{"title":"False negative interpretation of biologic indicators or the importance of using two biologic indicator test strips.","authors":"J R Scoville","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>The use of at least two biologic indicators to monitor a sterilization process greatly enhances failure detection. A retrospective study of a sterilization monitoring service's database revealed that 48.7% of tests using two test strips were single-strip positive. This indicates that there is a nonuniformity in the sterilization conditions within the sterilizer chamber and that there is a need for at least two test strips to detect this nonuniformity.</p>","PeriodicalId":55234,"journal":{"name":"Compendium-Continuing Education for Veterinarians","volume":"15 12","pages":"1472-4, 1476, 1478; quiz 1478"},"PeriodicalIF":0.0000,"publicationDate":"1994-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Compendium-Continuing Education for Veterinarians","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
The use of at least two biologic indicators to monitor a sterilization process greatly enhances failure detection. A retrospective study of a sterilization monitoring service's database revealed that 48.7% of tests using two test strips were single-strip positive. This indicates that there is a nonuniformity in the sterilization conditions within the sterilizer chamber and that there is a need for at least two test strips to detect this nonuniformity.