{"title":"Study on the Safety of the New Radial Artery Hemostasis Device","authors":"Baofeng Wu, Rui Zhang, C. Liang, Chengjie Zhang, Gang Qin","doi":"10.1155/2022/2345584","DOIUrl":null,"url":null,"abstract":"Objective At present, the use of particular radial hemostatic devices after coronary angiography (CAG) or percutaneous coronary intervention (PCI) has become the primary method of hemostasis. Most control studies are based on the products already on the market, while only a few studies are on the new hemostatic devices. The aim of this study is to compare a new radial artery hemostasis device which is transformed based on the invention patent (Application number: CN201510275446) with TR Band (Terumo Medical) to evaluate its clinical effects. Methods In a prospective randomized clinical trial, 60 patients after CAG or PCI were randomly divided into two groups, patients in the trial group (CD group) using a new radial artery hemostasis device to stop bleeding and the control group (TR group) using the TR Band. The method is to collect relevant data of the two groups and compare the differences in hemostasis, local complications, and patient discomfort between the two groups. Results The hemostatic devices in both groups achieved adequate hemostasis, and there was no failure to stop bleeding. The new radial artery hemostasis device was better than the TR band in pain and swelling (P < 0.05). There were no significant differences in bleeding, hematoma, ecchymosis, skin damage, and local infection between the two groups (P > 0.05). Conclusions The sample of the new radial artery hemostasis device can stop bleeding effectively at the puncture site after CAG or PCI and is not inferior to the TR Band balloon hemostatic device in safety and is better in comfort.","PeriodicalId":16329,"journal":{"name":"Journal of interventional cardiology","volume":" ","pages":""},"PeriodicalIF":1.6000,"publicationDate":"2022-04-05","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of interventional cardiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1155/2022/2345584","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 1

Abstract

Objective At present, the use of particular radial hemostatic devices after coronary angiography (CAG) or percutaneous coronary intervention (PCI) has become the primary method of hemostasis. Most control studies are based on the products already on the market, while only a few studies are on the new hemostatic devices. The aim of this study is to compare a new radial artery hemostasis device which is transformed based on the invention patent (Application number: CN201510275446) with TR Band (Terumo Medical) to evaluate its clinical effects. Methods In a prospective randomized clinical trial, 60 patients after CAG or PCI were randomly divided into two groups, patients in the trial group (CD group) using a new radial artery hemostasis device to stop bleeding and the control group (TR group) using the TR Band. The method is to collect relevant data of the two groups and compare the differences in hemostasis, local complications, and patient discomfort between the two groups. Results The hemostatic devices in both groups achieved adequate hemostasis, and there was no failure to stop bleeding. The new radial artery hemostasis device was better than the TR band in pain and swelling (P < 0.05). There were no significant differences in bleeding, hematoma, ecchymosis, skin damage, and local infection between the two groups (P > 0.05). Conclusions The sample of the new radial artery hemostasis device can stop bleeding effectively at the puncture site after CAG or PCI and is not inferior to the TR Band balloon hemostatic device in safety and is better in comfort.
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
新型桡动脉止血装置的安全性研究
目的目前,在冠状动脉造影(CAG)或经皮冠状动脉介入治疗(PCI)后使用特定的径向止血装置已成为主要的止血方法。大多数对照研究是基于市场上已有的产品,而只有少数研究是针对新的止血装置的。本研究的目的是比较一种基于发明专利(申请号:CN201510275446)改造的新型桡动脉止血装置与TR Band (Terumo Medical)的临床效果。方法采用前瞻性随机临床试验,将60例CAG或PCI术后患者随机分为两组,实验组(CD组)采用新型桡动脉止血装置止血,对照组(TR组)采用TR带止血。方法是收集两组的相关资料,比较两组在止血、局部并发症、患者不适等方面的差异。结果两组患者使用的止血装置均能充分止血,均无止血失败。新型桡动脉止血装置在疼痛和肿胀方面优于TR带(P < 0.05)。两组患者出血、血肿、瘀斑、皮肤损伤、局部感染的发生率比较,差异无统计学意义(P < 0.05)。结论新型桡动脉止血装置样品在CAG或PCI术后穿刺部位均能有效止血,其安全性不逊于TR带状球囊止血装置,舒适性更佳。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Journal of interventional cardiology
Journal of interventional cardiology CARDIAC & CARDIOVASCULAR SYSTEMS-
CiteScore
3.80
自引率
0.00%
发文量
81
审稿时长
6-12 weeks
期刊介绍: Journal of Interventional Cardiology is a peer-reviewed, Open Access journal that provides a forum for cardiologists determined to stay current in the diagnosis, investigation, and management of patients with cardiovascular disease and its associated complications. The journal publishes original research articles, review articles, and clinical studies focusing on new procedures and techniques in all major subject areas in the field, including: Acute coronary syndrome Coronary disease Congenital heart diseases Myocardial infarction Peripheral arterial disease Valvular heart disease Cardiac hemodynamics and physiology Haemostasis and thrombosis
期刊最新文献
A Comparative Analysis of Primary and Bailout ADR in CTO-PCI The Effect of Electromagnetic Interference Produced by Smartphones Using 5G Network on Patients With Permanent Pacemakers (EMS5G-PPM Study) Comparison of Sheathless and Sheathed Guiding Catheters in Transradial Percutaneous Coronary Interventions: A Systematic Review and Meta-Analysis Percutaneous Coronary Intervention Using the DynamX Sirolimus-Eluting Bioadaptor: 12-Month Clinical and Imaging Outcomes Vascular Complications and Bleeding After Transfemoral TAVR With Surgical Versus Percutaneous Approach: A Contemporary Prospective Study
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1