Nd:YAG Laser therapy device for caries control and prevention and its influence on temperature of isolated tooth pulp chamber

Xiaoyue Liang, Zhaohui Zou, Xiaoxi Dong, Ranran Du, Yu Zhang
{"title":"Nd:YAG Laser therapy device for caries control and prevention and its influence on temperature of isolated tooth pulp chamber","authors":"Xiaoyue Liang, Zhaohui Zou, Xiaoxi Dong, Ranran Du, Yu Zhang","doi":"10.3760/CMA.J.ISSN.1673-4181.2019.06.012","DOIUrl":null,"url":null,"abstract":"Objective \nTo analyze the influence of a dental caries phototherapy device using neodymium-yttrium-aluminum-garnet (Nd:YAG) laser on the temperature of tooth pulp chamber of different sites and morphologies under different irradiation duration and power. \n \n \nMethods \nFifty intact isolated teeth were collected and randomly divided into middle incisor group, upper right first premolar group, upper right second molar group, upper left first premolar group, and upper left second molar group. Each experimental group was irradiated according to the laser power 1.5, 1.6, 1.7, 1.8, 1.9 and 2.0 W, and the irradiation duration 30, 60, 90 s. The temperature rise of the pulp chamber was recorded with a thermocouple thermometer. \n \n \nResults \nAfter the laser irradiation, the temperature of the tooth pulp chamber increased, and the temperature rise was less than 5.5 ℃, i.e. the threshold leading to the dental pulp necrosis. \n \n \nConclusions \nThe use of laser to prevent dental caries will increase the temperature of the tooth pulp chamber, but the temperature rise in the range of 5.5 ° C is relatively safe and will not cause irreversible damage to the pulp tissue. \n \n \nKey words: \nOral phototherapy device; Laser; Pulp chamber; Temperature","PeriodicalId":61751,"journal":{"name":"国际生物医学工程杂志","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2019-12-28","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"国际生物医学工程杂志","FirstCategoryId":"1087","ListUrlMain":"https://doi.org/10.3760/CMA.J.ISSN.1673-4181.2019.06.012","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Objective To analyze the influence of a dental caries phototherapy device using neodymium-yttrium-aluminum-garnet (Nd:YAG) laser on the temperature of tooth pulp chamber of different sites and morphologies under different irradiation duration and power. Methods Fifty intact isolated teeth were collected and randomly divided into middle incisor group, upper right first premolar group, upper right second molar group, upper left first premolar group, and upper left second molar group. Each experimental group was irradiated according to the laser power 1.5, 1.6, 1.7, 1.8, 1.9 and 2.0 W, and the irradiation duration 30, 60, 90 s. The temperature rise of the pulp chamber was recorded with a thermocouple thermometer. Results After the laser irradiation, the temperature of the tooth pulp chamber increased, and the temperature rise was less than 5.5 ℃, i.e. the threshold leading to the dental pulp necrosis. Conclusions The use of laser to prevent dental caries will increase the temperature of the tooth pulp chamber, but the temperature rise in the range of 5.5 ° C is relatively safe and will not cause irreversible damage to the pulp tissue. Key words: Oral phototherapy device; Laser; Pulp chamber; Temperature
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Nd:YAG激光防龋装置及其对离体牙髓腔温度的影响
目的分析采用钕钇铝石榴石(Nd:YAG)激光治疗龋病的装置在不同照射时间和功率下对牙髓腔不同部位和形态温度的影响。方法收集50颗完整离体牙,随机分为中切牙组、右上第一前磨牙组、右上第二磨牙组、左上第一前磨牙组和左上第二磨牙组。实验组分别按激光功率1.5、1.6、1.7、1.8、1.9、2.0 W,照射时间30、60、90 s进行照射。用热电偶温度计记录了牙髓腔的温升。结果激光照射后牙髓腔温度升高,且温升小于5.5℃,即导致牙髓坏死的阈值。结论使用激光预防龋病会使牙髓腔温度升高,但温度升高在5.5℃范围内是相对安全的,不会对牙髓组织造成不可逆的损伤。关键词:口服光疗仪;激光;髓室;温度
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
自引率
0.00%
发文量
1974
期刊最新文献
A Study on Assessment of Depression, Anxiety and Stress among South Indian Population During the COVID-19 Pandemic Study on Antiviral Activities of Glycyrrhizin Improving Clinical Teaching of Fourth Year Anaesthesia Students: The Case of Wolaita Sodo University Hearing Aids of the Future: A Simulation Study Association Between the Distributions of Mean Corpuscular Hemoglobin and Red Blood Cell, and Mortality in a 3-Year Retrospective Study of Hemodialysis Patients
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1