Comparison of accuracy and reliability of CBCT and 3D laser scanner in the volumetric assessment of the root canal space.

IF 0.9 4区 医学 Q3 DENTISTRY, ORAL SURGERY & MEDICINE American journal of dentistry Pub Date : 2023-10-01
Anchu R Thomas, Htoo Htoo K Soe, Christine S Silva, Harpeven Kaur, Loshanii D Ganendrah, Lynette M Gomez
{"title":"Comparison of accuracy and reliability of CBCT and 3D laser scanner in the volumetric assessment of the root canal space.","authors":"Anchu R Thomas,&nbsp;Htoo Htoo K Soe,&nbsp;Christine S Silva,&nbsp;Harpeven Kaur,&nbsp;Loshanii D Ganendrah,&nbsp;Lynette M Gomez","doi":"","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To compare the accuracy and reliability of cone-beam computed tomography (CBCT) and laser scanner in measuring minor volume changes such as the root canal space.</p><p><strong>Methods: </strong>35 maxillary incisors were endodontically prepared. A dimensionally stable silicone material was injected into the root canal space and scanned with CBCT. The root canal volume was measured using Romexis 3.0.1 R software. Replicas were carefully removed from the teeth and scanned using an extraoral laser scanner. These images were exported to the Rhinoceros software for volume measurement. The volume of each replica was also assessed using the gravimetric method. To determine the accuracy, the volume obtained from both devices was compared with the gravimetric method. Statistical analysis was done using a paired t-test. The reliability was assessed using the intraclass correlation coefficient.</p><p><strong>Results: </strong>There was no statistically significant difference between the mean volume of CBCT 27.04 ± 7.25 mm³ and the mean volume of the gravimetric method 27.87 ± 7.17 mm³ (P< 0.05). A statistically significant difference was seen with the laser scanner at 25.31 ± 6.89 mm³ and the gravimetric method at 27.87 ± 7.17 mm³ (P< 0.05). CBCT showed a good degree of agreement (ICC 0.899), while the laser scanner showed a moderate degree of agreement (ICC 0.644) with the gravimetric method. CBCT proved accurate and reliable in measuring minor volumes like the root canal space, ideally in the range of 20-25 mm³. The laser scanner presented acceptable reliability.</p><p><strong>Clinical significance: </strong>The laboratory data showed satisfactory outcomes, providing an evidence-based approach and potentially motivating clinicians to integrate cone-beam computed tomography for volume analysis into clinical practice. The accuracy and reliability of laser scanners for small-volume analysis have not previously been evaluated. Consequently, the findings from this study warrant further clinical investigations.</p>","PeriodicalId":7538,"journal":{"name":"American journal of dentistry","volume":null,"pages":null},"PeriodicalIF":0.9000,"publicationDate":"2023-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"American journal of dentistry","FirstCategoryId":"3","ListUrlMain":"","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: To compare the accuracy and reliability of cone-beam computed tomography (CBCT) and laser scanner in measuring minor volume changes such as the root canal space.

Methods: 35 maxillary incisors were endodontically prepared. A dimensionally stable silicone material was injected into the root canal space and scanned with CBCT. The root canal volume was measured using Romexis 3.0.1 R software. Replicas were carefully removed from the teeth and scanned using an extraoral laser scanner. These images were exported to the Rhinoceros software for volume measurement. The volume of each replica was also assessed using the gravimetric method. To determine the accuracy, the volume obtained from both devices was compared with the gravimetric method. Statistical analysis was done using a paired t-test. The reliability was assessed using the intraclass correlation coefficient.

Results: There was no statistically significant difference between the mean volume of CBCT 27.04 ± 7.25 mm³ and the mean volume of the gravimetric method 27.87 ± 7.17 mm³ (P< 0.05). A statistically significant difference was seen with the laser scanner at 25.31 ± 6.89 mm³ and the gravimetric method at 27.87 ± 7.17 mm³ (P< 0.05). CBCT showed a good degree of agreement (ICC 0.899), while the laser scanner showed a moderate degree of agreement (ICC 0.644) with the gravimetric method. CBCT proved accurate and reliable in measuring minor volumes like the root canal space, ideally in the range of 20-25 mm³. The laser scanner presented acceptable reliability.

Clinical significance: The laboratory data showed satisfactory outcomes, providing an evidence-based approach and potentially motivating clinicians to integrate cone-beam computed tomography for volume analysis into clinical practice. The accuracy and reliability of laser scanners for small-volume analysis have not previously been evaluated. Consequently, the findings from this study warrant further clinical investigations.

分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
CBCT和3D激光扫描仪在根管间隙体积评估中的准确性和可靠性比较。
目的:比较锥形束计算机断层扫描(CBCT)和激光扫描仪测量根管间隙等微小体积变化的准确性和可靠性。方法:对35颗上颌切牙进行根管预备。将尺寸稳定的硅树脂材料注入根管间隙并用CBCT扫描。使用Romexis 3.0.1 R软件测量根管体积。从牙齿上仔细取出复制品,并使用口腔外激光扫描仪进行扫描。将这些图像导出到Rhinoceros软件中进行体积测量。每个复制品的体积也使用重量分析法进行评估。为了确定准确度,将从两个装置获得的体积与重量分析法进行比较。采用配对t检验进行统计分析。使用组内相关系数评估可靠性。结果:CBCT平均体积27.04±7.25 mm³与重量法平均体积27.87±7.17 mm³无统计学意义(P<0.05),激光扫描25.31±6.89 mm³与重力法27.87±7.1 7 mm³有统计学意义(P<0.01),而激光扫描仪显示出与重量分析法的中等程度的一致性(ICC 0.644)。CBCT在测量较小体积(如根管间隙)方面被证明是准确可靠的,理想范围为20-25 mm³。激光扫描仪具有可接受的可靠性。临床意义:实验室数据显示出令人满意的结果,提供了一种循证方法,并有可能激励临床医生将锥束计算机断层扫描用于体积分析纳入临床实践。用于小体积分析的激光扫描仪的准确性和可靠性以前没有得到评估。因此,这项研究的发现值得进一步的临床研究。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
American journal of dentistry
American journal of dentistry 医学-牙科与口腔外科
CiteScore
2.40
自引率
7.10%
发文量
57
审稿时长
1 months
期刊介绍: The American Journal of Dentistry, published by Mosher & Linder, Inc., provides peer-reviewed scientific articles with clinical significance for the general dental practitioner.
期刊最新文献
A comparison of polishing systems and thermal cycling on the surface roughness and color stability of a single-shade resin composite. Charcoal dentifrices: A laboratory study of their safety and efficacy. Current landscape of research on whitening toothpastes and their effects on dental hard tissue. Effects of salivary contamination on the shear bond strengths of universal adhesives to dentin. Framework's marginal adaptation evaluation of fixed partial denture using conventional and digital impression techniques.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1