Systematic Review of Diagnostic Accuracy of DiaSorin Liaison SARS-CoV-2 Antigen Immunoassay.

Q2 Medicine Electronic Journal of the International Federation of Clinical Chemistry and Laboratory Medicine Pub Date : 2022-08-08 eCollection Date: 2022-08-01
Giuseppe Lippi, Brandon M Henry, Mario Plebani, Khosrow Adeli
{"title":"Systematic Review of Diagnostic Accuracy of DiaSorin Liaison SARS-CoV-2 Antigen Immunoassay.","authors":"Giuseppe Lippi, Brandon M Henry, Mario Plebani, Khosrow Adeli","doi":"","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Quantification of SARS-CoV-2 antigens by means of rapid, high-throughput and fully-automated techniques has been proposed as a feasible alternative to overcome the current shortage of resources for routine molecular diagnostics. To this end, we provide here a systematic review of diagnostic accuracy of DiaSorin Liaison SARS-CoV-2 antigen immunoassay.</p><p><strong>Methods: </strong>An electronic search was conduced in Medline and Scopus, with no language or date restrictions (up to January 20, 2022), for identifying all published studies articles in which the diagnostic performance of the DiaSorin Liaison SARS-CoV-2 antigen immunoassay was compared with molecular diagnostic techniques.</p><p><strong>Results: </strong>The electronic search identified a final number of 11 studies, totalling 4449 oro- and naso-pharyngeal specimens. The pooled diagnostic sensitivity, specificity and area under the curve (AUC) of the Liaison SARS-CoV-2 antigen immunoassay in all samples were 0.51 (95%CI, 0.49-0.54), 1.00 (95%CI, 1.00-1.00) and 0.994 (95%CI, 0.990-0.998), respectively, whilst the overall concordance with molecular diagnostics was 82.1%. The pooled diagnostic sensitivity, specificity and AUC of the Liaison SARS-CoV-2 antigen immunoassay in specimens with high viral load (i.e., cycle threshold values <25-30) were 0.79 (95%CI, 0.75-0.82), 1.00 (95%CI, 0.99-1.00) and 0.911 (95%CI, 0.879-0.943), respectively, whilst the overall concordance with molecular diagnostics in such samples increased to 94.2%.</p><p><strong>Conclusion: </strong>The results of this systematic literature review suggest that there is sufficient accuracy of the DiaSorin Liaison SARS-CoV-2 antigen immunoassay in samples with high viral loads that would enable its reliable usage for identifying superspreaders, who are responsible for the vast majority of transmission events.</p>","PeriodicalId":37192,"journal":{"name":"Electronic Journal of the International Federation of Clinical Chemistry and Laboratory Medicine","volume":"33 2","pages":"94-104"},"PeriodicalIF":0.0000,"publicationDate":"2022-08-08","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://ftp.ncbi.nlm.nih.gov/pub/pmc/oa_pdf/4d/ab/ejifcc-33-094.PMC9562478.pdf","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Electronic Journal of the International Federation of Clinical Chemistry and Laboratory Medicine","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2022/8/1 0:00:00","PubModel":"eCollection","JCR":"Q2","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0

Abstract

Background: Quantification of SARS-CoV-2 antigens by means of rapid, high-throughput and fully-automated techniques has been proposed as a feasible alternative to overcome the current shortage of resources for routine molecular diagnostics. To this end, we provide here a systematic review of diagnostic accuracy of DiaSorin Liaison SARS-CoV-2 antigen immunoassay.

Methods: An electronic search was conduced in Medline and Scopus, with no language or date restrictions (up to January 20, 2022), for identifying all published studies articles in which the diagnostic performance of the DiaSorin Liaison SARS-CoV-2 antigen immunoassay was compared with molecular diagnostic techniques.

Results: The electronic search identified a final number of 11 studies, totalling 4449 oro- and naso-pharyngeal specimens. The pooled diagnostic sensitivity, specificity and area under the curve (AUC) of the Liaison SARS-CoV-2 antigen immunoassay in all samples were 0.51 (95%CI, 0.49-0.54), 1.00 (95%CI, 1.00-1.00) and 0.994 (95%CI, 0.990-0.998), respectively, whilst the overall concordance with molecular diagnostics was 82.1%. The pooled diagnostic sensitivity, specificity and AUC of the Liaison SARS-CoV-2 antigen immunoassay in specimens with high viral load (i.e., cycle threshold values <25-30) were 0.79 (95%CI, 0.75-0.82), 1.00 (95%CI, 0.99-1.00) and 0.911 (95%CI, 0.879-0.943), respectively, whilst the overall concordance with molecular diagnostics in such samples increased to 94.2%.

Conclusion: The results of this systematic literature review suggest that there is sufficient accuracy of the DiaSorin Liaison SARS-CoV-2 antigen immunoassay in samples with high viral loads that would enable its reliable usage for identifying superspreaders, who are responsible for the vast majority of transmission events.

Abstract Image

Abstract Image

分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
DiaSorin Liaison SARS-CoV-2 抗原免疫测定诊断准确性的系统回顾。
背景:通过快速、高通量和全自动技术对 SARS-CoV-2 抗原进行定量分析被认为是克服目前常规分子诊断资源短缺的一种可行的替代方法。为此,我们在此对 DiaSorin Liaison SARS-CoV-2 抗原免疫测定的诊断准确性进行了系统回顾:方法:我们在 Medline 和 Scopus 上进行了电子检索,没有语言和日期限制(截至 2022 年 1 月 20 日),以确定所有已发表的研究文章,其中对迪亚索林联络 SARS-CoV-2 抗原免疫测定的诊断性能与分子诊断技术进行了比较:电子检索最终确定了 11 项研究,共计 4449 份口腔和鼻咽标本。联络 SARS-CoV-2 抗原免疫测定法对所有样本的诊断灵敏度、特异性和曲线下面积(AUC)分别为 0.51(95%CI,0.49-0.54)、1.00(95%CI,1.00-1.00)和 0.994(95%CI,0.990-0.998),与分子诊断的总体一致性为 82.1%。联络 SARS-CoV-2 抗原免疫测定对高病毒载量(即周期阈值)标本的诊断灵敏度、特异性和 AUC 的总和 结论:对高病毒载量(即周期阈值)标本的诊断灵敏度、特异性和 AUC 的总和分别为 0.994(95%CI)和 0.990(95%CI):本系统文献综述的结果表明, DiaSorin Liaison SARS-CoV-2 抗原免疫测定在高病毒载量样本中具有足够的准确性,可以可靠地用于识别超级传播者,他们是绝大多数传播事件的罪魁祸首。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
CiteScore
2.30
自引率
0.00%
发文量
0
期刊最新文献
Adolopment of Recommendations for Standardized Reporting of Protein Electrophoresis in Pakistan. Continuous reference intervals for plasma cystatin C and creatinine in Vietnamese children. Correlation Analysis of Direct LDL Measurement and Calculated LDL Methods in Lipid Profile Assessment: A Comprehensive Study. Diagnostic Accuracy of Creatinine-Based Equations for eGFR Estimation in Pakistanis: Evaluation of the European Kidney Function Consortium Equation vs the CKD-EPI Pakistan Equation. Integrating Patient-Generated Health Data from Mobile Devices into Electronic Health Records: Best Practice Recommendations by the IFCC Committee on Mobile Health and Bioengineering in Laboratory Medicine (C-MHBLM).
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1