Post-Cure Polymerization and Depth-of-Cure Behaviors of Dental Bulk-Fill Resin-Based Composites.

Christy I Lee, Minju D Yi, Brandon M Gage, Lisa N Yarbrough, Brian J Kirkwood, Wen Lien
{"title":"Post-Cure Polymerization and Depth-of-Cure Behaviors of Dental Bulk-Fill Resin-Based Composites.","authors":"Christy I Lee,&nbsp;Minju D Yi,&nbsp;Brandon M Gage,&nbsp;Lisa N Yarbrough,&nbsp;Brian J Kirkwood,&nbsp;Wen Lien","doi":"","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>Polymerization for modern dental resin-based composites (RBCs) not only occurs immediately upon light exposure but also continues for another 24 hours, well beyond after light is terminated. However, many questions still remain about the role of polymerization kinetics in optimizing the physical properties of a new RBC type-the bulk-fill.</p><p><strong>Objective: </strong>The aim is to study the post-cure polymerization kinetics of bulk-fill RBCs and to compare their degree of polymeric conversion (DC) and depth-of-cure (DoC) with an incremental-fill, conventional RBC.</p><p><strong>Methods: </strong>Five representative bulk-fill RBCs [Surefil SDR+Stress Decreasing Resin Flow Plus (SDRFP), Tetric EvoCeram Bulk Fill (TECB), Filtek 1 Bulk Fill (F1B), Venus Bulk Fill (VB), and Sonicfill (SF3)] and one conventional RBC [Filtek Supreme Ultra (FSU)] were investigated. The upper surface per RBC specimen was exposed to a light curing unit (Paradigm, 3M-ESPE, irradiance=1221 ± 5 mW/cm2) for 20 seconds. The DC per RBC brand were measured at the bottom surface (specimen Ø=4 mm, thickness=3 mm and 5 mm) as a function of post-curing times using a Fourier transform infrared attenuated total reflection spectrometer. Real-time data recording for post-cure DC began immediately upon light exposure and continued at steady intervals, up to15 min, then again after 24 hours. The DoC of all six RBC brands (n=6 / group) were measured according to ISO-4049. Data were analyzed with nonlinear regression and analysis of variances (ANOVA)/Tukey (α=0.05).</p><p><strong>Results: </strong>Mean DC for the six RBCs with 5 mm curing height after 24 hours were: TECB=79.5%, VB=75.7%, SDRFP=69.2%, SF3=65.8%, F1B=51.8%, and FSU=44.0%. Bulk-fill RBCs showed higher DC efficiency than the conventional RBC for both the 3 mm and 5 mm curing heights. Significant differences in DoC were found amongst the six RBC brands: VB=5.1 mm, SDRFP=4.6 mm, F1B=3.8 mm, TECB=3.5 mm, FSU=3.0 mm, and SF3=2.7 mm.</p><p><strong>Conclusion: </strong>DCs were more affected by specimen thickness, through which the curing light was attenuated, than RBC types. Clinician should be aware not all bulk-fill RBCs have a DoC greater than or equal to 4 mm. Also, a bulk-fill RBC that has a high DC after a post-cure time of 24 hours may not have a high DoC, which is typically measured relatively soon after light exposure.</p>","PeriodicalId":74148,"journal":{"name":"Medical journal (Fort Sam Houston, Tex.)","volume":" PB 8-21-10/11/12","pages":"74-82"},"PeriodicalIF":0.0000,"publicationDate":"2021-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Medical journal (Fort Sam Houston, Tex.)","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Introduction: Polymerization for modern dental resin-based composites (RBCs) not only occurs immediately upon light exposure but also continues for another 24 hours, well beyond after light is terminated. However, many questions still remain about the role of polymerization kinetics in optimizing the physical properties of a new RBC type-the bulk-fill.

Objective: The aim is to study the post-cure polymerization kinetics of bulk-fill RBCs and to compare their degree of polymeric conversion (DC) and depth-of-cure (DoC) with an incremental-fill, conventional RBC.

Methods: Five representative bulk-fill RBCs [Surefil SDR+Stress Decreasing Resin Flow Plus (SDRFP), Tetric EvoCeram Bulk Fill (TECB), Filtek 1 Bulk Fill (F1B), Venus Bulk Fill (VB), and Sonicfill (SF3)] and one conventional RBC [Filtek Supreme Ultra (FSU)] were investigated. The upper surface per RBC specimen was exposed to a light curing unit (Paradigm, 3M-ESPE, irradiance=1221 ± 5 mW/cm2) for 20 seconds. The DC per RBC brand were measured at the bottom surface (specimen Ø=4 mm, thickness=3 mm and 5 mm) as a function of post-curing times using a Fourier transform infrared attenuated total reflection spectrometer. Real-time data recording for post-cure DC began immediately upon light exposure and continued at steady intervals, up to15 min, then again after 24 hours. The DoC of all six RBC brands (n=6 / group) were measured according to ISO-4049. Data were analyzed with nonlinear regression and analysis of variances (ANOVA)/Tukey (α=0.05).

Results: Mean DC for the six RBCs with 5 mm curing height after 24 hours were: TECB=79.5%, VB=75.7%, SDRFP=69.2%, SF3=65.8%, F1B=51.8%, and FSU=44.0%. Bulk-fill RBCs showed higher DC efficiency than the conventional RBC for both the 3 mm and 5 mm curing heights. Significant differences in DoC were found amongst the six RBC brands: VB=5.1 mm, SDRFP=4.6 mm, F1B=3.8 mm, TECB=3.5 mm, FSU=3.0 mm, and SF3=2.7 mm.

Conclusion: DCs were more affected by specimen thickness, through which the curing light was attenuated, than RBC types. Clinician should be aware not all bulk-fill RBCs have a DoC greater than or equal to 4 mm. Also, a bulk-fill RBC that has a high DC after a post-cure time of 24 hours may not have a high DoC, which is typically measured relatively soon after light exposure.

分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
牙体填充树脂基复合材料的固化后聚合和固化深度行为。
现代牙科树脂基复合材料(rbc)的聚合不仅在光照下立即发生,而且在光照终止后还会持续24小时。然而,关于聚合动力学在优化一种新型红细胞-填充体的物理性能中的作用,仍然存在许多问题。目的:研究体积填充型红细胞的固化后聚合动力学,并比较其与增量填充型常规红细胞的聚合转化率(DC)和固化深度(DoC)。方法:研究了5种典型的填充型红细胞[Surefil SDR+Stress - reduction Resin Flow Plus (SDRFP), Tetric EvoCeram Bulk Fill (TECB), Filtek 1 Bulk Fill (F1B), Venus Bulk Fill (VB)和Sonicfill (SF3)]和1种常规RBC [Filtek Supreme Ultra (FSU)]。每个RBC标本的上表面暴露在光固化装置(Paradigm, 3M-ESPE,辐照度=1221±5 mW/cm2)下20秒。使用傅里叶变换红外衰减全反射光谱仪测量每个RBC品牌的底面DC(样品Ø=4 mm,厚度=3 mm和5 mm)作为固化后时间的函数。固化后DC的实时数据记录在光照后立即开始,并以稳定的间隔持续记录,最长可达15分钟,然后在24小时后再次记录。6个RBC品牌(n=6 /组)的DoC均按ISO-4049标准测定。数据分析采用非线性回归和方差分析(ANOVA)/Tukey (α=0.05)。结果:固化高度为5mm的6种红细胞24h后DC均值为:TECB=79.5%, VB=75.7%, SDRFP=69.2%, SF3=65.8%, F1B=51.8%, FSU=44.0%。在3mm和5mm固化高度下,填充型红细胞的直流效率均高于常规红细胞。6种红细胞类型(VB=5.1 mm, SDRFP=4.6 mm, F1B=3.8 mm, TECB=3.5 mm, FSU=3.0 mm, SF3=2.7 mm)的DoC差异有统计学意义。结论:dc受标本厚度的影响更大,通过标本厚度使固化光减弱。临床医生应注意,并非所有填充型红细胞的DoC都大于或等于4mm。此外,在固化后24小时具有高DC的散装填充RBC可能没有高DoC,这通常是在光照后相对较短的时间内测量的。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
Closing the Trauma and Critical Care Gap: A Paradigm Shift through Virtual Reality and Augmented Reality. MRI Predictive Model's Utility in a Recruit Training Environment for Tibia Stress Fractures. Platelet-Rich Plasma Improves Strength and Speed of Recovery in an Active-Duty Soldier with Isolated Injury to the Lateral Collateral Ligament of the Knee: A Case Report. The Continued Misadventures of Intentional C4 Ingestion: A Case Cluster. Impact of a Novel Biplane User Interface on Ultrasound-Guided Vascular Access Performance: A Prospective, Randomized, Crossover Study.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1