Laurent Maïmoun , Sandrine Alonso , Krishna Kunal Mahadea , Vincent Boudousq , Thibault Mura , Denis Mariano-Goulart
{"title":"Cross-Calibration Study of The Stratos And Hologic QDR 4500A Dual-Energy X-Ray Absorptiometers to Assess Bone Mineral Density And Body Composition","authors":"Laurent Maïmoun , Sandrine Alonso , Krishna Kunal Mahadea , Vincent Boudousq , Thibault Mura , Denis Mariano-Goulart","doi":"10.1016/j.jocd.2023.101434","DOIUrl":null,"url":null,"abstract":"<div><p>The objective of the study was to assess the agreement between the Stratos (DMS) and QDR 4500A (Hologic) DXAs in determining whole body and regional aBMD, as well as whole body composition.</p><p>Fifty-five individuals (46 women: 84%) with a mean age of 41 ± 13.0 years (range: 20 to 64) and a mean BMI of 31.9 ± 10 kg/m² (range: 12.2 to 49.5) were consecutively scanned on the same day using the two devices. Predictive equations for areal bone mineral density (aBMD) and whole body composition (WBC) were derived from linear regression of the data.</p><p>The two DXAs were highly correlated (p<0.001 for all parameters) with a correlation coefficient (r) ranging from 0.89 to 0.99 for aBMD (r=0.89 for whole body, r=0.92 for radius, r=0.95 for femoral neck, r=0.96 for total hip, and r=0.99 for L1-L4). For WBC, the r value was 0.98 for lean tissue mass (LTM) and 1.0 for fat mass (FM). Paired t-tests indicated a statistically significant bias between the two DXAs for the majority of measurements, requiring the determination of specific cross-calibration equations. Compared to QDR 4500A, Stratos underestimated whole body aBMD and LTM and overestimated neck and hip aBMD and whole body FM. Conversely, no significant bias was demonstrated for mean aBMD at L1-L4 and radius. For whole body aBMD and FM, the concordance between the two DXAs was influenced by BMI.</p><p>Despite a high concordance between the two DXAs, the systematic bias for aBMD and WBC measurements illustrates the need to define cross-calibration equations to compare data across systems.</p></div>","PeriodicalId":1,"journal":{"name":"Accounts of Chemical Research","volume":null,"pages":null},"PeriodicalIF":16.4000,"publicationDate":"2023-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Accounts of Chemical Research","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1094695023000847","RegionNum":1,"RegionCategory":"化学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"CHEMISTRY, MULTIDISCIPLINARY","Score":null,"Total":0}
引用次数: 0
Abstract
The objective of the study was to assess the agreement between the Stratos (DMS) and QDR 4500A (Hologic) DXAs in determining whole body and regional aBMD, as well as whole body composition.
Fifty-five individuals (46 women: 84%) with a mean age of 41 ± 13.0 years (range: 20 to 64) and a mean BMI of 31.9 ± 10 kg/m² (range: 12.2 to 49.5) were consecutively scanned on the same day using the two devices. Predictive equations for areal bone mineral density (aBMD) and whole body composition (WBC) were derived from linear regression of the data.
The two DXAs were highly correlated (p<0.001 for all parameters) with a correlation coefficient (r) ranging from 0.89 to 0.99 for aBMD (r=0.89 for whole body, r=0.92 for radius, r=0.95 for femoral neck, r=0.96 for total hip, and r=0.99 for L1-L4). For WBC, the r value was 0.98 for lean tissue mass (LTM) and 1.0 for fat mass (FM). Paired t-tests indicated a statistically significant bias between the two DXAs for the majority of measurements, requiring the determination of specific cross-calibration equations. Compared to QDR 4500A, Stratos underestimated whole body aBMD and LTM and overestimated neck and hip aBMD and whole body FM. Conversely, no significant bias was demonstrated for mean aBMD at L1-L4 and radius. For whole body aBMD and FM, the concordance between the two DXAs was influenced by BMI.
Despite a high concordance between the two DXAs, the systematic bias for aBMD and WBC measurements illustrates the need to define cross-calibration equations to compare data across systems.
期刊介绍:
Accounts of Chemical Research presents short, concise and critical articles offering easy-to-read overviews of basic research and applications in all areas of chemistry and biochemistry. These short reviews focus on research from the author’s own laboratory and are designed to teach the reader about a research project. In addition, Accounts of Chemical Research publishes commentaries that give an informed opinion on a current research problem. Special Issues online are devoted to a single topic of unusual activity and significance.
Accounts of Chemical Research replaces the traditional article abstract with an article "Conspectus." These entries synopsize the research affording the reader a closer look at the content and significance of an article. Through this provision of a more detailed description of the article contents, the Conspectus enhances the article's discoverability by search engines and the exposure for the research.