{"title":"Composition, constitution, and interaction of bone with hydroxyapatite coatings determined by FT Raman microscopy","authors":"Bernd Dippel, Reinhold T. Mueller, Andreas Pingsmann, Bernhard Schrader","doi":"10.1002/(SICI)1520-6343(1998)4:6<403::AID-BSPY5>3.0.CO;2-M","DOIUrl":null,"url":null,"abstract":"<p>An optimized FT Raman microscope (inverted microscope with high throughput of radiation) was developed that allows minimal sample preparation and Raman spectroscopy without fluorescence. A quantitative determination of the mineralization of bone tissue and hydroxyapatite (HA) coatings of hip and knee prostheses was performed. The lateral resolution reached down to 10 μm. The distribution of the HA content in the coatings investigated was found to be similar all the time. This result was independent of the composition of the coatings and the history of the whole prosthesis. In the immediate vicinity of the prosthesis a large HA content could be observed that decreased to a minimum towards the periphery of the coating and increased at the site of the ongrown bone. For the interface between bone and HA coating a transitional zone was observed at a lateral distance of 30–40 μm to the implant. © 1998 John Wiley & Sons, Inc. Biospectroscopy 4: 403–412, 1998</p>","PeriodicalId":9037,"journal":{"name":"Biospectroscopy","volume":"4 6","pages":"403-412"},"PeriodicalIF":0.0000,"publicationDate":"1999-01-06","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1002/(SICI)1520-6343(1998)4:6<403::AID-BSPY5>3.0.CO;2-M","citationCount":"19","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Biospectroscopy","FirstCategoryId":"1085","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/%28SICI%291520-6343%281998%294%3A6%3C403%3A%3AAID-BSPY5%3E3.0.CO%3B2-M","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 19
Abstract
An optimized FT Raman microscope (inverted microscope with high throughput of radiation) was developed that allows minimal sample preparation and Raman spectroscopy without fluorescence. A quantitative determination of the mineralization of bone tissue and hydroxyapatite (HA) coatings of hip and knee prostheses was performed. The lateral resolution reached down to 10 μm. The distribution of the HA content in the coatings investigated was found to be similar all the time. This result was independent of the composition of the coatings and the history of the whole prosthesis. In the immediate vicinity of the prosthesis a large HA content could be observed that decreased to a minimum towards the periphery of the coating and increased at the site of the ongrown bone. For the interface between bone and HA coating a transitional zone was observed at a lateral distance of 30–40 μm to the implant. © 1998 John Wiley & Sons, Inc. Biospectroscopy 4: 403–412, 1998
用FT拉曼显微镜测定骨与羟基磷灰石涂层的组成、构成和相互作用
开发了一种优化的FT拉曼显微镜(具有高通量辐射的倒置显微镜),允许最少的样品制备和无荧光的拉曼光谱。定量测定髋关节和膝关节假体骨组织和羟基磷灰石(HA)涂层的矿化程度。横向分辨率可达10 μm。结果表明,所测涂料中HA含量的分布基本一致。这一结果与涂层的成分和整个假体的历史无关。在假体附近,可以观察到大量的HA含量,在涂层周围减少到最小,在长骨部位增加。在骨与羟基磷灰石涂层的界面处,在距种植体30-40 μm的横向距离处观察到一个过渡区。©1998 John Wiley &儿子,Inc。生物光谱学杂志,2004,31 (4):393 - 398
本文章由计算机程序翻译,如有差异,请以英文原文为准。