Activated partial thromboplastin time. A multicenter evaluation of 11 reagents in the screening of mild haemophilia A

N. Ciavarella, S. Coccheri, P. Mannucci, M. Canciani, G. Mariani, P. Mori, M. Orlando, L. Tentori, O. Ponari
{"title":"Activated partial thromboplastin time. A multicenter evaluation of 11 reagents in the screening of mild haemophilia A","authors":"N. Ciavarella, S. Coccheri, P. Mannucci, M. Canciani, G. Mariani, P. Mori, M. Orlando, L. Tentori, O. Ponari","doi":"10.1111/J.1600-0609.1981.TB01408.X","DOIUrl":null,"url":null,"abstract":"An internationally standardized preparation and 10 commercial kits widely used to perform the activated partial thromboplastin time (APTT) were compared in 4 laboratories for the purpose of assessing their ability to detect mild deficiencies of factor VIII activity. The participating laboratories were asked to carry out with each APTT reagent quadruplicate readings of 3 coded lyophilized plasmas containing varying levels of factor VIII (109, 26 and 17 U/dl respectively). An analysis of variance of clotting times showed significant differences between reagents and laboratories. All the reagents detected the abnormality of the plasma containing 17 U/dl, whereas a number of failures were found when the plasma with 26 U/dl was tested. When analysis of variance was carried out on ratios of factor-VIII deficient to normal plasma clotting times, the results showed less difference between laboratories and reagents. Clotting times of plasma with normal factor VIII level (109 U/dl) usually fell within the normal range indicated by manufacturers of the commercial reagents.","PeriodicalId":21489,"journal":{"name":"Scandinavian journal of haematology","volume":"75 1","pages":"308-317"},"PeriodicalIF":0.0000,"publicationDate":"2009-04-24","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"10","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Scandinavian journal of haematology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1111/J.1600-0609.1981.TB01408.X","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 10

Abstract

An internationally standardized preparation and 10 commercial kits widely used to perform the activated partial thromboplastin time (APTT) were compared in 4 laboratories for the purpose of assessing their ability to detect mild deficiencies of factor VIII activity. The participating laboratories were asked to carry out with each APTT reagent quadruplicate readings of 3 coded lyophilized plasmas containing varying levels of factor VIII (109, 26 and 17 U/dl respectively). An analysis of variance of clotting times showed significant differences between reagents and laboratories. All the reagents detected the abnormality of the plasma containing 17 U/dl, whereas a number of failures were found when the plasma with 26 U/dl was tested. When analysis of variance was carried out on ratios of factor-VIII deficient to normal plasma clotting times, the results showed less difference between laboratories and reagents. Clotting times of plasma with normal factor VIII level (109 U/dl) usually fell within the normal range indicated by manufacturers of the commercial reagents.
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
活化部分凝血活酶时间。11种试剂在轻度血友病A筛查中的多中心评价
一种国际标准化制剂和10种广泛用于执行活化部分凝血活素时间(APTT)的商用试剂盒在4个实验室进行比较,以评估其检测因子VIII活性轻度缺陷的能力。参与实验的实验室被要求使用每种APTT试剂对含有不同水平因子VIII(分别为109、26和17 U/dl)的3种编码冻干血浆进行四次重复读数。对凝血时间的方差分析显示试剂和实验室之间存在显著差异。所有的试剂都检测到含有17u /dl的血浆的异常,而当检测含有26u /dl的血浆时,发现了许多失败。当对因子- viii缺乏与正常血浆凝血时间的比率进行方差分析时,结果显示实验室和试剂之间的差异较小。血浆凝血时间正常因子VIII水平(109 U/dl)通常落在商业试剂制造商指示的正常范围内。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
Pyridoxine-responsive primary acquired sideroblastic anaemia. In vitro and in vivo effects of vitamin B6 on decreased 5-aminolaevulinate synthase activity. Permanent fixation of lymphoid cells exposed to dye exclusion test. Acute Renal Failure in Hereditary Spherocytosis. Report of a Case Studies on lymphocytes XIII. Nuclear volume measurement as a rapid approach to estimate proliferative fraction. The Influence of Technical Factors on the NBT Test
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1