Evaluating the sensitivity and specificity of Determine™ HIV-1/2 rapid test using a 0.01M phosphate-buffered saline produced at the Medical Research Council Unit The Gambia for the diagnosis of HIV.

IF 1.9 4区 医学 Q3 PUBLIC, ENVIRONMENTAL & OCCUPATIONAL HEALTH Transactions of The Royal Society of Tropical Medicine and Hygiene Pub Date : 2024-02-01 DOI:10.1093/trstmh/trad071
Anna Boté-Casamitjana, Tisbeh Faye-Joof, Ousman Bah, Sira Jallow, Alagie Camara, Olimatou Jallow, Nuredin Mohammed, Karen Forrest, Behzad Nadjm
{"title":"Evaluating the sensitivity and specificity of Determine™ HIV-1/2 rapid test using a 0.01M phosphate-buffered saline produced at the Medical Research Council Unit The Gambia for the diagnosis of HIV.","authors":"Anna Boté-Casamitjana, Tisbeh Faye-Joof, Ousman Bah, Sira Jallow, Alagie Camara, Olimatou Jallow, Nuredin Mohammed, Karen Forrest, Behzad Nadjm","doi":"10.1093/trstmh/trad071","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Human immunodeficiency virus (HIV) rapid diagnostic tests (RDTs) are widely used. However, buffer stockouts commonly lead to utilising non-approved liquids, resulting in errors. Our aim was to evaluate the diagnostic accuracy of an alternative buffer.</p><p><strong>Methods: </strong>Paired Determine HIV-1/2 rapid tests with commercial buffer and locally produced 0.01M phosphate-buffered saline (PBS) were performed on consecutive consenting individuals requiring HIV testing. Serum samples were sent for confirmation through the local gold-standard algorithm (Murex HIV Ag/Ab, Hexagon HIV with/without Geenius HIV 1/2). Test accuracy, κ and exact McNemar's test were also carried out.</p><p><strong>Results: </strong>Of 167 participants, 137 had confirmatory testing. The sensitivity of the Determine HIV-1/2 test using PBS compared with the gold standard was 100% (95% confidence interval [CI] 90.5 to 100) with a specificity of 98% (95% CI 92.9 to 99.8). The κ value was 0.94 compared with the gold standard and 0.92 compared with the Determine HIV-1/2 test using the commercial buffer. McNemar's test showed no evidence of differing sensitivities. Due to operational constraints, the study included 37 of the 49 positive cases as determined by the sample size calculation, resulting in an attained power of 80% instead of the intended 90%.</p><p><strong>Conclusions: </strong>These results suggest that 0.01M PBS is an alternative solution for Determine HIV-1/2 when buffer stockouts occur.</p>","PeriodicalId":23218,"journal":{"name":"Transactions of The Royal Society of Tropical Medicine and Hygiene","volume":null,"pages":null},"PeriodicalIF":1.9000,"publicationDate":"2024-02-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10833640/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Transactions of The Royal Society of Tropical Medicine and Hygiene","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1093/trstmh/trad071","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"PUBLIC, ENVIRONMENTAL & OCCUPATIONAL HEALTH","Score":null,"Total":0}
引用次数: 0

Abstract

Background: Human immunodeficiency virus (HIV) rapid diagnostic tests (RDTs) are widely used. However, buffer stockouts commonly lead to utilising non-approved liquids, resulting in errors. Our aim was to evaluate the diagnostic accuracy of an alternative buffer.

Methods: Paired Determine HIV-1/2 rapid tests with commercial buffer and locally produced 0.01M phosphate-buffered saline (PBS) were performed on consecutive consenting individuals requiring HIV testing. Serum samples were sent for confirmation through the local gold-standard algorithm (Murex HIV Ag/Ab, Hexagon HIV with/without Geenius HIV 1/2). Test accuracy, κ and exact McNemar's test were also carried out.

Results: Of 167 participants, 137 had confirmatory testing. The sensitivity of the Determine HIV-1/2 test using PBS compared with the gold standard was 100% (95% confidence interval [CI] 90.5 to 100) with a specificity of 98% (95% CI 92.9 to 99.8). The κ value was 0.94 compared with the gold standard and 0.92 compared with the Determine HIV-1/2 test using the commercial buffer. McNemar's test showed no evidence of differing sensitivities. Due to operational constraints, the study included 37 of the 49 positive cases as determined by the sample size calculation, resulting in an attained power of 80% instead of the intended 90%.

Conclusions: These results suggest that 0.01M PBS is an alternative solution for Determine HIV-1/2 when buffer stockouts occur.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
评估Determine的敏感性和特异性™ 使用冈比亚医学研究委员会单位生产的0.01M磷酸盐缓冲盐水进行HIV-1/2快速检测,用于诊断HIV。
背景:人类免疫缺陷病毒(HIV)快速诊断试验(RDTs)应用广泛。然而,缓冲器缺货通常会导致使用未经批准的液体,从而导致错误。我们的目的是评估替代缓冲液的诊断准确性。方法:用商业缓冲液和本地生产的0.01M磷酸盐缓冲盐水(PBS)对需要HIV检测的连续同意个体进行配对测定HIV-1/2快速检测。通过当地金标准算法(Murex HIV Ag/Ab,Hexagon HIV带/不带Geenius HIV 1/2)发送血清样本进行确认。还进行了检验准确性、κ和精确的McNemar检验。结果:在167名参与者中,137人进行了验证性测试。与金标准相比,使用PBS测定HIV-1/2试验的灵敏度为100%(95%置信区间[CI]90.5至100),特异性为98%(95%可信区间92.9至99.8)。与金标准比较,κ值为0.94,与使用商业缓冲液测定HIV-1/2试验相比为0.92。McNemar的测试没有显示出不同敏感性的证据。由于操作限制,该研究包括了样本量计算确定的49例阳性病例中的37例,结果获得了80%的功率,而不是预期的90%。结论:这些结果表明,当缓冲液缺货时,0.01M PBS是确定HIV-1/2的替代解决方案。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Transactions of The Royal Society of Tropical Medicine and Hygiene
Transactions of The Royal Society of Tropical Medicine and Hygiene 医学-公共卫生、环境卫生与职业卫生
CiteScore
4.00
自引率
9.10%
发文量
115
审稿时长
4-8 weeks
期刊介绍: Transactions of the Royal Society of Tropical Medicine and Hygiene publishes authoritative and impactful original, peer-reviewed articles and reviews on all aspects of tropical medicine.
期刊最新文献
Foetal outcome of Leptospira and Rickettsial infections during pregnancy: a systematic review. A combined tropical medicine and psychiatry approach to patients with possible delusional infestation. 'My feet cannot stand on their own': podoconiosis patient healthcare expenditures and income impacts in Rwanda. Risk of human leptospirosis in Colombia: spatiotemporal analysis and related hydroclimatic factors. Understanding noma: WHO's recognition and the path forward in global health.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1