Simultaneous Determination of Uranium and Depleted Uranium Isotopic Ratio Using Inductively Coupled Mass Spectrometry
Maria A. Velez-Quinones, Hanna Xu, Nhu Vo, Joanna M. Gaitens, Melissa A. McDiarmid, Michael R. Lewin-Smith, Frederick G. Strathmann
{"title":"Simultaneous Determination of Uranium and Depleted Uranium Isotopic Ratio Using Inductively Coupled Mass Spectrometry","authors":"Maria A. Velez-Quinones, Hanna Xu, Nhu Vo, Joanna M. Gaitens, Melissa A. McDiarmid, Michael R. Lewin-Smith, Frederick G. Strathmann","doi":"10.1002/cptx.59","DOIUrl":null,"url":null,"abstract":"<p>We have developed and validated a method for the simultaneous quantitative measurement of total uranium (TU) and uranium <sup>235</sup>U/<sup>238</sup>U isotopic ratio (UIR) in urine by inductively coupled plasma mass spectrometry (ICP-MS) using a Thermo Scientific iCAP-Q instrument. The performance characteristics of the assay were determined to be in compliance with clinical laboratory standards. The assay was linear in the concentration range of 1.0 to 500.0 ng/liter TU. The method was precise and accurate with limits of detection of 2.5 ng/liter for TU and 9.8 ng/liter for UIR. The accuracy was >93% and the coefficient of variation (% CV) was <5.0% for both TU and UIR. All results were within established guidelines and agreed-upon criteria, and the results fell within the certified range for the reference controls. The method has thus been shown to be effective as a simple, precise, and sensitive analytical technique for testing urine samples. © 2018 by John Wiley & Sons, Inc.</p>","PeriodicalId":72743,"journal":{"name":"Current protocols in toxicology","volume":"78 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2018-10-04","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1002/cptx.59","citationCount":"4","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Current protocols in toxicology","FirstCategoryId":"1085","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/cptx.59","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 4
Abstract
We have developed and validated a method for the simultaneous quantitative measurement of total uranium (TU) and uranium 235U/238U isotopic ratio (UIR) in urine by inductively coupled plasma mass spectrometry (ICP-MS) using a Thermo Scientific iCAP-Q instrument. The performance characteristics of the assay were determined to be in compliance with clinical laboratory standards. The assay was linear in the concentration range of 1.0 to 500.0 ng/liter TU. The method was precise and accurate with limits of detection of 2.5 ng/liter for TU and 9.8 ng/liter for UIR. The accuracy was >93% and the coefficient of variation (% CV) was <5.0% for both TU and UIR. All results were within established guidelines and agreed-upon criteria, and the results fell within the certified range for the reference controls. The method has thus been shown to be effective as a simple, precise, and sensitive analytical technique for testing urine samples. © 2018 by John Wiley & Sons, Inc.
电感耦合质谱法同时测定铀和贫铀同位素比值
我们开发并验证了一种同时定量测量尿液中总铀(TU)和铀235U/238U同位素比(UIR)的方法,该方法采用电感耦合等离子体质谱(ICP-MS),使用Thermo Scientific iCAP-Q仪器。测定的性能特征符合临床实验室标准。本方法在1.0 ~ 500.0 ng/l TU浓度范围内呈线性关系,检测限为TU 2.5 ng/l, UIR 9.8 ng/l。TU和UIR的准确度为93%,变异系数(% CV)为5.0%。所有结果都在既定的指导方针和商定的标准范围内,并且结果落在参考对照的认证范围内。因此,该方法已被证明是一种简单、精确和敏感的尿液样本检测分析技术。©2018 by John Wiley &儿子,Inc。
本文章由计算机程序翻译,如有差异,请以英文原文为准。