Correlation analysis between radiation exposure and the image quality of cone-beam computed tomography in the dental clinical environment.

IF 1.7 Q3 DENTISTRY, ORAL SURGERY & MEDICINE Imaging Science in Dentistry Pub Date : 2022-09-01 Epub Date: 2022-05-13 DOI:10.5624/isd.20220129
Chang-Ho Song, Han-Gyeol Yeom, Jo-Eun Kim, Kyung-Hoe Huh, Won-Jin Yi, Min-Suk Heo, Sam-Sun Lee
{"title":"Correlation analysis between radiation exposure and the image quality of cone-beam computed tomography in the dental clinical environment.","authors":"Chang-Ho Song,&nbsp;Han-Gyeol Yeom,&nbsp;Jo-Eun Kim,&nbsp;Kyung-Hoe Huh,&nbsp;Won-Jin Yi,&nbsp;Min-Suk Heo,&nbsp;Sam-Sun Lee","doi":"10.5624/isd.20220129","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>This study was conducted to measure the radiation exposure and image quality of various cone-beam computed tomography (CBCT) machines under common clinical conditions and to analyze the correlation between them.</p><p><strong>Materials and methods: </strong>Seven CBCT machines used frequently in clinical practice were selected. Because each machine has various sizes of fields of view (FOVs), 1 large FOV and 1 small FOV were selected for each machine. Radiation exposure was measured using a dose-area product (DAP) meter. The quality of the CBCT images was analyzed using 8 image quality parameters obtained using a dental volume tomography phantom. For statistical analysis, regression analysis using a generalized linear model was used.</p><p><strong>Results: </strong>Polymethyl-methacrylate (PMMA) noise and modulation transfer function (MTF) 10% showed statistically significant correlations with DAP values, presenting positive and negative correlations, respectively (P<0.05). Image quality parameters other than PMMA noise and MTF 10% did not demonstrate statistically significant correlations with DAP values.</p><p><strong>Conclusion: </strong>As radiation exposure and image quality are not proportionally related in clinically used equipment, it is necessary to evaluate and monitor radiation exposure and image quality separately.</p>","PeriodicalId":51714,"journal":{"name":"Imaging Science in Dentistry","volume":null,"pages":null},"PeriodicalIF":1.7000,"publicationDate":"2022-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://ftp.ncbi.nlm.nih.gov/pub/pmc/oa_pdf/a2/53/isd-52-283.PMC9530299.pdf","citationCount":"2","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Imaging Science in Dentistry","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.5624/isd.20220129","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2022/5/13 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 2

Abstract

Purpose: This study was conducted to measure the radiation exposure and image quality of various cone-beam computed tomography (CBCT) machines under common clinical conditions and to analyze the correlation between them.

Materials and methods: Seven CBCT machines used frequently in clinical practice were selected. Because each machine has various sizes of fields of view (FOVs), 1 large FOV and 1 small FOV were selected for each machine. Radiation exposure was measured using a dose-area product (DAP) meter. The quality of the CBCT images was analyzed using 8 image quality parameters obtained using a dental volume tomography phantom. For statistical analysis, regression analysis using a generalized linear model was used.

Results: Polymethyl-methacrylate (PMMA) noise and modulation transfer function (MTF) 10% showed statistically significant correlations with DAP values, presenting positive and negative correlations, respectively (P<0.05). Image quality parameters other than PMMA noise and MTF 10% did not demonstrate statistically significant correlations with DAP values.

Conclusion: As radiation exposure and image quality are not proportionally related in clinically used equipment, it is necessary to evaluate and monitor radiation exposure and image quality separately.

Abstract Image

Abstract Image

Abstract Image

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
牙科临床环境中放射照射与锥束计算机断层成像质量的相关性分析。
目的:测量临床常用条件下各种锥形束ct (cone-beam computed tomography, CBCT)机器的辐射暴露和图像质量,并分析两者之间的相关性。材料与方法:选取临床常用的7台CBCT机。由于每台机器有不同大小的视场(FOV),所以每台机器选择1个大FOV和1个小FOV。使用剂量面积积(DAP)计测量辐射暴露。利用牙体容积断层成像体获得的8个图像质量参数对CBCT图像质量进行分析。统计分析采用广义线性模型进行回归分析。结果:聚甲基丙烯酸甲酯(PMMA)噪声和调制传递函数(MTF) 10%与DAP值具有统计学意义,分别呈正相关和负相关(p)结论:在临床使用的设备中,由于辐射暴露与图像质量不成比例相关,有必要分别对辐射暴露和图像质量进行评估和监测。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Imaging Science in Dentistry
Imaging Science in Dentistry DENTISTRY, ORAL SURGERY & MEDICINE-
CiteScore
2.90
自引率
11.10%
发文量
42
期刊最新文献
Impact of the spatial orientation of the patient's head, metal artifact reduction, and tube current on cone-beam computed tomography artifact expression adjacent to a dental implant: A laboratory study using a simulated surgical guide. Optimizing cone-beam computed tomography exposure for an effective radiation dose and image quality balance. Temporomandibular joint ankylosis suspected to be associated with ankylosing spondylitis based on cervical computed tomography images: A pictorial essay. Vertical root fracture diagnosis in teeth with metallic posts: Impact of metal artifact reduction and sharpening filters. A novel method for determining dose distribution on panoramic reconstruction computed tomography images from radiotherapy computed tomography.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1