Performance evaluation of Rayto RT-7600Vet hematology analyzer in side-by-side comparison with manual hematological methods for apparently healthy Cholistani cattle blood.

IF 2.9 3区 综合性期刊 Q1 MULTIDISCIPLINARY SCIENCES PLoS ONE Pub Date : 2025-03-11 eCollection Date: 2025-01-01 DOI:10.1371/journal.pone.0302617
Umer Farooq, Mushtaq Hussain Lashari, Zia Ur Rehman, Musadiq Idris, Haroon Rashid, Shagufta Nasreen, Farah Laraib, Rubaisha Ameer, Maryam Chauhdary, Iram Fatima
{"title":"Performance evaluation of Rayto RT-7600Vet hematology analyzer in side-by-side comparison with manual hematological methods for apparently healthy Cholistani cattle blood.","authors":"Umer Farooq, Mushtaq Hussain Lashari, Zia Ur Rehman, Musadiq Idris, Haroon Rashid, Shagufta Nasreen, Farah Laraib, Rubaisha Ameer, Maryam Chauhdary, Iram Fatima","doi":"10.1371/journal.pone.0302617","DOIUrl":null,"url":null,"abstract":"<p><p>The present study is the first from Pakistan being reported with an objective to assess performance of Rayto RT-7600Vet hematology analyzer (HA) for Cholistani cattle blood (n = 134), in comparison to the manual hematological methods. The four hematological attributes viz. total erythrocyte count (TEC), hemoglobin (Hb), packed cell volume (PCV) and platelet count (PLT) were deduced through HA (A) and manual (M) methods. Various statistical tests were implied to ascertain level of interrelationship, accuracy and level of agreement between the two methods. All attributes attained through manual methods had high positive, leptokurtic distribution (having many outliers) except for PLT-M and PCV-M. The coefficient of variation for attributes attained through HA and manual methods ranged from 16-24% and from 16-59%, respectively. Comparison between the overall results revealed that all the studied attributes, except TEC, were significantly (P≤0.05) different for both methods. A weak relationship was noticed between the attributes attained through two methods as indicated by weak r-values and adjusted r-square values. The reliability level of estimating Hb and PCV had highest intraclass correlation coefficient value of 0.722 and 0.555 for average measures, respectively. However, accuracy level, as determined through Lin's concordance correlation coefficient was highest for TEC (0.9504) for both analytical methods. Poor level of agreement, in general, was shown for the two methods of analysis regarding all four hematological attributes through Bland and Altman test. In conclusion, the Rayto RT-7600Vet) may present data having higher skewness, kurtosis, and CV%, however, they are valid for multi-species hematological analysis. Caution must however, be taken in interpreting their results with corrected reference intervals and CV% for each machine and for each tested attribute.</p>","PeriodicalId":20189,"journal":{"name":"PLoS ONE","volume":"20 3","pages":"e0302617"},"PeriodicalIF":2.9000,"publicationDate":"2025-03-11","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"PLoS ONE","FirstCategoryId":"103","ListUrlMain":"https://doi.org/10.1371/journal.pone.0302617","RegionNum":3,"RegionCategory":"综合性期刊","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/1/1 0:00:00","PubModel":"eCollection","JCR":"Q1","JCRName":"MULTIDISCIPLINARY SCIENCES","Score":null,"Total":0}
引用次数: 0

Abstract

The present study is the first from Pakistan being reported with an objective to assess performance of Rayto RT-7600Vet hematology analyzer (HA) for Cholistani cattle blood (n = 134), in comparison to the manual hematological methods. The four hematological attributes viz. total erythrocyte count (TEC), hemoglobin (Hb), packed cell volume (PCV) and platelet count (PLT) were deduced through HA (A) and manual (M) methods. Various statistical tests were implied to ascertain level of interrelationship, accuracy and level of agreement between the two methods. All attributes attained through manual methods had high positive, leptokurtic distribution (having many outliers) except for PLT-M and PCV-M. The coefficient of variation for attributes attained through HA and manual methods ranged from 16-24% and from 16-59%, respectively. Comparison between the overall results revealed that all the studied attributes, except TEC, were significantly (P≤0.05) different for both methods. A weak relationship was noticed between the attributes attained through two methods as indicated by weak r-values and adjusted r-square values. The reliability level of estimating Hb and PCV had highest intraclass correlation coefficient value of 0.722 and 0.555 for average measures, respectively. However, accuracy level, as determined through Lin's concordance correlation coefficient was highest for TEC (0.9504) for both analytical methods. Poor level of agreement, in general, was shown for the two methods of analysis regarding all four hematological attributes through Bland and Altman test. In conclusion, the Rayto RT-7600Vet) may present data having higher skewness, kurtosis, and CV%, however, they are valid for multi-species hematological analysis. Caution must however, be taken in interpreting their results with corrected reference intervals and CV% for each machine and for each tested attribute.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
求助全文
约1分钟内获得全文 去求助
来源期刊
PLoS ONE
PLoS ONE 生物-生物学
CiteScore
6.20
自引率
5.40%
发文量
14242
审稿时长
3.7 months
期刊介绍: PLOS ONE is an international, peer-reviewed, open-access, online publication. PLOS ONE welcomes reports on primary research from any scientific discipline. It provides: * Open-access—freely accessible online, authors retain copyright * Fast publication times * Peer review by expert, practicing researchers * Post-publication tools to indicate quality and impact * Community-based dialogue on articles * Worldwide media coverage
期刊最新文献
Conditioned medium of engineering macrophages combined with soluble microneedles promote diabetic wound healing. Comparison between Go-GutDx, a novel diagnostic stool test kit with potential impact in low-income countries, and BioFire test. The causal effects of inflammatory bowel disease on its ocular manifestations: A Mendelian randomization study. Coping with chronic periprosthetic joint infection after failed revision of total knee and hip arthroplasty: a qualitative study on patient's experiences in treatment and healing. Correction: LightGBM hybrid model based DEM correction for forested areas.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1