Outcomes of novel 3D-printed fully porous titanium cup and a cemented highly cross-linked polyethylene liner in complex and revision total hip arthroplasty.

IF 4.3 4区 医学 Q2 ORTHOPEDICS Arthroplasty Pub Date : 2022-12-02 DOI:10.1186/s42836-022-00152-5
Ittai Shichman, Lyndsay Somerville, William B Lutes, Stephen A Jones, Richard McCalden, Ran Schwarzkopf
{"title":"Outcomes of novel 3D-printed fully porous titanium cup and a cemented highly cross-linked polyethylene liner in complex and revision total hip arthroplasty.","authors":"Ittai Shichman,&nbsp;Lyndsay Somerville,&nbsp;William B Lutes,&nbsp;Stephen A Jones,&nbsp;Richard McCalden,&nbsp;Ran Schwarzkopf","doi":"10.1186/s42836-022-00152-5","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>A novel fully porous acetabular titanium shell has been designed to reduce stiffness mismatch between bone and implant and promote osseointegration in complex (cTHA) and revision total hip arthroplasty (rTHA). A highly cross-linked polyethylene (XLPE) liner is cemented within the cup to reduce wear rates and increase survivorship. This study reported the outcomes of an XLPE liner cemented into a novel 3D-printed fully porous cup in cTHA and rTHA.</p><p><strong>Methods: </strong>Presented was a multicenter retrospective review of 40 patients (6 cTHA and 34 rTHA) who underwent THA with a fully porous titanium acetabular cup and cemented XLPE liner. Data were collected on demographics, surgical information, outcomes, including osseointegration and migration and implant survivorship.</p><p><strong>Results: </strong>On average, patients were 71.42 ± 9.97 years old and obese (BMI: 30.36 ± 6.88 kg/m<sup>2</sup>) and were followed up for a mean time of 2.21 ± 0.77 years. Six patients underwent cTHA and 34 patients underwent rTHA. The mean hospital length of stay was 5.34 ± 3.34 days. Three (7.5%) 90-day readmissions were noted. Harris Hip Scores improved, on average, from 53.87 ± 12.58 preoperatively to 83.53 ± 12.15 postoperatively (P<0.001). One case of acetabular shell aspetic loosening with migration was noted. Thirty-nine of the 40 acetabular components were fully osseointegrated without migration. Two patients underwent re-revision surgery for PJI and one patient received acetabular shell+liner re-revision due to aseptic loosening. Kaplan-Meier analysis showed an all-cause revision-free survival rate of 95.0% at 6 months and 1 year, and 92.0% at 4-years. Aseptic acetabular cup, liner dislocation/loosening, and fracture-free survival was 100% at 6 months and 1-year, and 97.1% at 2 years.</p><p><strong>Conclusion: </strong>The combined use of a novel 3D-printed fully porous titanium acetabular shell and cemented XLPE acetabular liner yielded excellent rates of osseointegration, and all-cause and acetabular aseptic loosening survivorship at a minimum 1-year follow-up. Further long-term studies are needed to assess the longevity of this construct.</p>","PeriodicalId":52831,"journal":{"name":"Arthroplasty","volume":" ","pages":"51"},"PeriodicalIF":4.3000,"publicationDate":"2022-12-02","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9717502/pdf/","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Arthroplasty","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1186/s42836-022-00152-5","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"ORTHOPEDICS","Score":null,"Total":0}
引用次数: 1

Abstract

Introduction: A novel fully porous acetabular titanium shell has been designed to reduce stiffness mismatch between bone and implant and promote osseointegration in complex (cTHA) and revision total hip arthroplasty (rTHA). A highly cross-linked polyethylene (XLPE) liner is cemented within the cup to reduce wear rates and increase survivorship. This study reported the outcomes of an XLPE liner cemented into a novel 3D-printed fully porous cup in cTHA and rTHA.

Methods: Presented was a multicenter retrospective review of 40 patients (6 cTHA and 34 rTHA) who underwent THA with a fully porous titanium acetabular cup and cemented XLPE liner. Data were collected on demographics, surgical information, outcomes, including osseointegration and migration and implant survivorship.

Results: On average, patients were 71.42 ± 9.97 years old and obese (BMI: 30.36 ± 6.88 kg/m2) and were followed up for a mean time of 2.21 ± 0.77 years. Six patients underwent cTHA and 34 patients underwent rTHA. The mean hospital length of stay was 5.34 ± 3.34 days. Three (7.5%) 90-day readmissions were noted. Harris Hip Scores improved, on average, from 53.87 ± 12.58 preoperatively to 83.53 ± 12.15 postoperatively (P<0.001). One case of acetabular shell aspetic loosening with migration was noted. Thirty-nine of the 40 acetabular components were fully osseointegrated without migration. Two patients underwent re-revision surgery for PJI and one patient received acetabular shell+liner re-revision due to aseptic loosening. Kaplan-Meier analysis showed an all-cause revision-free survival rate of 95.0% at 6 months and 1 year, and 92.0% at 4-years. Aseptic acetabular cup, liner dislocation/loosening, and fracture-free survival was 100% at 6 months and 1-year, and 97.1% at 2 years.

Conclusion: The combined use of a novel 3D-printed fully porous titanium acetabular shell and cemented XLPE acetabular liner yielded excellent rates of osseointegration, and all-cause and acetabular aseptic loosening survivorship at a minimum 1-year follow-up. Further long-term studies are needed to assess the longevity of this construct.

Abstract Image

Abstract Image

Abstract Image

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
新型3d打印全多孔钛杯和胶结高度交联聚乙烯衬垫在复杂和翻修全髋关节置换术中的效果。
一种新型的全多孔髋臼钛壳被设计用于减少骨与植入物之间的刚度不匹配,促进复杂(cTHA)和翻修全髋关节置换术(rTHA)中的骨整合。在杯内粘接高交联聚乙烯(XLPE)衬管,以降低磨损率,提高寿命。本研究报告了在cTHA和rTHA中,将XLPE衬套粘合到新型3d打印全多孔杯中的结果。方法:对40例(6例cTHA和34例rTHA)采用全孔钛髋臼杯和骨水泥XLPE衬垫行THA的患者进行多中心回顾性分析。收集的数据包括人口统计学、手术信息、结果,包括骨整合、迁移和种植体存活。结果:患者平均年龄为71.42±9.97岁,肥胖(BMI: 30.36±6.88 kg/m2),平均随访时间为2.21±0.77年。6例患者行cTHA, 34例患者行rTHA。平均住院时间为5.34±3.34天。90天再入院3例(7.5%)。Harris髋关节评分平均从术前的53.87±12.58分提高到术后的83.53±12.15分。结论:新型3d打印全多孔钛髋臼壳和骨水泥XLPE髋臼衬套的联合使用在至少1年的随访中获得了极好的骨整合率,以及全因和髋臼无菌性松动的生存率。需要进一步的长期研究来评估这种结构的寿命。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Arthroplasty
Arthroplasty ORTHOPEDICS-
CiteScore
2.20
自引率
0.00%
发文量
49
审稿时长
15 weeks
期刊最新文献
Risk analysis index demonstrates superior predictive performance compared to traditional frailty metrics in elderly female patients undergoing inpatient total shoulder arthroplasty. Anatomical relationship between the intertrochanteric line and femoral neck osteotomy level in direct anterior approach total hip arthroplasty: a 3D morphometric and cadaveric validation study. Restricted kinematic alignment total knee arthroplasty using augmented reality technology to maintain limb alignment within targeted boundaries. National evidence linking robotic total knee arthroplasty to reduced 90-day readmissions, complications, and readmission costs. Exploring the optimal age for total knee arthroplasty to minimize risk of adverse outcomes: machine learning analysis of a statewide cohort.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1