{"title":"Comparison of optical performances of intraocular lenses with different depth of field","authors":"Federico Tommasi ,&nbsp;Alessio Giorgetti ,&nbsp;Caterina Monnetti ,&nbsp;Lorenzo Fini ,&nbsp;Giovanni Romualdi ,&nbsp;Rita Mencucci ,&nbsp;Massimo Gurioli ,&nbsp;Stefano Cavalieri","doi":"10.1016/j.optom.2025.100546","DOIUrl":null,"url":null,"abstract":"<div><h3>Purpose</h3><div>This study presents a detailed optical characterization of three intraocular lenses (IOLs) comparing two so called “enhanced range of field”, Evolux IOL and Tecnis Eyhance IOL and one so called “narrow range of field”, a standard monofocal IOL Acrysof IQ.</div></div><div><h3>Methods</h3><div>The measurements are performed using the PMTF optical bench, basing on the Modulation Transfer Function (MTF) to evaluate the optical performance of each lens. The MTFa, representing the area under the MTF curve, is utilized as a key and synthetic metric to quantify performance across different spatial frequencies, providing insights into the IOLs’ behavior as defocus varies.</div></div><div><h3>Results</h3><div>The results highlight the strengths and the weaknesses of the IOLs, with the Evolux showing a broadest depth of focus. The USAF resolution target is used to qualitatively assess the image reproduction at far, intermediate, and near distances, offering a visual representation of the IOLs’ capabilities.</div></div><div><h3>Conclusions</h3><div>These findings provide valuable information for selecting IOLs based on patient-specific visual requirements in cataract and refractive surgery.</div></div>","PeriodicalId":46407,"journal":{"name":"Journal of Optometry","volume":"18 2","pages":"Article 100546"},"PeriodicalIF":1.8000,"publicationDate":"2025-03-29","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Optometry","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1888429625000123","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose

This study presents a detailed optical characterization of three intraocular lenses (IOLs) comparing two so called “enhanced range of field”, Evolux IOL and Tecnis Eyhance IOL and one so called “narrow range of field”, a standard monofocal IOL Acrysof IQ.

Methods

The measurements are performed using the PMTF optical bench, basing on the Modulation Transfer Function (MTF) to evaluate the optical performance of each lens. The MTFa, representing the area under the MTF curve, is utilized as a key and synthetic metric to quantify performance across different spatial frequencies, providing insights into the IOLs’ behavior as defocus varies.

Results

The results highlight the strengths and the weaknesses of the IOLs, with the Evolux showing a broadest depth of focus. The USAF resolution target is used to qualitatively assess the image reproduction at far, intermediate, and near distances, offering a visual representation of the IOLs’ capabilities.

Conclusions

These findings provide valuable information for selecting IOLs based on patient-specific visual requirements in cataract and refractive surgery.
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
不同景深人工晶体光学性能的比较
本研究详细介绍了三种人工晶体(IOL)的光学特性,比较了两种所谓的“增强视野范围”(Evolux IOL)和Tecnis Eyhance IOL,以及一种所谓的“窄视野范围”(narrow range of field),一种标准的单焦点IOL (Acrysof IQ)。方法采用PMTF光学实验台进行测量,基于调制传递函数(MTF)评价各透镜的光学性能。MTFa,表示MTF曲线下的面积,是一个关键的综合指标,用于量化不同空间频率下的性能,从而深入了解离焦变化时iol的行为。结果结果突出了人工晶状体的优点和缺点,其中Evolux显示出最广泛的聚焦深度。USAF分辨率目标用于定性地评估远、中、近距离的图像再现,提供人工晶状体能力的视觉表现。结论本研究结果为在白内障和屈光手术中根据患者特定视力要求选择人工晶状体提供了有价值的信息。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Journal of Optometry
Journal of Optometry OPHTHALMOLOGY-
CiteScore
5.20
自引率
0.00%
发文量
60
审稿时长
66 days
期刊最新文献
Standardization of distance visual acuity measurement using optonet computerized vision charts: A multicentric study. Home-based perceptual learning augments high-frequency contrast sensitivity and stereopsis after esotropia surgery: A retrospective cohort study. Transient ocular effects of red light exposure in young adults. Analysis of patient adherence to emerging treatment tools for improving visual functions in amblyopia: A systematic review and meta-analysis. Diurnal variation in corneo-scleral morphology.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1