Is in vivo sensing in a total hip replacement a possibility? A review on past systems and future challenges

IF 5 Q1 ENGINEERING, BIOMEDICAL Progress in biomedical engineering (Bristol, England) Pub Date : 2021-01-01 DOI:10.1088/2516-1091/ac1b7f
O. Vickers, P. Culmer, G. Isaac, R. Kay, M. P. Shuttleworth, T. Board, S. Williams
{"title":"Is in vivo sensing in a total hip replacement a possibility? A review on past systems and future challenges","authors":"O. Vickers, P. Culmer, G. Isaac, R. Kay, M. P. Shuttleworth, T. Board, S. Williams","doi":"10.1088/2516-1091/ac1b7f","DOIUrl":null,"url":null,"abstract":"Surgery to implant a total hip replacement (THR) is very successful in reducing pain and restoring function. This procedure has become more prevalent, and projections estimate further increases in demand. However, complications can arise, and current diagnostic techniques often fail to expose underlying issues before they result in a catastrophic failure that requires revision surgery. An instrumented implant, with embedded sensors capable of real time condition monitoring, would be an attractive proposition to incorporate within a THR. Continued advances in the performance and miniaturisation of electronic components, embedded systems, sensing and wireless communications have given the tools and resources medical device manufacturers need to innovate in the field of implantable medical devices. Smart implants are already being widely used in healthcare including pacemakers, cochlear implants, glucose monitors and insulin pumps however, a widely used smart THR has not yet been realised. Since the implantation of the first instrumented hip implant in the 1960s there have been several in vitro studies monitoring levels of implant loosening. Additionally, significant research has been conducted using instrumented THRs to perform in vivo measurement of biomechanical metrics, including force and moments. To date less than 100 patients have successfully received an instrumented implant. The results of these studies have aided researchers, designers and surgeons in wider research projects, however, the motivation behind the work was to provide discrete biomechanical data sets and not provide real-time condition monitoring of an implants performance or highlight early indications for revision surgery. If in vivo sensing within a THR is to be achieved and adopted in regular clinical practice then the following challenges need to be addressed: choice of the sensing method, biocompatibility and integration within the implant, power supply, communication, and regulatory considerations.","PeriodicalId":74582,"journal":{"name":"Progress in biomedical engineering (Bristol, England)","volume":null,"pages":null},"PeriodicalIF":5.0000,"publicationDate":"2021-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"2","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Progress in biomedical engineering (Bristol, England)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1088/2516-1091/ac1b7f","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 2

Abstract

Surgery to implant a total hip replacement (THR) is very successful in reducing pain and restoring function. This procedure has become more prevalent, and projections estimate further increases in demand. However, complications can arise, and current diagnostic techniques often fail to expose underlying issues before they result in a catastrophic failure that requires revision surgery. An instrumented implant, with embedded sensors capable of real time condition monitoring, would be an attractive proposition to incorporate within a THR. Continued advances in the performance and miniaturisation of electronic components, embedded systems, sensing and wireless communications have given the tools and resources medical device manufacturers need to innovate in the field of implantable medical devices. Smart implants are already being widely used in healthcare including pacemakers, cochlear implants, glucose monitors and insulin pumps however, a widely used smart THR has not yet been realised. Since the implantation of the first instrumented hip implant in the 1960s there have been several in vitro studies monitoring levels of implant loosening. Additionally, significant research has been conducted using instrumented THRs to perform in vivo measurement of biomechanical metrics, including force and moments. To date less than 100 patients have successfully received an instrumented implant. The results of these studies have aided researchers, designers and surgeons in wider research projects, however, the motivation behind the work was to provide discrete biomechanical data sets and not provide real-time condition monitoring of an implants performance or highlight early indications for revision surgery. If in vivo sensing within a THR is to be achieved and adopted in regular clinical practice then the following challenges need to be addressed: choice of the sensing method, biocompatibility and integration within the implant, power supply, communication, and regulatory considerations.
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
全髋关节置换术中的体内传感是可能的吗?回顾过去的制度和未来的挑战
手术植入全髋关节置换术(THR)在减轻疼痛和恢复功能方面非常成功。这一程序已变得更加普遍,预计需求将进一步增加。然而,并发症可能会出现,目前的诊断技术往往不能在导致需要翻修手术的灾难性失败之前暴露潜在的问题。一种带有嵌入式传感器的仪器植入物,能够实时监测状态,将是一个有吸引力的提议,纳入THR。电子元件、嵌入式系统、传感和无线通信的性能和小型化的持续进步,为医疗设备制造商提供了在植入式医疗设备领域进行创新所需的工具和资源。智能植入物已经广泛应用于医疗保健领域,包括起搏器、耳蜗植入物、血糖监测仪和胰岛素泵,然而,广泛使用的智能THR尚未实现。自20世纪60年代首次植入人工髋关节以来,已经进行了几项体外研究,监测人工髋关节松动的水平。此外,已经进行了大量研究,使用仪器化的THRs进行生物力学指标的体内测量,包括力和力矩。迄今为止,只有不到100名患者成功接受了器械植入。这些研究的结果在更广泛的研究项目中为研究人员、设计师和外科医生提供了帮助,然而,这项工作背后的动机是提供离散的生物力学数据集,而不是提供植入物性能的实时状态监测或突出修复手术的早期适应症。如果要在THR内实现体内传感并在常规临床实践中采用,则需要解决以下挑战:传感方法的选择、植入物内的生物相容性和整合、电源、通信和监管考虑。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
CiteScore
9.40
自引率
0.00%
发文量
0
期刊最新文献
Lifetime engineering of bioelectronic implants with mechanically reliable thin film encapsulations The stochastic digital human is now enrolling for in silico imaging trials – Methods and tools for generating digital cohorts Updates on polyurethane and its multifunctional applications in biomedical engineering Wearable facial electromyography: in the face of new opportunities A systematic review of cardiac in-silico clinical trials.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1