Biomechanical Assessment of Three Osteosynthesis Constructs by Periprosthetic Humerus Fractures.

IF 1.2 Q3 ORTHOPEDICS Advances in Orthopedics Pub Date : 2020-10-26 eCollection Date: 2020-01-01 DOI:10.1155/2020/8872419
Afif Harb, Bastian Welke, Emmanouil Liodakis, Sam Razaeian, Dafang Zhang, Christian Krettek, Christof Hurschler, Nael Hawi
{"title":"Biomechanical Assessment of Three Osteosynthesis Constructs by Periprosthetic Humerus Fractures.","authors":"Afif Harb,&nbsp;Bastian Welke,&nbsp;Emmanouil Liodakis,&nbsp;Sam Razaeian,&nbsp;Dafang Zhang,&nbsp;Christian Krettek,&nbsp;Christof Hurschler,&nbsp;Nael Hawi","doi":"10.1155/2020/8872419","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Biomechanical stability assessment of 3 different constructs for proximal fixation of a locking compression plate (LCP) in treating a Worland type C periprosthetic fracture after total shoulder arthroplasty.</p><p><strong>Methods: </strong>27 Worland type C fractures after shoulder arthroplasty in synthetic humeri were treated with 14-hole LCP that is proximally fixed using the following: (1) 1 × 1.5 mm cerclage wires and 2x unicortical-locking screws, (2) 3 × 1.5 mm cerclage wires, or (3) 2x bicortical-locking attachment plates. Torsional stiffness was assessed by applying an internal rotation moment of 5 Nm and then after unloading the specimen, an external rotation moment of 5 Nm at the same rate was applied. Axial stiffness was assessed by applying a 50 N preload, and then applying a cyclic load of 250 N, then increasing the load by 50 N each time, until a maximum axial load of 2500 N was reached or specimen failure occurred.</p><p><strong>Results: </strong>With regard to internal as well as external rotational stiffness, group 1 showed a mean stiffness of 0.37 Nm/deg and 0.57 Nm/deg, respectively, group 2 had a mean stiffness of 0.51 Nm/deg and 0.39 Nm/deg, respectively, while group 3 had a mean stiffness of 1.34 Nm/deg and 1.31 Nm/deg, respectively. Concerning axial stiffness, group 1 showed an average stiffness of 451.0 N/mm, group 2 had a mean stiffness of 737.5 N/mm, whereas group 3 had a mean stiffness of 715.8 N/mm.</p><p><strong>Conclusion: </strong>Group 3 displayed a significantly higher torsional stiffness while a comparable axial stiffness to group 2.</p>","PeriodicalId":7358,"journal":{"name":"Advances in Orthopedics","volume":"2020 ","pages":"8872419"},"PeriodicalIF":1.2000,"publicationDate":"2020-10-26","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1155/2020/8872419","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Advances in Orthopedics","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1155/2020/8872419","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2020/1/1 0:00:00","PubModel":"eCollection","JCR":"Q3","JCRName":"ORTHOPEDICS","Score":null,"Total":0}
引用次数: 1

Abstract

Background: Biomechanical stability assessment of 3 different constructs for proximal fixation of a locking compression plate (LCP) in treating a Worland type C periprosthetic fracture after total shoulder arthroplasty.

Methods: 27 Worland type C fractures after shoulder arthroplasty in synthetic humeri were treated with 14-hole LCP that is proximally fixed using the following: (1) 1 × 1.5 mm cerclage wires and 2x unicortical-locking screws, (2) 3 × 1.5 mm cerclage wires, or (3) 2x bicortical-locking attachment plates. Torsional stiffness was assessed by applying an internal rotation moment of 5 Nm and then after unloading the specimen, an external rotation moment of 5 Nm at the same rate was applied. Axial stiffness was assessed by applying a 50 N preload, and then applying a cyclic load of 250 N, then increasing the load by 50 N each time, until a maximum axial load of 2500 N was reached or specimen failure occurred.

Results: With regard to internal as well as external rotational stiffness, group 1 showed a mean stiffness of 0.37 Nm/deg and 0.57 Nm/deg, respectively, group 2 had a mean stiffness of 0.51 Nm/deg and 0.39 Nm/deg, respectively, while group 3 had a mean stiffness of 1.34 Nm/deg and 1.31 Nm/deg, respectively. Concerning axial stiffness, group 1 showed an average stiffness of 451.0 N/mm, group 2 had a mean stiffness of 737.5 N/mm, whereas group 3 had a mean stiffness of 715.8 N/mm.

Conclusion: Group 3 displayed a significantly higher torsional stiffness while a comparable axial stiffness to group 2.

Abstract Image

Abstract Image

Abstract Image

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
肱骨假体周围骨折三种植骨装置的生物力学评价。
背景:对全肩关节置换术后Worland型C假体周围骨折的近端锁定加压钢板(LCP) 3种不同固定方式的生物力学稳定性进行评估。方法:对27例人工肱骨肩关节置换术后Worland C型骨折采用14孔LCP治疗,近端采用以下方法固定:(1)1 × 1.5 mm环扎针和2枚单皮质锁定螺钉,(2)3 × 1.5 mm环扎针,或(3)2枚双皮质锁定连接钢板。通过施加5 Nm的内旋转力矩来评估扭转刚度,然后在卸载试件后,以相同的速率施加5 Nm的外旋转力矩。轴向刚度通过施加50 N的预载荷,然后施加250 N的循环载荷,然后每次增加50 N的载荷来评估,直到最大轴向载荷达到2500 N或试样发生破坏。结果:在内、外旋转刚度方面,组1的平均刚度分别为0.37 Nm/deg和0.57 Nm/deg,组2的平均刚度分别为0.51 Nm/deg和0.39 Nm/deg,组3的平均刚度分别为1.34 Nm/deg和1.31 Nm/deg。轴向刚度方面,组1平均刚度为451.0 N/mm,组2平均刚度为737.5 N/mm,组3平均刚度为715.8 N/mm。结论:组3表现出明显更高的扭转刚度,而组2的轴向刚度相当。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
CiteScore
2.40
自引率
0.00%
发文量
36
审稿时长
21 weeks
期刊介绍: Advances in Orthopedics is a peer-reviewed, Open Access journal that provides a forum for orthopaedics working on improving the quality of orthopedic health care. The journal publishes original research articles, review articles, and clinical studies related to arthroplasty, hand surgery, limb reconstruction, pediatric orthopaedics, sports medicine, trauma, spinal deformities, and orthopaedic oncology.
期刊最新文献
Addressing Knee Osteoarthritis Pathology Through Platelet-Rich Plasma Treatment: A Comprehensive Review. Is Anterior Cruciate Ligament Reconstruction "Silva Technique" Equal to All-Inside Techniques? A Prospective Single-Center Study: An Alternative ACL All-Inside Reconstruction Technique Using a Tibial Tunnel and Bone Graft. Implant Survival of an Uncemented Modular Femoral Implant in Patients With Severe Femoral Bone Loss and 2-Stage Hip Revision. Regular Nonsteroidal Anti-Inflammatory Drug Use Increases Stress Fracture Risk in the General Population: A Retrospective Case-Control Study. From Etiology to Intervention: A Holistic Review of Bunion Pathophysiology and Care.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1