Accuracy of the Freestyle Libre 3 Continuous Glucose Monitoring System in Hypo- and Euglycemic Cats

IF 2.1 2区 农林科学 Q1 VETERINARY SCIENCES Journal of Veterinary Internal Medicine Pub Date : 2025-03-19 DOI:10.1111/jvim.70048
Antonio M. Tardo, Chiquitha Crews, Jocelyn Mott, Lauren T. Porter, Christopher Adin, Chen Gilor
{"title":"Accuracy of the Freestyle Libre 3 Continuous Glucose Monitoring System in Hypo- and Euglycemic Cats","authors":"Antonio M. Tardo,&nbsp;Chiquitha Crews,&nbsp;Jocelyn Mott,&nbsp;Lauren T. Porter,&nbsp;Christopher Adin,&nbsp;Chen Gilor","doi":"10.1111/jvim.70048","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Background</h3>\n \n <p>The FreeStyle Libre 3 (FSL3) has several improvements compared to previous FreeStyle Libre systems, but its accuracy has not yet been determined in cats. In diabetic people, FSL3 offers increased accuracy, and its smaller size could be advantageous for use in veterinary patients.</p>\n </section>\n \n <section>\n \n <h3> Objectives</h3>\n \n <p>Assess the accuracy of FSL3 in cats with experimentally induced hypoglycemia.</p>\n </section>\n \n <section>\n \n <h3> Animals</h3>\n \n <p>Seven healthy, purpose-bred cats.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>Hyperinsulinemic-hypoglycemic clamps were performed. Interstitial glucose concentration (IG), measured by FSL3, was compared to blood glucose concentration (BG) measured by AlphaTrak2. Data were analyzed for all paired measurements (<i>n</i> = 474) and during stable BG (≤ 1 mg/dL/min change over 10 min). Pearson's r test, Bland–Altman test, and Parkes Error Grid analysis (EGA) respectively used to determine correlation, bias, and clinical accuracy.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>Blood glucose concentration and IG correlated strongly (<i>r</i> = 0.86, <i>p</i> &lt; 0.0001) in stable glycemia and moderately at all rates of change (<i>r</i> = 0.73, <i>p</i> &lt; 0.0001). Analytical accuracy was not achieved, whereas clinical accuracy was demonstrated with 99%–100% of the results in zones A + B of the Parkes EGA. Interstitial glucose concentration underestimated BG in euglycemia and mild hypoglycemia (mean −11.7 ± 11.2, −5.5 ± 9.1, −1.5 ± 6.0 mg/dL in the ranges 91–120, 66–90, and 56–65 mg/dL, respectively), but overestimated BG in marked hypoglycemia (mean 6.3 ± 5.7, 15.7 ± 5.6 mg/dL in the ranges 46–55 and &lt; 45 mg/dL, respectively).</p>\n </section>\n \n <section>\n \n <h3> Conclusions</h3>\n \n <p>The FSL3 underestimates BG across most of the hypo-euglycemic range but overestimates BG in marked hypoglycemia (&lt; 55 mg/dL). Recognizing the proportional, glycemic-dependent bias of FSL3 improves the safety of its clinical application in cats.</p>\n </section>\n </div>","PeriodicalId":49958,"journal":{"name":"Journal of Veterinary Internal Medicine","volume":"39 2","pages":""},"PeriodicalIF":2.1000,"publicationDate":"2025-03-19","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1111/jvim.70048","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Veterinary Internal Medicine","FirstCategoryId":"97","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1111/jvim.70048","RegionNum":2,"RegionCategory":"农林科学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"VETERINARY SCIENCES","Score":null,"Total":0}
引用次数: 0

Abstract

Background

The FreeStyle Libre 3 (FSL3) has several improvements compared to previous FreeStyle Libre systems, but its accuracy has not yet been determined in cats. In diabetic people, FSL3 offers increased accuracy, and its smaller size could be advantageous for use in veterinary patients.

Objectives

Assess the accuracy of FSL3 in cats with experimentally induced hypoglycemia.

Animals

Seven healthy, purpose-bred cats.

Methods

Hyperinsulinemic-hypoglycemic clamps were performed. Interstitial glucose concentration (IG), measured by FSL3, was compared to blood glucose concentration (BG) measured by AlphaTrak2. Data were analyzed for all paired measurements (n = 474) and during stable BG (≤ 1 mg/dL/min change over 10 min). Pearson's r test, Bland–Altman test, and Parkes Error Grid analysis (EGA) respectively used to determine correlation, bias, and clinical accuracy.

Results

Blood glucose concentration and IG correlated strongly (r = 0.86, p < 0.0001) in stable glycemia and moderately at all rates of change (r = 0.73, p < 0.0001). Analytical accuracy was not achieved, whereas clinical accuracy was demonstrated with 99%–100% of the results in zones A + B of the Parkes EGA. Interstitial glucose concentration underestimated BG in euglycemia and mild hypoglycemia (mean −11.7 ± 11.2, −5.5 ± 9.1, −1.5 ± 6.0 mg/dL in the ranges 91–120, 66–90, and 56–65 mg/dL, respectively), but overestimated BG in marked hypoglycemia (mean 6.3 ± 5.7, 15.7 ± 5.6 mg/dL in the ranges 46–55 and < 45 mg/dL, respectively).

Conclusions

The FSL3 underestimates BG across most of the hypo-euglycemic range but overestimates BG in marked hypoglycemia (< 55 mg/dL). Recognizing the proportional, glycemic-dependent bias of FSL3 improves the safety of its clinical application in cats.

Abstract Image

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
求助全文
约1分钟内获得全文 去求助
来源期刊
CiteScore
4.50
自引率
11.50%
发文量
243
审稿时长
22 weeks
期刊介绍: The mission of the Journal of Veterinary Internal Medicine is to advance veterinary medical knowledge and improve the lives of animals by publication of authoritative scientific articles of animal diseases.
期刊最新文献
Weekly Biological Variation of Urine Protein Creatinine Ratio and Urine Specific Gravity in Healthy Dogs Accuracy of the Freestyle Libre 3 Continuous Glucose Monitoring System in Hypo- and Euglycemic Cats Assessment of Exenatide Extended-Release for Maintenance of Diabetic Remission in Cats Apoptosis Inhibitor of Macrophage (CD5-Like Antigen) in Healthy Dogs and Dogs With Acute Kidney Injury Facial Myokymia With or Without Concurrent Neurological Deficits in Seven Dogs and Two Cats
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1