":"48 4","pages":"Article 104469"},"PeriodicalIF":1.2000,"publicationDate":"2025-03-08","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal Francais D Ophtalmologie","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0181551225000518","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose

To compare iStent inject® W + phacoemulsification with XEN45 implant + phacoemulsification over 12 months in mild to moderate glaucoma patients needing cataract surgery at the Hospital General Universitario Dr. Balmis, Alicante, Spain.

Methods

Retrospective clinical cohort study. Consecutive patients who underwent phacoemulsification either combined with XEN45 implantation or iStent inject® W from 2020 to 2022 were included. The main outcome measure was surgical success at 12 months postoperatively. Intraocular pressure (IOP), number of antiglaucoma medications, intraoperative and postoperative complications, number of revision surgeries, use of the operating room and required visits were also evaluated.

Results

A total of 167 eyes were included in the study, with 101 eyes of 61 patients undergoing iStent + PHACO and 66 eyes of 46 patients undergoing XEN45 + PHACO. The proportion of patients achieving a complete success (IOP  5 and ≤ 18 mmHg without medication) at the conclusion of the 12-month follow-up was 47.9% (35/73) in the iStent + PHACO and 48.2% (27/56) in the XEN45 + PHACO surgery group, P = 1.000. The mean (95% CI) IOP reduction at the conclusion of the study follow-up was –2.4 (–3.5 to –1.3 mmHg, P < 0.001, iStent + PHACO) and –3.10 (–4.8 to –1.4 mmHg, P < 0.001, XEN45 + PHACO). The mean number of antiglaucoma medications was significantly reduced in both study groups. Rates of intraoperative complications (5 vs. 15.2%), postoperative complications (1 vs. 46.2%), reoperations (0 vs. 24.6%), mean number of operating room visits (1 vs. 1.41), and mean number of required postoperative visits (4.50 vs. 11.13) were all statistically significantly higher in the XEN45 + PHACO group (< 0.001).

Conclusions

Both procedures achieved similar surgical success rates, with comparable reductions in IOP and the number of antiglaucoma medications. After one year, a modest IOP reduction of 10.83% was observed in the iStent group and 13.65% in the Xen group. However, the iStent inject® W demonstrated a better intraoperative and postoperative safety profile.

Objectif

L’objectif était de comparer l’iStent inject® W associé à une phacoémulsification avec l’implant XEN45 associé à une phacoémulsification sur une période de 12 mois chez des patients atteints de glaucome léger à modéré nécessitant une chirurgie de la cataracte à l’Hôpital General Universitario Dr Balmis, à Alicante, en Espagne.

Méthodes

Étude clinique rétrospective de cohorte. Des patients consécutifs ayant subi une phacoémulsification combinée avec l’implantation du XEN45 ou de l’iStent inject® W entre 2020 et 2022 ont été inclus. La principale mesure de résultat était le succès chirurgical à 12 mois postopératoires. La pression intraoculaire (PIO), le nombre de médicaments antiglaucomateux, les complications peropératoires et postopératoires, le nombre de chirurgies de révision, l’utilisation de la salle d’opération et les visites nécessaires ont également été évalués.

Résultats

Au total, 167 yeux ont été inclus dans l’étude, avec 101 yeux de 61 patients ayant subi la procédure iStent + PHACO et 66 yeux de 46 patients ayant subi la procédure XEN45 + PHACO. La proportion de patients ayant atteint un succès complet (PIO  5 et ≤ 18 mmHg sans traitement) à la fin des 12 mois de suivi était de 47,9 % (35/73) dans le groupe iStent + PHACO et de 48,2 % (27/56) dans le groupe XEN45 + PHACO, p = 1,000. La réduction moyenne de la PIO (IC 95 %) à la fin de l’étude était de –2,4 (–3,5 à –1,3 mmHg, p < 0,001, iStent + PHACO) et de –3,10 (–4,8 à –1,4 mmHg, p < 0,001, XEN45 + PHACO). Le nombre moyen de médicaments antiglaucomateux a été significativement réduit dans les deux groupes de l’étude. Les taux de complications peropératoires (5 vs 15,2 %), de complications postopératoires (1 vs 46,2 %), de réinterventions (0 vs 24,6 %), le nombre moyen de visites en salle d’opération (1 vs 1,41), et le nombre moyen de visites postopératoires nécessaires (4,50 vs 11,13) étaient tous significativement plus élevés dans le groupe XEN45 + PHACO (p < 0,001).

Conclusion

Les deux procédures ont atteint des taux de succès chirurgicaux similaires, avec des réductions comparables de la PIO et du nombre de médicaments antiglaucomateux. Après un an, une réduction modeste de la PIO de 10,83 % a été observée dans le groupe iStent et de 13,65 % dans le groupe XEN-45. Cependant, l’iStent inject® W a démontré un meilleur profil de sécurité peropératoire et postopératoire.
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
求助全文
约1分钟内获得全文 去求助
来源期刊
CiteScore
1.10
自引率
8.30%
发文量
317
审稿时长
49 days
期刊介绍: The Journal français d''ophtalmologie, official publication of the French Society of Ophthalmology, serves the French Speaking Community by publishing excellent research articles, communications of the French Society of Ophthalmology, in-depth reviews, position papers, letters received by the editor and a rich image bank in each issue. The scientific quality is guaranteed through unbiased peer-review, and the journal is member of the Committee of Publication Ethics (COPE). The editors strongly discourage editorial misconduct and in particular if duplicative text from published sources is identified without proper citation, the submission will not be considered for peer review and returned to the authors or immediately rejected.
期刊最新文献
Long-term surgical outcomes of pediatric retinal detachment associated with primary congenital glaucoma Comparison of ChatGPT-4o, Google Gemini 1.5 Pro, Microsoft Copilot Pro, and Ophthalmologists in the management of uveitis and ocular inflammation: A comparative study of large language models Clinical validity of automated refraction with the Chronos binocular refraction system compared to standard refraction techniques Performances visuelles après implantation bilatérale de lentilles intraoculaires à profondeur de champ étendue AE2UV/ZOE en mini-monovision Capsulotomie circulaire ou radiaire pour le traitement du capsulophimosis : quelle serait la meilleure option ?
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1