Late dislocation of the capsular bag-intraocular lens-modified capsular tension ring complex after knotless transscleral suturing using 9-0 polypropylene.
Natalia S Anisimova, Lisa B Arbisser, Natalya F Shilova, Roman V Kirtaev, Daria A Dibina, Boris E Malyugin
{"title":"Late dislocation of the capsular bag-intraocular lens-modified capsular tension ring complex after knotless transscleral suturing using 9-0 polypropylene.","authors":"Natalia S Anisimova, Lisa B Arbisser, Natalya F Shilova, Roman V Kirtaev, Daria A Dibina, Boris E Malyugin","doi":"10.5693/djo.02.2020.04.001","DOIUrl":null,"url":null,"abstract":"<p><p>We report a case of late breakage of a 9-0 polypropylene transscleral suture used for fixation of a dislocated capsular bag-intraocular lens-modified capsular tension ring complex in a 52-year-old woman with Marfan syndrome. Breakage occurred despite use of a cow-hitch technique for external and internal fixation. We believe breakage was caused by the suture chafing on the sharp edges of the modified capsular tension ring eyelet. Cross-sectional analysis of Malyugin-modified capsular tension rings from two different manufacturers revealed a difference with respect to radius of curvature. Suturing intraocular implants with relatively sharp edges may cause suture breakage; further studies are needed to identify the critical parameters for the surface quality of sutured intraocular implants.</p>","PeriodicalId":38112,"journal":{"name":"Digital journal of ophthalmology : DJO","volume":"26 2","pages":"7-16"},"PeriodicalIF":0.0000,"publicationDate":"2020-06-26","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8031962/pdf/djo-19-122.pdf","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Digital journal of ophthalmology : DJO","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.5693/djo.02.2020.04.001","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2021/2/1 0:00:00","PubModel":"eCollection","JCR":"Q3","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0
Abstract
We report a case of late breakage of a 9-0 polypropylene transscleral suture used for fixation of a dislocated capsular bag-intraocular lens-modified capsular tension ring complex in a 52-year-old woman with Marfan syndrome. Breakage occurred despite use of a cow-hitch technique for external and internal fixation. We believe breakage was caused by the suture chafing on the sharp edges of the modified capsular tension ring eyelet. Cross-sectional analysis of Malyugin-modified capsular tension rings from two different manufacturers revealed a difference with respect to radius of curvature. Suturing intraocular implants with relatively sharp edges may cause suture breakage; further studies are needed to identify the critical parameters for the surface quality of sutured intraocular implants.