Characterization of Effective Orifice Areas of Mitral Prosthetic Heart Valves: An In-Vitro Study.

Q3 Medicine Journal of Heart Valve Disease Pub Date : 2017-11-01
Morgane Evin, Julien Magne, Stuart M Grieve, Régis Rieu, Philipe Pibarot
{"title":"Characterization of Effective Orifice Areas of Mitral Prosthetic Heart Valves: An In-Vitro Study.","authors":"Morgane Evin,&nbsp;Julien Magne,&nbsp;Stuart M Grieve,&nbsp;Régis Rieu,&nbsp;Philipe Pibarot","doi":"","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Reference values of hemodynamic parameters for the assessment of prosthetic heart valves are necessary, and ideally need to be provided by entities independent of the valve manufacturers. Thus, the study aim was to provide, in vitro, normal reference values of the effective orifice area (EOA) for different models and sizes of mitral prosthetic valve, and to assess the determinants of EOA and mean transvalvular pressure gradient (mTPG).</p><p><strong>Methods: </strong>Four models of mechanical prostheses were tested (one mono-leaflet, three bi-leaflet) and four models of bioprostheses (two bovine pericardial, two porcine) on a double-activation pulsed duplicator that was specifically designed and optimized for assessing the hemodynamic performance of mitral prosthetic valves. The hemodynamic conditions were standardized and included for bioprostheses: two mitral flow volumes, three mean aortic pressures, two heart rates, and three E/A ratios. The EOAs were measured with Doppler echocardiography, using the same method (continuity equation) as was used in the clinical setting. Overestimation in term of EOA was defined according to guidelines as >0.25 cm2.</p><p><strong>Results: </strong>EOA reference values were recorded. For mono-leaflet prostheses (Medtronic Hall 7700, size 25 to 31 mm) 2.29 and 3.49; for bi-leaflet prostheses (St. Jude Medical Master and Master HP, sizes 25 to 33 mm and On-X valve, sizes 27-29 mm) 1.34 and 4.74 cm2; for porcine bioprostheses (Medtronic Mosaic CINCH, sizes 25 to 31 mm and St. Jude Epic 100, sizes 25 to 33 mm) 1.35 and 3.56 cm2; for bovine pericardial bioprosthetic valves (Edwards Perimount 6900P and Magna Ease 7300, sizes 25 to 33 mm) 1.67 and 2.36 cm2. There were some discrepancies between the normal reference EOAs measured compared to those provided by the prosthesis manufacturers, or in published reports. The bioprosthetic EOAs were shown to be smaller than the manufacturers' values in 32% of valves (by an average of 0.57 ± 0.28 cm2) versus in 7% of valves when compared to values reported elsewhere (by an average of 0.43 ± 0.17 cm2). The relationship between EOA and internal orifice area (IOA) varied according to the type of prosthesis. The EOA was close to the IOA in mechanical valves (regression slopes 0.87-0.99) but was much smaller than the IOA in bioprosthetic valves (slopes 0.25-0.30). The EOA was influenced by prosthesis diameter, prosthesis stent diameter and height, while the mTPG was influenced by EOA and heart rate.</p><p><strong>Conclusions: </strong>The present study has provided normal reference values of EOAs for several frequently used mitral prostheses. This information may be helpful for identifying and quantifying prosthetic valve dysfunction and prosthesis-patient mismatch.</p>","PeriodicalId":50184,"journal":{"name":"Journal of Heart Valve Disease","volume":"26 6","pages":"677-687"},"PeriodicalIF":0.0000,"publicationDate":"2017-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Heart Valve Disease","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0

Abstract

Background: Reference values of hemodynamic parameters for the assessment of prosthetic heart valves are necessary, and ideally need to be provided by entities independent of the valve manufacturers. Thus, the study aim was to provide, in vitro, normal reference values of the effective orifice area (EOA) for different models and sizes of mitral prosthetic valve, and to assess the determinants of EOA and mean transvalvular pressure gradient (mTPG).

Methods: Four models of mechanical prostheses were tested (one mono-leaflet, three bi-leaflet) and four models of bioprostheses (two bovine pericardial, two porcine) on a double-activation pulsed duplicator that was specifically designed and optimized for assessing the hemodynamic performance of mitral prosthetic valves. The hemodynamic conditions were standardized and included for bioprostheses: two mitral flow volumes, three mean aortic pressures, two heart rates, and three E/A ratios. The EOAs were measured with Doppler echocardiography, using the same method (continuity equation) as was used in the clinical setting. Overestimation in term of EOA was defined according to guidelines as >0.25 cm2.

Results: EOA reference values were recorded. For mono-leaflet prostheses (Medtronic Hall 7700, size 25 to 31 mm) 2.29 and 3.49; for bi-leaflet prostheses (St. Jude Medical Master and Master HP, sizes 25 to 33 mm and On-X valve, sizes 27-29 mm) 1.34 and 4.74 cm2; for porcine bioprostheses (Medtronic Mosaic CINCH, sizes 25 to 31 mm and St. Jude Epic 100, sizes 25 to 33 mm) 1.35 and 3.56 cm2; for bovine pericardial bioprosthetic valves (Edwards Perimount 6900P and Magna Ease 7300, sizes 25 to 33 mm) 1.67 and 2.36 cm2. There were some discrepancies between the normal reference EOAs measured compared to those provided by the prosthesis manufacturers, or in published reports. The bioprosthetic EOAs were shown to be smaller than the manufacturers' values in 32% of valves (by an average of 0.57 ± 0.28 cm2) versus in 7% of valves when compared to values reported elsewhere (by an average of 0.43 ± 0.17 cm2). The relationship between EOA and internal orifice area (IOA) varied according to the type of prosthesis. The EOA was close to the IOA in mechanical valves (regression slopes 0.87-0.99) but was much smaller than the IOA in bioprosthetic valves (slopes 0.25-0.30). The EOA was influenced by prosthesis diameter, prosthesis stent diameter and height, while the mTPG was influenced by EOA and heart rate.

Conclusions: The present study has provided normal reference values of EOAs for several frequently used mitral prostheses. This information may be helpful for identifying and quantifying prosthetic valve dysfunction and prosthesis-patient mismatch.

分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
二尖瓣人工心脏瓣膜有效孔面积的表征:一项体外研究。
背景:评估人工心脏瓣膜的血流动力学参数参考值是必要的,理想情况下需要由独立于瓣膜制造商的实体提供。因此,本研究旨在为不同型号和尺寸的二尖瓣人工瓣膜提供体外正常的有效孔面积(EOA)参考值,并评估影响EOA和平均跨瓣压力梯度(mTPG)的因素。方法:采用专为二尖瓣人工瓣膜血流动力学性能评估设计和优化的双激活脉冲复制仪,对4个机械假体模型(1个单叶、3个双叶)和4个生物假体模型(2个牛心包、2个猪)进行测试。血液动力学条件标准化并纳入生物假体:2个二尖瓣血流容量,3个平均主动脉压,2个心率和3个E/A比。用多普勒超声心动图测量eoa,采用与临床相同的方法(连续性方程)。根据指南,EOA的高估定义为>0.25 cm2。结果:记录了EOA参考值。单叶假体(Medtronic Hall 7700,尺寸25至31 mm) 2.29和3.49;双叶假体(St. Jude Medical Master和Master HP,尺寸25 - 33毫米,On-X瓣膜尺寸27-29毫米)1.34和4.74平方厘米;猪生物假体(美敦力Mosaic CINCH,尺寸25至31毫米,St. Jude Epic 100,尺寸25至33毫米)1.35和3.56平方厘米;用于牛心包生物假体瓣膜(Edwards perimont 6900P和Magna Ease 7300,尺寸25至33 mm) 1.67和2.36 cm2。测量的正常参考eoa与假体制造商提供的或发表的报告之间存在一些差异。在32%的瓣膜中,生物假体的EOAs比制造商的值小(平均0.57±0.28 cm2),而在其他地方报道的瓣膜中,只有7%的瓣膜比制造商的值小(平均0.43±0.17 cm2)。内孔面积(IOA)与内孔面积(IOA)的关系因假体类型的不同而不同。机械瓣膜的IOA接近IOA(回归斜率0.87 ~ 0.99),但远小于生物假体瓣膜的IOA(斜率0.25 ~ 0.30)。假体直径、假体支架直径和高度影响EOA, mTPG受EOA和心率影响。结论:本研究为几种常用的二尖瓣假体提供了正常的EOAs参考值。这一信息可能有助于识别和量化假体瓣膜功能障碍和假体-患者不匹配。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Journal of Heart Valve Disease
Journal of Heart Valve Disease 医学-心血管系统
CiteScore
1.00
自引率
0.00%
发文量
0
审稿时长
4-8 weeks
期刊介绍: The Journal of Heart Valve Disease (ISSN 0966-8519) is the official journal of The Society for Heart Valve Disease. It is indexed/abstracted by Index Medicus, Medline, Medlar, PubMed, Science Citation Index, Scisearch, Research Alert, Biomedical Products, Current Contents/Clinical Medicine. It is issued bi-monthly in one indexed volume by ICR Publishers Ltd., Crispin House, 12A South Approach, Moor Park, Northwood HA6 2ET, United Kingdom. This paper meets the requirements of ANSI standard Z39.48-1992 (Permanence of Paper).
期刊最新文献
ABO Blood Group: A Risk Factor for a Cardiovascular Disease in Adults in Morocco Anti-Hypertensive Medication in Geriatric Population with HFpEF Comparison of Clinical Characteristics and Outcomes in Patients with STEMI in Croatia and Kosovo Radiofrequency Ablation of Ventricular Arrhythmias on Bioprosthetic Valve, a Never Found Case Before An Interesting Case of Heart Failure in a Patient with Bicuspid Aortic Valve - Unveiling the Diagnosis of Amyloidosis
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1